COAGULOOP

K991314 · American Medical Systems, Inc. · FAS · Jun 29, 1999 · Gastroenterology, Urology

Device Facts

Record IDK991314
Device NameCOAGULOOP
ApplicantAmerican Medical Systems, Inc.
Product CodeFAS · Gastroenterology, Urology
Decision DateJun 29, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The AMS Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder.

Device Story

Electrosurgical device; used with resectoscopes and monopolar electrosurgical generators. Input: radio frequency electrical current. Function: removes and coagulates soft tissue via RF energy. Used in clinical settings (e.g., OR) by physicians. Output: tissue resection/coagulation. Benefit: enables surgical removal and hemostasis of prostate/bladder tissue.

Clinical Evidence

Bench testing only. Biocompatibility testing followed ISO-10993 recommendations. Sterilization validated per ANSI/AAMI and ISO 11137 standards.

Technological Characteristics

Monopolar electrosurgical resection electrode; compatible with standard resectoscopes and generators. Sterilization: gamma radiation (SAL 10^-6). Biocompatibility: ISO-10993 compliant.

Indications for Use

Indicated for resection, ablation, and fulguration of prostate gland and bladder tissue.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 2 9 1999 Image /page/0/Picture/2 description: The image is a black and white logo. The logo is an abstract design that resembles the letter 'A'. The 'A' is formed by a series of vertical and horizontal lines. The lines are arranged in a way that creates a sense of depth and dimension. The logo is simple, yet visually striking. # AMERICAN MEDICAL SYSTEMS # 510(k) SUMMARY #### SUBMITTER I. Name and Address: American Medical Systems, Inc. 10700 Bren Road West Minnetonka, MN 55343 USA 2183959 David Worrell April 15, 1999 2183959 - Establishment Registration Number: Contact Person: Date of Summary Preparation: #### II. MANUFACTURER Name and Address: Device Trade Name: Classification Name: American Medical Systems 10700 Bren Road West Minnetonka, MN 55343-9679 Establishment Registration Number #### III. DEVICE NAME Device Common or Usual Name: Resection Loop or Resection Electrode AMS Coaguloop™ Resection Electrode Endoscopic electrosurgical unit with accessories (21 CFR 876.4300) {1}------------------------------------------------ #### DEVICE DESCRIPTION IV. The AMS Coaguloop™ Resection Electrode is an electrosurgical device designed for use with commercially available resectoscopes and general purpose monopolar electrosurgical generators. The Coaguloop™ Resection Electrode is designed to remove and coagulate soft tissue by means of radio frequency electrical current. #### V. INDICATION FOR USE The Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder. #### SUBSTANTIAL EQUIVALENCE SUPPORT VI. American Medical Systems' Coaguloop™ Resection Electrode is similar in indications, design, and features to various resectoscope electrodes that are in commercial distribution. The predicate devices that AMS is claiming substantial equivalence to are the AMS Coaguloop™ Resection Electrode K971512 and the Circon/ACMI Vaportome™ Resection Loop Electrode K973820. #### STERILIZATION METHODOLOGY VII. Sterilization is based on the recommendations in ANSI/AAMI and ISO 11137 Guideline for the Sterilization of health care products - Requirements for validation and routine control - Radiation sterilization. The devices are terminally sterilized by exposure to gamma radiation. This gamma radiation sterilization cycle will achieve a minimum Sterility Assurance Level (SAL) of 10°. ### VIII. -BIOCOMPATIBILITY Biocompatibility testing generally followed the recommendations of FDA Blue Book Memorandum #G95-1 entitled "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." ## CONCLUSION In summary, the data and information provided within this 510(k) Premarket Notification adequately support that the AMS Coaguloop™ Resection Electrode is substantially equivalent to the AMS Coaguloop™ Resection Electrode K971512 and the Circon/ACMI Vaportome™ Resection Loop Electrode K973820. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 2 9 1999 Mr. David Worrell Regulatory Affairs Specialist American Medical Systems, Inc. 10700 Bren Road West Minnetonka, MN 55343 Re: K991314 AMS Coaguloop™ Resection Electrode Dated: April 15, 1999 Received: April 16, 1999 Requlatory Class: II 21 CFR §876.4300/Procode: 78 FAS Dear Mr. Worrell: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page _1_ of _1_ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ 510(k) Number (if Known): ____________________________________________________________________________________________________________________________________________________ AMS Coaguloop™ Resection Electrode Device Name: Indications For Use: The AMS Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--|--------------------------------------------------------|--| |--|--------------------------------------------------------|--| | Prescription Use | <div style="text-align:left;">/</div> | |----------------------|---------------------------------------| | (Per 21 CFR 801.109) | | OR | Over-The-Counter Use | _________________ | |----------------------|-------------------| |----------------------|-------------------| | (Division Sign-Off) | | |--------------------------------------------------------------------|---------| | Division of Reproductive, Abdominal, ENT, and Radiological Devices | | | 510(k) Number | K991314 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...