INDICATIONS FOR USE RESECTION ELECTRODES: The Omnitech HF Resection electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation, or removal of soft tissue and where hemostasis is required. The specific urological indications in the prostate, bladder and bladder neck. The procedures for which the devices can be used are: Transurethral resection in saline (TURis). Transurethral prostatectomy, transurethral resection of the prostate (TURP), for benign prostatic hyperplasia, Transurethral incision of the prostate (TUIP) or bladder neck, Transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate. INDICATIONS FOR USE VAPORIZATION ELECTRODE: The Omnitech HF Vaporization electrode for plasma vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving vaporization, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include: Use in the prostate, bladder, and bladder neck. The specific treatment indications include benign prostate hyperplasia (BPH), bladder cancer, tumors, lesions and neoplasms. The specific urological indications include Transurethral Electro vaporization (TUVP), (TVP), (TVP), (TVP), (TVP), (TUEVP) also known as Transurethral Vapor Resection of the prostate (TUVRP), Transurethral Vaporization in Saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.
Device Story
Omnitech HF Resection and Vaporization electrodes are bipolar electrosurgical instruments for endoscopic urological procedures. Devices operate in irrigated environments to perform tissue resection, ablation, vaporization, and coagulation. Used by urologists in clinical settings to treat BPH, bladder tumors, and neoplasms. Electrodes deliver high-frequency electrical energy to target tissue; hemostasis is achieved concurrently with tissue removal. Output is controlled by the surgeon via an electrosurgical generator. Benefits include precise tissue management and reduced bleeding during transurethral procedures.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use.
Technological Characteristics
Bipolar electrosurgical instruments; designed for use in irrigated environments; compatible with standard electrosurgical generators for high-frequency energy delivery.
Indications for Use
Indicated for patients requiring endoscopic urological soft tissue resection, ablation, vaporization, or coagulation in the prostate, bladder, or bladder neck. Specific conditions include benign prostatic hyperplasia (BPH), bladder tumors, lesions, and neoplasms. Contraindicated for the treatment of prostate cancer.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Related Devices
K153055 — ACE Electrosurgical Resection and Vaporization Electrodes Series · Ace Medical Devices Pvt. , Ltd. · Jul 11, 2016
K100275 — WORKING ELEMENT, RESECTION BUTTON ELECTRODE FOR PLASMA VAPORIZATON, HF-RESECTION ELECTRODE LOOPS AND BANDS · Olympus Winter & Ibe GmbH · Jul 1, 2010
K120418 — ELECTROSUGRICAL RESECTION AND VAPORIZATION ELECTRODE SERIES · Olympus Winter & Ibe GmbH · Jun 20, 2012
K050488 — TURBT-RF ELECTRODES AND ELECTROSURGICAL DEVICES · Prosurg, Inc. · Apr 1, 2005
K102781 — RESECTION BUTTON ELECTRODE FOR PLASMA VAPORIZATION, HF-RESECTION ELECTRODE ROLLERS AND NEEDLES, LOOPS AND BANDS · Olympus Winter & Ibe GmbH · Jan 11, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three human profiles facing right, stacked on top of each other, with a stylized design below them that could represent water or waves.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 27, 2016
Omnitech Systems, Inc. Jon Barrett President 450 South Campbell St., Suite #2 Valparaiso. IN 46385
Re: K151976 Trade/Device Name: Omnitech HF Resection Electrodes Omnitech HF Vaporization Electrode Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: FAS Dated: December 18, 2015 Received: May 25, 2016
Dear Jon Barrett,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151976
Device Name Omnitech HF Resection Electrodes Omnitech HF Vaporization Electrode
Indications for Use (Describe)
#### INDICATIONS FOR USE RESECTION ELECTRODES:
The Omnitech HF Resection electrodes are a bipolar instrument series designed and intended for use in endoscopic urological surgical procedures involving the resection, ablation, or removal of soft tissue and where hemostasis is required. The specific urological indications in the prostate, bladder and bladder neck. The procedures for which the devices can be used are: Transurethral resection in saline (TURis). Transurethral prostatectomy, transurethral resection of the prostate (TURP), for benign prostatic hyperplasia, Transurethral incision of the prostate (TUIP) or bladder neck, Transurethral resection of bladder tumors (TURBT) and cystodiathermy. These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.
### INDICATIONS FOR USE VAPORIZATION ELECTRODE:
The Omnitech HF Vaporization electrode for plasma vaporization is a bipolar instrument designed and intended for use in urological surgical procedures involving vaporization, cutting, removal of soft tissue and coagulation where hemostasis is required. The specific soft tissue indications include: Use in the prostate, bladder, and bladder neck. The specific treatment indications include benign prostate hyperplasia (BPH), bladder cancer, tumors, lesions and neoplasms. The specific urological indications include Transurethral Electro vaporization (TUVP), (TVP), (TVP), (TVP), (TVP), (TUEVP) also known as Transurethral Vapor Resection of the prostate (TUVRP), Transurethral Vaporization in Saline (TUVis). These devices are intended to be used in an irrigated environment. These devices are not intended to be used in treating cancer of the prostate.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.