QIS · Esophageal, Mucosal, Electrical Characterization

Gastroenterology, Urology · 21 CFR 876.1450 · Class 2

Overview

Product CodeQIS
Device NameEsophageal, Mucosal, Electrical Characterization
Regulation21 CFR 876.1450
Device ClassClass 2
Review PanelGastroenterology, Urology

Identification

An esophageal tissue characterization system is a device intended for obtaining measurements of electrical properties within esophageal tissue.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) All patient contacting components of the device must be demonstrated to be biocompatible. (2) Performance testing must demonstrate the device can accurately measure the designated electrical characteristics. (3) Mechanical safety testing must demonstrate that the device will withstand forces encountered during use. (4) Software verification, validation, and hazard analysis must be performed. (5) Electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety of the device must be performed. (6) Performance data must validate the reprocessing instructions for any reusable components of the device. (7) Labeling must include: (i) Specific instructions regarding the proper placement and use of the device; (ii) Instructions for reprocessing of any reusable components; and (iii) An expiration date for single use components.

Recent Cleared Devices (3 of 3)

RecordDevice NameApplicantDecision DateDecision
K233974Mucosal Impedance Measurement SystemAlandra Medical Sapi DE CVSep 6, 2024SESE
K230056MiVu™ Esophageal Endo CapDiversatek HealthcareApr 25, 2023SESE
DEN180067Mucosal Integrity Conductivity (MI) Test SystemDiversatek Healthcare, Inc.Dec 23, 2019DENG

Top Applicants

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...