← Product Code [FDI](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI) · K183289

# ClearGrasp Snare (K183289)

_Finemedix Co., Ltd. · FDI · Mar 6, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K183289

## Device Facts

- **Applicant:** Finemedix Co., Ltd.
- **Product Code:** [FDI](/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI.md)
- **Decision Date:** Mar 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Intended Use

The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.

## Device Story

ClearGrasp Snare is a monopolar electrosurgical instrument for endoscopic mucosal resection (EMR). Device consists of flexible rope wire loop, catheter tube, and handle with control knobs (handle and slider). Operated by physician during endoscopic procedures; device passes through endoscope working channel. User selects loop shape (oval or crescent), size, and rotational capability based on lesion characteristics. When connected to an electrosurgical generator, loop delivers monopolar current to resect targeted GI tract lesions. Single-patient use; supplied sterile. Benefits include precise grasping and resection of polyps, facilitating removal during standard GI endoscopy.

## Clinical Evidence

Bench testing only. Biocompatibility testing per ISO 10993-1, -5, -10, -11. Performance testing per IEC 60601-2-2 and ANSI/AAMI HF18 (appearance, dimension, continuity, tensile strength, withstand voltage, leakage current). Sterilization validation per ISO 11737 and shelf-life testing per ASTM F1980.

## Technological Characteristics

Monopolar electrosurgical snare. Materials: SUS 304, PTFE, ABS. Dimensions: 1600-2400mm working length; 6-30mm loop sizes. Connectivity: Connects to external electrosurgical generator. Sterilization: Ethylene Oxide. Design: Flexible catheter with handle-actuated loop extension/rotation. No software/algorithm.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Polypectomy Snare ([K172729](/device/K172729.md))
- Lariat Snare ([K151197](/device/K151197.md))
- Rotatable Snares, Non-Rotatable Snares ([K172758](/device/K172758.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

March 6, 2019

FINEMEDIX CO., LTD. % April Lee Consultant Withus Group, Inc. 106 Superior Irvine, CA 92620

Re: K183289

> Trade/Device Name: ClearGrasp Snare Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FDI Dated: January 29, 2019 Received: February 5, 2019

Dear April Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark J. Antonino -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183289

Device Name ClearGrasp Snare

Indications for Use (Describe)

The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, pedunculated polyps and tissue from within the GI tract.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for FINEMEDIX. The logo consists of the word "FINEMEDIX" in a bold, sans-serif font, with an orange circle above the "E" in "FINEMEDIX". The word "FINEMEDIX" is in a dark blue color. The logo is simple and modern.

#### 510(k) Summary

#### Submitter

FINEMEDIX CO., LTD. Hee Kyung Kwon 60, Maeyeo-ro, Dong-gu, Daegu 41065 South Korea Email: hkyung1230@finemedix.com Tel. +82-53-741-8388 Fax. +82-53-741-8168

#### Official Correspondent

Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122

#### Device Information

- Trade Name: ClearGrasp Snare
- Common Name: Snare, Flexible
- Classification Name: Endoscopic electrosurgical unit and accessories
- Product Code: FDI
- Panel: Gastroenterology/Urology
- Regulation Number: 21 CFR 876.4300
- Device Class: Class II ●
- . Date Prepared: 11/23/2018

#### Predicate Devices:

The subject device is substantially equivalent to the following predicate devices:

- . K 172729, Polypectomy Snare manufactured by Hangzhou AGS MedTech Co., Ltd.
- K151197, Lariat Snare by United States Endoscopy Group, Inc. ●
- K172758, Rotatable Snares, Non-Rotatable Snares by Zhejiang Chuangxiang Medical Technology ● Co., LTD.

#### Indication for Use:

The ClearGrasp Snare is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.

#### Device Description:

ClearGrasp Snare is a monopolar electrosurgical instrument intended for EMR which is Endoscopic Mucosal Resection by applying high-frequency current during endoscopic electrosurgical procedures. It is used for grasping and resecting the targeted lesion within the digestive tract by opening and closing the loop. Users can choose an oval or a crescent type based on their preference and the characteristic of the lesion. The loop sizes can be chosen in accordance with the size of the lesion and the rotational type helps approaching and grasping the lesion. There is a stiff type distinguished by the strength of loop.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "FINEMEDIX" in large, bold, purple letters. Above the "E" in "FINEMEDIX" is a small orange circle. The background of the image is white. 

