K973251 · KARL STORZ Endoscopy-America, Inc. · FEO · Jun 22, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K973251
Device Name
KSEA CALCUSON
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
FEO · Gastroenterology, Urology
Decision Date
Jun 22, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4480
Device Class
Class 2
Attributes
Therapeutic
Intended Use
These instruments are intended for use by qualified surgeons and provide for the fragmentation of renal, ureteral and bladder calculi.
Device Story
The Karl Storz Calcuson is an ultrasonic lithotriptor designed for the fragmentation of renal, ureteral, and bladder calculi. It is operated by qualified surgeons in a clinical or surgical setting. The device utilizes ultrasonic energy to break down stones into smaller fragments, facilitating their removal. The surgeon controls the device to target calculi; the ultrasonic energy provides the mechanical force necessary for fragmentation. This process assists in the treatment of urolithiasis, potentially reducing the need for more invasive surgical procedures for stone extraction.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Ultrasonic lithotriptor; utilizes ultrasonic energy for stone fragmentation. Classified under 21 CFR 876.4480 (Product codes: 78 FFK and FEO).
Indications for Use
Indicated for fragmentation of renal, ureteral, and bladder calculi in patients requiring surgical intervention for stone removal.
Regulatory Classification
Identification
An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”
Related Devices
K993046 — FIBRA SONICS ULTRASONIC LITHOTRIPTOR LITHOSONIC, MODEL LS-2000 · Fibra Sonics, Inc. · Nov 22, 1999
K160198 — UreTron MultiProbe Lithotripter System · Med-Sonics Corp. · Feb 23, 2016
K011700 — MODULITH SLK LITHOTRIPTER · Karl Storz Endoscopy · Aug 16, 2002
K052680 — CALCULASE HOLMIUM LASER SYSTEM AND ACCESSORIES · KARL STORZ Endoscopy-America, Inc. · Jan 23, 2006
K042427 — GENESTONE EXEL (XL) EXTRACORPOREAL SHOCK WAVE LITHOTRIPTER · Genemed, Inc. · Dec 3, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for Karl Storz Endoscopy. The logo is vertically oriented with the words "Karl Storz Endoscopy" stacked on the left side. The word "STORZ" is written in large, bold letters on the right side of the logo. The letter "O" in "STORZ" is stylized as a circle.
K916335
## 2661 2 2 A CONSULER ONA ATTANS AO ANAMINUS (A) OF S
NOC
roghelwould a A.H.S. In the comments of them and our to currently of theming in the dir. 114 Co Tox Яг. Соведов 20 (АПМӘ) — Алама) — Салан (1)ОТ 30 хташты кіңі
| Applicant: | Karl Storz Endoscopy - America, Inc. 600 Corporate Pointe Drive Culver City, CA 90230 (310) 558-1500 |
|------------------------|---------------------------------------------------------------------------------------------------------|
| Contact: | Kevin Kennan Regulatory Affairs Specialist |
| Device Identification: | |
| Common Name: | Ultrasonic Lithotriptor |
| Trade Name: (optional) | |
calculi visninu 10 notistnemark sinossalu att 10 bengizeb 21 metrys novelse zuote 1187 1187 118118215017
Karl Storz Calcuson
noitisoquoo bus sezis tsom to ilusles neemgeri of theiriffus si əsinə sinfi və bəpiyord vəzifə ədə çəkin əldə əldə çəkilə olunmuşdur. Timmə - 10 mandanın mən of parties in the of personales and service commodial oinosusion as si nosusle arotz lus and suasantes and
devices ອະອຸປີ 10 ຈະມ bebruari 10 ແບ່ນວານນີ້ ລວມຄາຕາງກອງ ອປີ ຖວ 10 ຈາກຕ່າ ຂອງກອນອົງງາມ ອຸດາ ຂອງ ເຊຍ ລາວ burğu 30 sənəsi və və və və əsərin əsinə ən buna nozusları məsənəd məx əsinə aid bir nəfər və əsasən bəbinə ini bax sənələri əsasən əsasən əsasən ələnindən Substantial Equivalence: and of the consisted on included on and and and and and and
:paugiṣṣ Kevin Kevin Kevin Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Keving Ke
Regulatory Affairs Specialist
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 22 1998
Mr. Kevin Kennan Senior Regulatory Affairs Specialist Karl Storz Endoscopy-America, Inc. 600 Corporate Pointe Culver City, California 90230-7600 Re: K973251
Calcuson Ultrasonic Lithotripter and Accessories Dated: August 29, 1997 Received: August 29, 1997 Regulatory class: III 21 CFR §876.4480/Product codes: 78 FFK and FEO
Dear Mr. Kennan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
hiliau Yu
Lillian Yin, Ph.D.
Lillian Yin, Ph.D Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
510(k) Number (if known): Not yet assigned (9 7 3 J T S 7 3 J 1
Device Name: Calcuson Ultrasonic Lithotriptor
Indications for Use: These instruments are intended for use by qualified surgeons and provide for the fragmentation of renal, ureteral and bladder calculi.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div style="text-align:left;"> <p>Robert R. Sathing/</p> <p>(Division Sign-Off)</p> <p>Division of Reproductive, Abdominal, ENT, and Radiological Devices</p> </div> |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K973251 |
| Prescription Use: | <div style="display:inline-block;"> ✓ </div> |
|----------------------|--------------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
OR Over-The-Counter Use:
(Optional Format 1-2-96)000003
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.