Data from 30+ unique clinical sites, with 27 new sites compared to the predicate device, representing 61% of the cumulative data.
Not specified
Indications for Use
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance. The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
Device Story
Software-based computer-aided detection (CADe) system; analyzes high-definition endoscopic video in real time. Inputs: SDI video feed from compatible endoscopes (Olympus EVIS EXERA III, FUJI EC760). Processing: AI-based algorithm detects potential colorectal polyps; generates blue rectangular bounding box on monitor. Features: Device status indicator (green 'S' for active, gray 'S' for tool presence, red 'X' for error); mode selection button for bypass. Used in clinical settings by gastroenterologists. Output: Visual markers on endoscopic display. Benefit: Assists clinicians in identifying polyps during screening/surveillance. Safety: Detection marker disabled when biopsy tools detected or lighting inadequate; bypass mode available for software errors.
Clinical Evidence
Bench testing only. Software verification and validation conducted per FDA guidance. Bench testing confirmed adherence to 21 CFR 876.1520 special controls regarding algorithm performance, pixel degradation, and video latency. No new clinical data; clinical performance relies on data from K213686.
Technological Characteristics
Software-based CADe system; operates on hospital mains power. Inputs: SDI video feed. Outputs: GUI-based visual markers (blue rectangles). Connectivity: Standalone system. Software: AI-based detection algorithm; OS and library updates. Safety: Automated tool detection and bypass mode. Compatible with Olympus EVIS EXERA III and FUJI EC760 series endoscopes.
Indications for Use
Indicated for adult patients undergoing colorectal cancer screening or surveillance colonoscopy. Intended for use by qualified and trained gastroenterologists during white light colonoscopy to assist in identifying potential colorectal polyps.
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out to the right of the square.
July 3, 2024
Iterative Scopes, Inc. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market St. Floor 23 Philadelphia, Pennsylvania 19103
Re: K241508
Trade/Device Name: SKOUT® system Regulation Number: 21 CFR 876.1520 Regulation Name: Gastrointestinal Lesion Software Detection System Regulatory Class: Class II Product Code: QNP Dated: May 28, 2024 Received: May 28, 2024
# Dear Janice Hogan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241508
Device Name
SKOUT® system
#### Indications for Use (Describe)
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(K) SUMMARY
# SKOUT® system
# K241508
### Submitter Contact Information:
Address: Iterative Health, Inc. (Iterative Scopes) 14 Arrow Street Floor 3 Cambridge. MA 02138
Phone: (617) 209-9773
Caitlyn Seidl, VP, Clinical, Regulatory and Medical Affairs Contact:
Date Prepared: May 27, 2024
Name of Device: SKOUT® system
Classification Name: Gastrointestinal lesion software detection system (21 CFR 876.1520, Product Code QNP)
Predicate Device: SKOUT® system, Iterative Scopes, K240781
### Device Description
The SKOUT system is a software-based computer aided detection (CADe) system for the analysis of high-definition endoscopic video during colonoscopy procedures. The SKOUT system is intended to aid gastroenterologists with the detection of potential colorectal polyps during colonoscopy by providing an informational visual aid on the endoscopic monitor using trained software that processes the endoscopic video in real time.
Users will primarily interact with the SKOUT system by observing the software display, including the polyp detection box and device status indicator signal.
#### Polyp Detection Notification
The SKOUT system has a main graphical user interface (GUI) feature of the polyp detection notification. The polyp detection is a two-dimensional blue rectangular outline generated around any suspected polyps on the endoscopic video feed. If there is no polyp detected, the bounding box does not appear. SKOUT system pauses polyp detection when an endoscopic tools are detected in the video feed to ensure that the bounding box does not hinder any surgical procedure, biopsy, or resection or this may also occur when lighting conditions are deemed to be inadequate.
The polyp detection notification enables users to:
- . Detect potential colorectal polyps during colonoscopy examinations in adult patients undergoing a colorectal cancer screening or surveillance procedure.
- Utilize a tool that provides additional information for endoscopic observation. ●
{5}------------------------------------------------
# Device Status Indicator
The SKOUT system has an additional GUI feature that notifies users of the current device status (active or error):
- a two-dimensional green box with letter (S) when the device is powered on and actively ● processing video.
- . a two-dimensional gray box with letter (S) when a surgical tool is present.
- a red (X) with an error message; when there is an error with the video processing function of the SKOUT system, the green box will be replaced with a red X and error message to indicate an error has occurred.
