SPEEDBAND SUPERVIEW SUPER 7, INJECTION SPEEDBAND SUPERVIEW SUPER 7, MODELS 4205 AND 4238

K020824 · Boston Scientific Corp · FHN · Apr 12, 2002 · Gastroenterology, Urology

Device Facts

Record IDK020824
Device NameSPEEDBAND SUPERVIEW SUPER 7, INJECTION SPEEDBAND SUPERVIEW SUPER 7, MODELS 4205 AND 4238
ApplicantBoston Scientific Corp
Product CodeFHN · Gastroenterology, Urology
Decision DateApr 12, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Speedband™ Superview Super7 is indicated for use in endoscopic ligation of esophageal varices and anorectal hemorrhoids. The Injection Speedband™ Superview Super 7 is indicated for endoscopic treatment of esophageal varices utilizing combination ligation therapy. The Injection Speedband™ Superview Super7 is also indicated for band ligation treatment of anorectal hemorrhoids.

Device Story

Endoscopic ligator for esophageal varices and anorectal hemorrhoids; consists of ligating unit and handle assembly. Ligating unit features cylindrical housing with flexible silicone endoscope adapter and seven elastic ligating bands; suture threaded around bands passes through working channel of endoscope to handle assembly. Handle assembly includes knob, spool, and tripwire; tripwire secures suture during setup. Injection model includes septum port for injection therapy needle. Operated by physician via endoscope; suction maintained through septum. Device enables mechanical ligation of varices/hemorrhoids; injection model allows concurrent injection therapy. Benefits include minimally invasive treatment of variceal/hemorrhoidal conditions.

Clinical Evidence

Bench testing only.

Technological Characteristics

Ligating unit with cylindrical housing, silicone endoscope adapter, and seven elastic ligating bands. Handle assembly with knob, spool, and tripwire. Injection model includes septum with injection therapy needle port. Manual operation via endoscope working channel.

Indications for Use

Indicated for patients requiring endoscopic ligation of esophageal varices or anorectal hemorrhoids. Injection model indicated for combination ligation/injection therapy for esophageal varices.

Regulatory Classification

Identification

A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.

Special Controls

*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020824 Page 1 of 2 Image /page/0/Picture/2 description: The image shows the logo for Boston Scientific Microvasive. The words "Boston Scientific" are stacked on top of each other in a serif font, with a line underneath. Below that, the word "MICROVASIVE" is written in a bold, sans-serif font. APR 1 2 2002 MA 01760- www.bsci.com Section 9 510(k) Summary #### 510(K) SUMMARY #### 1. SUBMITTER: Boston Scientific Corporation One Boston Scientific Place Natick, MA 01760 Contact: Kathleen Morahan, Regulatory Affairs Manager Date Prepared: March 12, 2002 ### 2. DEVICE: Trade Name: Speedband™ Superview Super7 Injection Speedband™ Superview Super7 Common Name: Hemorrhoidal Ligator and Esophageal Variceal Ligator Class: II Classification Name: Ligator, Hemorrhoidal (78 FHN) Ligator, Esophageal (78 MND) ## 3. PREDICATE DEVICE: Modified Injection Speedband TM Superview Modified Speedband™ Superview ## 4. DEVICE DESCRIPTION: The Speedband Superview Super7 and Injection Speedband Superview Super7 consist of two main components; a ligating unit and a handle assembly. The ligating unit consists of a cylindrical housing with a flexible silicone endoscope adapter, and seven elastic ligating bands. Suture is threaded around each band and passes through suture slots around the perimeter of the housing. The suture enters the endoscope through the working channel of an endoscope and is secured the tripwire on the handle assembly durning set-up. Confidential {1}------------------------------------------------ The handle assembly consists of a handle knob and plastic spool mounted on a bracket with a stem. A tripwire is attached to the spool and passes through the septum and into the stem. The stem fits into the working channel of the endoscope. The septum seals the channel so suction can be maintained. On the Injection models, the septum has a port for insertion of an injection therapy needle. ### 5. INTENDED USE: The Speedband™ Superview Super7 is indicated for use in endoscopic ligation of esophageal varices and anorectal hemorrhoids. The Injection Speedband™ Superview Super 7 is indicated for endoscopic treatment of esophageal varices utilizing combination ligation therapy. The Injection Speedband™ Superview Super7 is also indicated for band ligation treatment of anorectal hemorrhoids. ## 6. TECHNOLOGICAL CHARACTERISTICS: The intended use and the materials are identical predicate devices. The proposed ligating unit contains seven bands whereas the predicate device has up to eight bands. Other minor housing and band modifications were made. The proposed and predicate device handle assemblies are the same except for an increase in the handle knob diameter and shape for ergonomics. #### 7. PERFORMANCE DATA: Bench testing was conducted to aid with the establishment of substantial equivalence of the proposed devices to the predicate device. #### 8. CONCLUSION: BSC has demonstrated that the Speedband Superview Super 7 and the Injection Speedband Superview Super 7 are substantially equivalent to the predicate devices. {2}------------------------------------------------ 510(k) Number: # KO20824 Device Name: Speedband™ Superview Super7 Indication for Use: The Speedband™ Superview Super7 is indicated for use in endoscopic ligation of esophageal varices and anorectal hemorrhoids. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.1091) (Optional Format 1-2-96) OR Over-The-Counter Use {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is an abstract image of a human figure with three lines extending from the head, resembling wings or feathers. # APR 1 2 2002 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Kathleen Morahan Manager, Regulatory Affairs Microvasive Endoscopy Boston Scientific Corporation One Boston Scientific Place NATICK MA 01760-1537 Re: K020824 ﻬﺎ ﺑﻬ Trade/Device Name: Injection Speedband™ Superview Super7TM, Model 4238 and Speedband™M Superview Super7TM, Model 4205 Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Codes: 78 FHN and 78 MND Dated: March 13, 2002 Received: March 14, 2002 Dear Ms. Morahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### K020824 510(k) Number: Device Name: Injection Speedband™ Superview Super7 ## Indication for Use: The Injection Speedband™ Superview Super7 is indicated for use in endoscopic treatment of esophageal varices utilizing combination ligation/injection therapy. The Injection Speedband™ Superview Super7 is also indicated for band ligation treatment of anorectal hemorrhoids. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.1091)<br>(Optional Format 1-2-96) | | |-----------------------------------------------------------------------|---------------------------------| | OR | Over-The-Counter Use __________ | (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K020824 | |---------------|---------| |---------------|---------| Confidential 000029
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...