The device consists of a flexible Rope Wire and various size of loop, which can be extended and rotated from the flexible Catheter Tube using two handle control knobs such as Handle and Slider. When connected to an electrosurgical generator and activated, the loop delivers a monopolar electrical current to the surgical site. This device passes through the working channel of endoscope, and the average contact time of the product and the mucosa of the human digestive tract is less than 1 hour. This device is supplied sterile for single-patient use and shall be not reused or re-sterilized.

|                                  | Subject Device                                                                                                                                                                       | Primary<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference Device                                                                                                                                                                   | Reference Device                                                                                                                                                                                                               |                                              |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Company                          | Finemedix Co.,<br>Ltd.                                                                                                                                                               | Hangzhou AGS<br>MedTech Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | United States<br>Endoscopy Group,<br>Inc.                                                                                                                                          | Zhejiang Chuangxiang<br>Medical Technology<br>Co., LTD.                                                                                                                                                                        |                                              |
| Device Name                      | ClearGrasp Snare                                                                                                                                                                     | Polypectomy Snare                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Lariat Snare                                                                                                                                                                       | Rotatable Snares,<br>Non-Rotatable Snares                                                                                                                                                                                      |                                              |
| 510(k) Number                    | NA                                                                                                                                                                                   | K172729                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K151197                                                                                                                                                                            | K172758                                                                                                                                                                                                                        |                                              |
| Device<br>Classification<br>Name | Endoscopic<br>electrosurgical unit<br>and accessories                                                                                                                                | Endoscopic<br>electrosurgical unit<br>and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Endoscopic<br>electrosurgical unit<br>and accessories                                                                                                                              | Endoscopic<br>electrosurgical unit<br>and accessories                                                                                                                                                                          |                                              |
| Product Code                     | FDI                                                                                                                                                                                  | FDI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | FDI                                                                                                                                                                                | FDI                                                                                                                                                                                                                            |                                              |
| Regulation<br>Number             | 876.4300                                                                                                                                                                             | 876.4300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 876.4300                                                                                                                                                                           | 876.4300                                                                                                                                                                                                                       |                                              |
| Indications for Use              | The ClearGrasp<br>Snare is used<br>endoscopically in<br>the removal of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated<br>polyps and tissue<br>from within the GI<br>tract. | The Polypectomy<br>Snare is used<br>endoscopically in<br>the removal of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated<br>polyps and tissue<br>from within the GI<br>tract.                                                                                                                                                                                                                                                                                                                   | The Lariat snare is<br>an electrosurgical<br>device designed to<br>be used to<br>endoscopically<br>grasp, dissect and<br>transect tissue<br>during GI<br>endoscopic<br>procedures. | The Polypectomy<br>Snare is used<br>endoscopically in the<br>removal and/or and<br>cauterization of<br>diminutive polyps,<br>sessile polyps,<br>pedunculated polys<br>and tissue from within<br>the gastrointestinal<br>tract. |                                              |
| Principle of<br>Operation        | Monopolar<br>electrosurgical<br>instrument<br>intended for EMR                                                                                                                       | Monopolar<br>electrosurgical<br>instrument<br>intended for EMR                                                                                                                                                                                                                                                                                                                                                                                                                                          | Monopolar<br>electrosurgical<br>instrument intended<br>for EMR                                                                                                                     | Monopolar<br>electrosurgical<br>instrument intended<br>for EMR                                                                                                                                                                 |                                              |
|                                  | Loop<br>Catheter<br>Tube<br>Handle                                                                                                                                                   | SUS 304                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | SUS 304                                                                                                                                                                            | Unknown                                                                                                                                                                                                                        | Unknown                                      |
| Material                         |                                                                                                                                                                                      | PTFE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | PTFE                                                                                                                                                                               |                                                                                                                                                                                                                                |                                              |
|                                  |                                                                                                                                                                                      | ABS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ABS                                                                                                                                                                                |                                                                                                                                                                                                                                |                                              |
|                                  | Components                                                                                                                                                                           | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug                                                                                                                                    | Loop, Catheter<br>Tube, Handle,<br>Slider, Plug                                                                                                                                                                                | Loop, Catheter Tube,<br>Handle, Slider, Plug |
|                                  | Shape of Loop                                                                                                                                                                        | Oval and Crescent                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Oval, Polygon,<br>Duck Bill,<br>Hexagonal and<br>Round                                                                                                                             | Oval, Hexagonal<br>and Diamond                                                                                                                                                                                                 | Oval, Hexagonal,<br>Crescent, Round          |
|                                  | Working Length                                                                                                                                                                       | 1600,1800,2200,<br>2400 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 1800, 2300 mm                                                                                                                                                                      | 2300mm                                                                                                                                                                                                                         | 1600, 1800, 2400mm                           |
|                                  | Loop Size                                                                                                                                                                            | 6,10,13,15,<br>24,30 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 10,15,25,32 mm                                                                                                                                                                     | 6,10,30mm                                                                                                                                                                                                                      | 10,15,22,25,32mm                             |
|                                  | Rotational/Non-<br>Rotational                                                                                                                                                        | Both                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Both                                                                                                                                                                               | Non-Rotational                                                                                                                                                                                                                 | Both                                         |
|                                  | Sterile/Non-Sterile                                                                                                                                                                  | Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Sterile                                                                                                                                                                            | Sterile                                                                                                                                                                                                                        | Sterile                                      |
|                                  | Sterilization<br>Method                                                                                                                                                              | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ethylene Oxide                                                                                                                                                                     | Ethylene Oxide                                                                                                                                                                                                                 | Ethylene Oxide                               |
|                                  | Single Use                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                | Yes                                                                                                                                                                                                                            | Yes                                          |
|                                  | Used with<br>Electrosurgical<br>Unit                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                | Yes                                                                                                                                                                                                                            | Yes                                          |
|                                  | Similarities                                                                                                                                                                         | The subject device is substantially equivalent to the predicate device in Indications for Use,<br>materials, components, shape of loop, rotationality, sterilization methods and principle of<br>operation.                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                    |                                                                                                                                                                                                                                |                                              |
|                                  | Differences                                                                                                                                                                          | The difference between the subject and primary predicate are shape of loop, working<br>length, and loop size. To support the difference of the shape of loop such as the crescent<br>shape and working length such as 1600 and 2400mm, we added the K172758 as the<br>reference device. To support the difference of the loop size such as 6mm, we added the<br>K151197 as the reference device. These differences do not raise any questions of<br>substantial equivalence to the declared predicates. |                                                                                                                                                                                    |                                                                                                                                                                                                                                |                                              |