# Intended Use / Indications for Use
The SKOUT® system is a software device designed to detect potential colorectal polyps in real time during colonoscopy examinations. It is indicated as a computer-aided detection tool providing colorectal polyps location information to assist qualified and trained gastroenterologists in identifying potential colorectal polyps during colonoscopy examinations in adult patients undergoing colorectal cancer screening or surveillance.
The SKOUT® system is only intended to assist the gastroenterologist in identifying suspected colorectal polyps and the gastroenterologist is responsible for reviewing SKOUT® suspected polyp areas and confirming the presence of a polyp based on their own medical judgment. SKOUT® is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. SKOUT® is indicated for white light colonoscopy only.
# Summary of Technological Characteristics
The subject device is a modified version of the predicate device, which was FDA-cleared on April 19, 2024 (K240781). The key characteristics of the subject SKOUT system and the predicate SKOUT system are compared in the following table.
| | Subject Device:<br>SKOUT® system (K241508) | Predicate Device:<br>SKOUT® system (K240781) | Comparison |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | QNP | QNP | Same |
| Regulation | 21 CFR 876.1520 | 21 CFR 876.1520 | Same |
| Device Class | Class II | Class II | Same |
| Indications for<br>Use | The SKOUT® system is a<br>software device designed to<br>detect potential colorectal polyps<br>in real time during colonoscopy<br>examinations. It is indicated as a<br>computer-aided detection tool<br>providing colorectal polyps<br>location information to assist<br>qualified and trained | The SKOUT® system is a<br>software device designed to<br>detect potential colorectal polyps<br>in real time during colonoscopy<br>examinations. It is indicated as a<br>computer-aided detection tool<br>providing colorectal polyps<br>location information to assist<br>qualified and trained | Same |
| | Subject Device:<br>SKOUT® system (K241508) | Predicate Device:<br>SKOUT® system (K240781) | Comparison |
| | gastroenterologists in identifying<br>potential colorectal polyps during<br>colonoscopy examinations in<br>adult patients undergoing<br>colorectal cancer screening or<br>surveillance. | gastroenterologists in identifying<br>potential colorectal polyps during<br>colonoscopy examinations in<br>adult patients undergoing<br>colorectal cancer screening or<br>surveillance. | |
| | The SKOUT® system is only<br>intended to assist the<br>gastroenterologist in identifying<br>suspected colorectal polyps and<br>the gastroenterologist is<br>responsible for reviewing<br>SKOUT® suspected polyp areas<br>and confirming the presence or<br>absence of a polyp based on<br>their own medical judgment.<br>SKOUT® is not intended to<br>replace a full patient evaluation,<br>nor is it intended to be relied<br>upon to make a primary<br>interpretation of endoscopic<br>procedures, medical diagnosis,<br>or recommendations of<br>treatment/course of action for<br>patients. SKOUT® is indicated<br>for white light colonoscopy only. | The SKOUT® system is only<br>intended to assist the<br>gastroenterologist in identifying<br>suspected colorectal polyps and<br>the gastroenterologist is<br>responsible for reviewing<br>SKOUT® suspected polyp areas<br>and confirming the presence or<br>absence of a polyp based on<br>their own medical judgment.<br>SKOUT® is not intended to<br>replace a full patient evaluation,<br>nor is it intended to be relied<br>upon to make a primary<br>interpretation of endoscopic<br>procedures, medical diagnosis,<br>or recommendations of<br>treatment/course of action for<br>patients. SKOUT® is indicated<br>for white light colonoscopy only. | |
| User<br>Population | Adult patients undergoing<br>colorectal cancer screening or<br>surveillance colonoscopy. | Adult patients undergoing<br>colorectal cancer screening or<br>surveillance colonoscopy. | Same |
| Fundamental<br>Technological<br>Characteristics | The SKOUT system is<br>composed of a single piece<br>hardware design and software<br>designed to highlight portions of<br>the colon where the device<br>detects potential colorectal<br>polyps. | The SKOUT system is<br>composed of a single piece<br>hardware design and software<br>designed to highlight portions of<br>the colon where the device<br>detects potential colorectal<br>polyps. | Same |
| Software<br>Algorithm | The SKOUT system utilizes an<br>artificial intelligence-based<br>algorithm to perform the polyp<br>detection function. | The SKOUT system utilizes an<br>artificial intelligence-based<br>algorithm to perform the polyp<br>detection function. | Same |
| Power Source | Hospital mains power | Hospital mains power | Same |
| | Subject Device:<br>SKOUT® system (K241508) | Predicate Device:<br>SKOUT® system (K240781) | Comparison |
