EXPEL NEPHROURETERAL DRAINAGE STENT WITH TWIST-LOC HUB SYSTEM, EXPEL URETERAL DRAINAGE STENT SYSTEM
K141344 · Boston Scientific Corp · FAD · Oct 17, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K141344
Device Name
EXPEL NEPHROURETERAL DRAINAGE STENT WITH TWIST-LOC HUB SYSTEM, EXPEL URETERAL DRAINAGE STENT SYSTEM
Applicant
Boston Scientific Corp
Product Code
FAD · Gastroenterology, Urology
Decision Date
Oct 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Expel™ Nephroureteral Stent System with Twist-Loc™ Hub: The Expel Nephroureteral Stents are delivered percutaneously and are intended to establish internal drainage from the ureteropelvic junction to the bladder while maintaining external access to the stent, as well as providing external drainage. Expel™ Ureteral Stent System: The Expel Ureteral Stent System is delivered percutaneously and is intended to establish drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
Device Story
Single-lumen, nonvascular, indwelling ureteral stents; percutaneously inserted to provide internal/external drainage between ureteropelvic junction and bladder. Nephroureteral version features Twist-Loc™ hub for proximal pigtail locking; Ureteral version uses gravity-based pressure differential. Includes radiopaque marker bands, tapered tips, and hydrophilic coating. Used by clinicians in clinical settings. Accessories include cannulas, stabilizers, plugs/caps, and pigtail straighteners. Stents facilitate urine flow; benefit patients by relieving obstruction or maintaining ureteral patency. Indwelling duration limited to 30 days.
Single-lumen, nonvascular, indwelling stents. Materials include new shaft material with hydrophilic coating. Features radiopaque marker bands. Diameters: 6.3F, 8.3F, 10.3F. Lengths: 12-28 cm. Drainage via gravity/pressure differential. Twist-Loc™ hub mechanism for nephroureteral version. Sterilized device.
Indications for Use
Indicated for patients requiring percutaneous ureteral drainage or stenting from the ureteropelvic junction to the bladder, with or without external access, for up to 30 days.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Percuflex™ Nephroureteral Stent and Percuflex™ Ureteral Stent (K924608)
Related Devices
K161236 — Universa Firm Ureteral Stents and Stent Sets · Cook Incorporated · Jan 27, 2017
K243039 — Ureteral Stents (AF-D series) · Alton (Shanghai) Medical Instruments Co., Ltd. · Jun 18, 2025
K190312 — Dornier Ureteral Stent · Dornier Medtech America, Inc. · Nov 15, 2019
K960148 — CIRCON-SURGITEK CLASSIC DOUBLE PIGTAIL URETERAL STENT · Cabot Medical Corp. · Mar 25, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three bars above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 17, 2014
Boston Scientific Corp Matt Beauchane Regulatory Affairs Specialist One Scimed Place Maple Grove, MN 55311-1566
Re: K141344
Trade/Device Name: Expel™ Nephroureteral Drainage Stent With Twist-Loc Hub System Expel™ Ureteral Drainage Stent System Regulation Number: 21 CFR 876.4620 Regulation Name: Uretal Stent Regulatory Class: Class II Product Code: FAD Dated: October 9, 2014 Received: October 10, 2014
Dear Matt Beauchane.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a simple, sans-serif font. The text is arranged in two lines, with "Herbert P." on the first line and "Lerner -S" on the second line. The letters are black against a white background, and there is a faint, gray FDA logo in the background.
for Benjamin Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141344
Device Name
ExpelTM Nephroureteral Stent System with Twist-LocTM Hub
Indications for Use (Describe)
The Expel Nephroureteral Stents are delivered percutaneously and are intended to establish internal drainage from the ureteropelvic junction to the bladder while maintaining external access to the stent, as well as providing external drainage.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141344
Device Name Expel™ Ureteral Stent System
Indications for Use (Describe)
The Expel Ureteral Stent System is delivered percutaneously and is intended to establish drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Per 21 CFR §807.92
| Submitter's Name<br>and Address | Boston Scientific Corporation<br>One Scimed Place<br>Maple Grove, MN 55311<br>USA |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name<br>and Information | Matt Beauchane<br>Regulatory Affairs Specialist<br>Phone: 763-494-1789<br>Fax: 763-494-2222<br>Email: matt.beauchane@bsci.com |
| Date Prepared | 20-Aug-2014 |
| Proprietary Names | Expel™ Nephroureteral Stent System with Twist-Loc™ Hub |
| | Expel™ Ureteral Stent System |
| Common Name | Ureteral Stent |
| Classification | Classification: Class II<br>Regulation: 21 CFR 876.4620<br>Product Code: FAD<br>Classification Panel: Gastroenterology/Urology |
| Predicate Device | Percuflex™ Nephroureteral Stent and Percuflex™ Ureteral Stent<br>(K924608, 26-Jan-1994) |
| Intended Use /<br>Indications for Use | Expel™ Nephroureteral Stent System with Twist-Loc™ Hub: |
| | The Expel Nephroureteral Stents are delivered percutaneously and<br>are intended to establish internal drainage from the ureteropelvic<br>junction to the bladder while maintaining external access to the<br>stent, as well as providing external drainage.<br><br>Expel™ Ureteral Stent System: |
The Expel Ureteral Stent System is delivered percutaneously and is intended to establish drainage from the ureteropelvic junction to the bladder and stenting of the ureter for all patients in whom it is desirable to place a drain which does not extend externally.