### Summary of Technological Characteristics:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in a dark blue font, with a small orange circle above the letters 'NE'. The font is sans-serif and appears to be bolded.

## FINEMEDIX CO., LTD.

60, Maeyeo-ro, Dong-gu, Daegu, Republic of Korea (41065)

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for FINEMEDIX. The logo is in a dark blue color, with an orange circle above the letters 'ME'. The text is in all capital letters and is a sans-serif font. The logo is simple and modern.

#### Non-clinical testing data:

The subject device was tested to evaluate its substantial equivalence according to the following standards.

- Biocompatibility testing according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 ● and ISO 10993-11:2006
- . Performance testing such as appearance, dimension, continuity, tensile strength test of electrode cable, withstand voltage, high frequency leakage current test of electrode cable according to IEC 60601-2-2:2017 and ANSI/AAMI HF18:2001
- EO Sterilization residuals testing according to ISO 10993-7:2008

Below tests were performed on our own device and they can be leveraged for the subject device:

- EO Sterilization Validation Testing according to ISO 11737-1:2006 and ISO 11737-2:2009 ●
- . Shelf Life Testing according to ASTM F1980

The biocompatibility evaluation for ClearGrasp Snare was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing within a Risk Management Process " and FDA's biocompatibility guidance, G95-1 Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" (May 1, 1995). The following tests were completed: Cytotoxicity, Sensitization, Intracutaneous reactivity, Acute Systemic Toxicity and Pyrogen Test.

Performance testing such as appearance, dimension, and continuity were performed as per Finemedix's design control system.

The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.

#### Conclusion:

ClearGrasp Snare constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate device. Therefore, ClearGrasp Snare and its predicate are substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K183289](https://fda.innolitics.com/submissions/GU/subpart-e%E2%80%94surgical-devices/FDI/K183289)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