| Safety<br>Features | The Mode Selection Button<br>allows for instantaneous toggling<br>between the SKOUT video feed<br>and the bypass video feed in the<br>event of software error that<br>affects video quality.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention.<br><br>SKOUT system GUI also has a<br>device status indicator that<br>notifies users of the current<br>device status (active or error):<br>a two-dimensional green<br>box with letter (S) when<br>the device is powered on<br>and actively processing<br>video.a two-dimensional gray<br>box with letter (S) when<br>a surgical tool is<br>present.a red (X) with an error<br>message; when there is<br>an error with the video<br>processing function of<br>the SKOUT system, the<br>green box will be<br>replaced with a red X<br>and error message to<br>indicate an error has<br>occurred. | The Mode Selection Button<br>allows for instantaneous toggling<br>between the SKOUT video feed<br>and the bypass video feed in the<br>event of software error that<br>affects video quality.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent<br>obstruction of the area of interest<br>during intervention.<br><br>SKOUT system GUI also has a<br>device status indicator that<br>notifies users of the current<br>device status (active or error):<br>a two-dimensional green<br>box with letter (S) when<br>the device is powered on<br>and actively processing<br>video.a two-dimensional gray<br>box with letter (S) when<br>a surgical tool is<br>present.a red (X) with an error<br>message; when there is<br>an error with the video<br>processing function of<br>the SKOUT system, the<br>green box will be<br>replaced with a red X<br>and error message to<br>indicate an error has<br>occurred. | Same |
| Device Output | SKOUT system generates<br>markers in the form of blue<br>rectangles superimposed on the<br>endoscopic video when potential<br>colorectal polyps are identified.<br>SKOUT markers are not<br>accompanied by a sound.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent | SKOUT system generates<br>markers in the form of blue<br>rectangles superimposed on the<br>endoscopic video when potential<br>colorectal polyps are identified.<br>SKOUT markers are not<br>accompanied by a sound.<br><br>The polyp detection marker is<br>disabled if a biopsy tool enters<br>the field of view to prevent | Same |
| | Subject Device:<br>SKOUT® system (K241508) | Predicate Device:<br>SKOUT® system (K240781) | Comparison |
| | obstruction of the area of interest<br>during intervention. | obstruction of the area of interest<br>during intervention. | |
| Video Input | Serial Digital Interface (SDI); de-<br>interlacing applied during pre-<br>processing | Serial Digital Interface (SDI); de-<br>interlacing applied during pre-<br>processing | Same |
| Compatible<br>Endoscopes | Olympus EVIS EXERA III and<br>FUJI EC760 series | Olympus EVIS EXERA III and<br>FUJI EC760 series | Same |
# Table 1: Comparison of Key Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The subject device includes the following minor changes to the technological characteristics compared to the predicate device:
- Refinement/retraining of polyp detection algorithm, including:
- Utilization of same system level and algorithm level architectures as predicate device O
- Utilization of data from 30+ unique clinical sites, of which 27 were new compared to o the predicate device, with new data representing 61% of the cumulative data
- Refinement of polyp tracking algorithm ●
- Update to current version of system operating system ●
- Update to current version of software libraries ●
- Bug fixes
### Performance Testing
Non-clinical performance testing was conducted to demonstrate that the SKOUT system is as safe and effective as the predicate device.
- Software verification and validation testing was conducted to confirm the SKOUT system ● software meets design requirements for its intended use, per the device's design documentation in line with recommendations outlined in General Principles of Software Validation, Guidance for Industry and FDA Staff.
- Additional bench software testing was performed to confirm the device meets the special ● controls in 21 CFR 876.1520 for algorithm performance, pixel degradation and video delays.
SKOUT system demonstrated passing results in all applicable testing.
#### Conclusions
The SKOUT system has the same intended use, indications for use, technological characteristics, and principles of operation as its predicate device. The minor differences in software do not affect its safety and effectiveness when used as labeled. The inference algorithms the same architecture and meet the same performance requirements as the predicate device, therefore clinical performance remains unchanged from the clinical performance submitted in K213686. Performance data demonstrates that the SKOUT system is as safe and effective as the predicate device. Thus, the SKOUT system is substantially equivalent.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.