{5}------------------------------------------------
Device Description The Expel Nephroureteral and Ureteral Stents are single-lumen, nonvascular intervention stents inserted using percutaneous access to provide internal and/or external drainage of the ureteropelvic junction to the bladder and stenting of the ureter. They are long-term indwelling devices not to exceed 30 days in the body.
> Expel Nephroureteral Stents are available in outer diameters of 8.3 and 10.3 French, with working lengths ranging from 22 to 28 cm. They contain two pigtails each containing drainage holes. The locking, proximal pigtail forms in the renal pelvis, while the distal, concentric pigtail forms in the bladder. Drainage can occur internally (from the ureteropelvic junction to the bladder) or externally (from the bladder and ureteropelvic junction to the outside of the patient). Each piqtail has an adjacent radiopaque marker band. The distal end of the stent has a tapered tip and hydrophilic coating. The locking, proximal pigtail is activated by the Twist-Loc™ hub on the proximal end of the stent.
Expel Ureteral Stents are available in outer diameters of 6.3, 8.3, and 10.3 French, with working lengths ranging from 12 to 28 cm. The distal and proximal ends of the stent contain drainage holes within concentric pigtails. The pigtails form in opposite directions, with the proximal pigtail forming in the renal pelvis, and the distal pigtail forming in the bladder. Drainage occurs through pressure differential via gravity between the body cavities.
Each pigtail has an adjacent radiopaque marker band, and the distal end of the stent has a tapered tip and hydrophilic coating. A suture is looped through the most proximal drainage hole which allows for stent adjustment or removal during placement, and the suture is removed once placement is complete.
The Expel Nephroureteral and Ureteral Stents come with the following accessories that aid device placement.
- . Cannulas
- Stabilizer .
- Plug/Cap ●
- Pigtail Straightener .
The Expel Nephroureteral and Ureteral Stents are similar in fundamental design, function, device materials, sterilization, operating principle, intended use / indication for use and fundamental technology as the predicate device, the Percuflex™ Nephroureteral Stent and Percuflex™ Ureteral Stent. The modifications from the predicate device include:
- . Modified hub designs
- . New stent shaft material
- . Addition of radiopaque marker bands
- . Extended sizes/platforms
- Modified packaging ●
## Comparison of Technological Characteristics
{6}------------------------------------------------
- Performance Data The following bench testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Expel Nephroureteral and Ureteral Stents, including packaging, met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.
- · Stent Dimensions OD and Working Length
- Stent Shaft Tensile Strength
- · RO Marker Band Tensile Strength
- Hub to Shaft Tensile Strength
- · Tip Tensile Strength
- Distal Tip Robustness
- · Pigtail Recovery
- · Pigtail Retention/Removal Force
- Resistance to Deformation
- Kink Resistance
- Flow Recovery Post Kinking
- · Resistance to Liguid Leakage -Under Pressure and During Aspiration/ Vacuum
- · Stent / RO Marker Radiopacity
- MRI Compatibility
- · Urine Compatibility
- Cannula / Stent Connection Force
- · Cannula to Stent / Stabilizer Compatibility
- · Guidewire to Stent / Cannula Compatibility
- · Flexible Stiffening Cannula Hub to Shaft Tensile Strength
- · Stabilizer to Flexible Stiffening Cannula Connection Force
- · Coefficient of Friction
- · Sterile Barrier Integrity
- · Pouch Seal Strength
- Thermoformed Tray Visual
- Shelf Life
- · Sterilization
- · Biocompatibility
## Conclusion
Boston Scientific has demonstrated that the modifications made for the Expel Nephroureteral and Ureteral Stents are substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principle, and intended use / indication for use as the predicate device. Percuflex™ Nephroureteral Stent and Percuflex™ Ureteral Stent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.