Triopsy Actuator (TMSDGB); Triopsy Biopsy Needle (BN-1825-36-01); Triopsy Biopsy Needle (BN-1825-55-01)

K242228 · Triopsy Medical, Inc. · KNW · Jan 10, 2025 · Gastroenterology, Urology

Device Facts

Record IDK242228
Device NameTriopsy Actuator (TMSDGB); Triopsy Biopsy Needle (BN-1825-36-01); Triopsy Biopsy Needle (BN-1825-55-01)
ApplicantTriopsy Medical, Inc.
Product CodeKNW · Gastroenterology, Urology
Decision DateJan 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Triopsy Biopsy System (instrument and needles) is intended for use in obtaining biopsies from the prostate.

Device Story

System comprises reusable actuator and disposable biopsy needle; used by clinicians for prostate tissue sampling. Actuator provides variable penetration depth settings; needle features patented ridges to collect full-length, unbroken apex-to-base core samples. Device operation involves manual activation of spring-loaded mechanism to penetrate tissue and extract core. Output is physical tissue sample for pathological analysis. Benefits include increased accuracy in sample length and improved patient safety by preventing over/undershooting during biopsy procedure.

Clinical Evidence

Bench testing only. Performance verified per FDA guidance for biopsy devices, including depth projection, mechanical durability, penetration, spring activation force, and sample extraction. Human factors and usability engineering studies conducted with intended user population in simulated clinical conditions validated intended use.

Technological Characteristics

System consists of a reusable mechanical actuator and disposable biopsy needle. Actuator utilizes spring-loaded energy for needle penetration. Features variable penetration depth settings and specialized needle geometry (ridges) for sample integrity. No software or electronic components described.

Indications for Use

Indicated for obtaining prostate core biopsy samples in patients requiring prostate tissue diagnosis. Prescription use only.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". January 10, 2025 Triopsy Medical, Inc. % John Mann Chief Executive Officer Insight Navigation LLC 501 N Ford Street Golden, Colorado 80403 Re: K242228 Trade/Device Name: Triopsy Actuator (TMSDGB); Triopsy Biopsv Needle (BN-1825-36-01): Triopsy Biopsy Needle (BN-1825-55-01) Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: December 9, 2024 Received: December 9, 2024 Dear John Mann: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark J. Antonino -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Submission Number (if known) K242228 Device Name Triopsy Actuator (TMSDGB); Triopsy Biopsy Needle (BN-1825-36-01); Triopsy Biopsy Needle (BN-1825-55-01) Indications for Use (Describe) The Triopsy Biopsy System (instrument and needles) is intended for use in obtaining biopsies from the prostate. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ | 510(k) #: | K242228 | |-----------|---------| |-----------|---------| | | K242228 Page 1 of 2 | |--|-------------------------| | | Prepared on: 2024-12-07 | | Contact Details | 21 CFR 807.92(a)(1) | |-----------------|---------------------| |-----------------|---------------------| | Applicant Name | Triopsy Medical, Inc. | |---------------------------------|----------------------------------------------------| | Applicant Address | 151 St. Andrews Ct. Mankato MN 56001 United States | | Applicant Contact Telephone | (804) 677-8527 | | Applicant Contact | Dr. David Bostwick | | Applicant Contact Email | dbostwick@triopsy.com | | Correspondent Name | Insight Navigation LLC | | Correspondent Address | 501 N Ford St Golden CO 80403 United States | | Correspondent Contact Telephone | 7208082416 | | Correspondent Contact | Mr. John Mann | | Correspondent Contact Email | john@insightnav.com | | Device Name | 21 CFR 807.92(a)(2) | |-------------|---------------------| |-------------|---------------------| | Device Trade Name | Triopsy Actuator (TMSDGB);<br>Triopsy Biopsy Needle (BN-1825-36-01);<br>Triopsy Biopsy Needle (BN-1825-55-01) | |---------------------|---------------------------------------------------------------------------------------------------------------| | Common Name | Gastroenterology-urology biopsy instrument | | Classification Name | Instrument, Biopsy | | Regulation Number | 876.1075 | | Product Code(s) | KNW, FCG | | Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | |------------------------------------|---------------------| |------------------------------------|---------------------| | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | |-------------|----------------------------------------------------------|--------------| | K934370 | Magnum Reusable Core Biopsy Instrument | KNW | | K934371 | Magnum Needle | FCG | | Device Description Summary | 21 CFR 807.92(a)(4) | |----------------------------|---------------------| |----------------------------|---------------------| The Triopsy Medical Actuator and Biopsy Needle is comprehensive system that aids in the diagnosis and treatment planning of prostate cancer. The system utilizes an actuator instrument and biopsy needle (disposable) with the intention of gathering prostate tissue and properly identifying its pathology. The needle collects full-length, unbroken apex-to-base core samples with patented ridges that holdproperly identifying its pathology. The needle collects full-length, unbroken apex-to-base core samples with patented ridges that hold and protect the sample's integrity. The actuator allows for variable sample lengths for each area of the prostate to be biopsied increasing accuracy and patient safety by avoiding under or overshooting. {5}------------------------------------------------ # Intended Use/Indications for Use The Triopsy Biopsy System (instrument and needles) is intended for use in obtaining biopsies from the prostate. ## Indications for Use Comparison The subject and the predicate device have similar indications for use statements and intended use for acquisition of core biopsy samples, device is prescription only. ## Technological Comparison There are no differences between the indications for use of the predicate, Bard Magnum Reusable Core Biopsy Instrument & Bard Magnum Disposable Biopsy Core Needle (K934370 & K934371) and the subject device The Triopsy™ Reusable Biopsy" Single-Use Biopsy Needle. The primary differences between the predicate Bard Magnum Reusable Core Biopsy Instrument & Bard Magnum Disposable Biopsy Core Needle (K934370 & K934371) and the subject device The Triopsy™ Reusable Biopsy " Single-Use Biopsy Needle are the variable penetration depth setting of the Triopsy™ Actuator and the sample length notch of the Triopsy™ Needles. Based on substantial equivalence testing neither of these differences raise new questions of safety or efficacy. #### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807,92(b) The performance requirements of the device were verified through performance with the intended use of the device and in accordance with the FDA Guidance Content of Premarket Notifications for Biopsy Devices Used in Gastroenterology and Urology (February 1993) including: - Depth Projection - - Mechanical Durability - - -Penetration - -Activation Force (Spring) - -Sample Extraction In Addition, Following FDA Guidance Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016), the human factors studies were conducted with the intended user population, use scenarios to simulate clinical conditions. Results of the human factors testing demonstrate validation of the intended use. In addition, ### Not Applicable The Triopsy™ Reusable Biopsy Actuator & Triopsy™ Single-Use Biopsy Needle was verified and validated in accordance with 21 CFR §820.30. Testing was completed to demonstrate substantial equivalence and that any technological differences do not raise new or different questions of safety and effectiveness. The device succesfully passed all of the results demonstrate the device is safe, effective, and performs as well or better than the predicate device. The differences between the subject device do not introduce or raise concerns regarding the safe and effective use of the subject device. The subject device, Triopsy™ Reusable Biopsy Mingle-Use Biopsy Needle, is substantially equivalent to the legally marketed predicate device, Bard Magnum Reusable Core Biopsy Instrument & Bard Magnum Disposable Biopsy Core Needle (1934370 & K934371) with respect to the intended use/indications for use, target populations, treatment method, and technological characteristics. 21 CFR 807.92(a)(5) ### 21 CFR 807.92(a)(6) 21 CFR 807.92(a)(5) 21 CFR 807.92(a)(5)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...