MEGACHANNEL ENDOSCOPIC OVERTUBE, MODEL OT22

K080550 · Minos Medical · ODB · Jun 10, 2008 · Gastroenterology, Urology

Device Facts

Record IDK080550
Device NameMEGACHANNEL ENDOSCOPIC OVERTUBE, MODEL OT22
ApplicantMinos Medical
Product CodeODB · Gastroenterology, Urology
Decision DateJun 10, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

To be used with endoscopes as a channel for multiple insertions of the endoscope into the lower gastrointestinal tract. The Megachannel Endoscopic Overtube is indicated for use in conjunction with an endoscope for tissue or foreign body manipulations and/or where multiple removal and reinsertions of the endoscope are required.

Device Story

Megachannel Endoscopic Overtube is a disposable flexible PVC tube used with endoscopes; facilitates multiple insertions of endoscope into lower gastrointestinal tract. Device includes proximal handle with insufflation sealing cap accommodating 12.8 mm diameter colonoscope; removable introducer plug at distal tip aids introduction through GI tract. Used in clinical settings by physicians; provides protected channel for endoscope access; enables tissue or foreign body manipulation; allows repeated endoscope removal/reinsertion without re-traversing tract. Benefits patient by reducing trauma during multiple endoscope passes.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Disposable flexible PVC tube; includes proximal handle with insufflation sealing cap and removable distal introducer plug. Accommodates 12.8 mm diameter colonoscopes. Non-powered, mechanical device.

Indications for Use

Indicated for use in conjunction with an endoscope for tissue or foreign body manipulations and/or where multiple removal and reinsertions of the endoscope are required in the lower gastrointestinal tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K080550 ## Minos Medical 15560-C Rockfield Boulevard ● Irvine, CA 92618 Phone: 949-215-1835 |Fax: 949-625-8625 ## 5 510(k) Summary | Submitter: | Minos Medical<br>15560-C Rockfield Boulevard<br>Irvine, CA 928618 | JUN 10 2008 | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Contact Person: | Tom Colonna,<br>Director of Quality | | | Date Prepared: | February 27, 2008 | | | Trade name: | Megachannel Endoscopic Overtube | | | Classification<br>Name: | Endoscope and Accessories (21 CFR 876.1500) | | | Name:<br>Predicate<br>Devices: | US Endoscopy Guardus Overtube (K040836) | | | | Spirus Medical Endo-Ease Overtube (K0520840) | | | Device<br>Description: | The Megachannel Endoscopic Overtube is a disposable<br>flexible PVC tube that is to be used with an endoscope.<br>The overtube includes a proximal handle with<br>insufflation sealing cap that accommodates a standard<br>12.8 mm diameter colonoscope. A removable<br>introducer plug is attached to the distal tip to facilitate<br>the introduction of the overtube through the<br>gastrointestinal tract. | | | Intended Use: | To be used with endoscopes as a channel for multiple<br>insertions of the endoscope into the lower<br>gastrointestinal tract. | | | Comparative<br>Analysis: | The Megachannel Endoscopic Overtube has been<br>demonstrated to be as safe and effective as the<br>predicate devices for its intended use. | | | Functional/Safety<br>Testing: | The Megachannel Endoscopic Overtube has<br>successfully undergone functional testing. These<br>products have been shown to be equivalent to the<br>predicate devices. | | | Conclusion: | The Megachannel Endoscopic Overtube is substantially<br>equivalent to the predicate devices. | | : {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three bars representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 0 2008 Mr. Tom Colonna Director of Quality Minos Medical 15560 -C Rockfield Blvd. IRVINE CA 92618 Re: K080550 Trade/Device Name: Megachannel Endoscopic Overtube Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODB Dated: May 2, 2008 Received: May 5, 2008 Dear Mr. Colonna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR. 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Nancy Cbrogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K080550 Minos Medical 15560-C Rockfield Boulevard · Irvine, CA 92618 Phone: 949-215-1835 Fax: 949-625-8625 ## Indications for Use Statement 4 510(k) Number (if known): (if known): K080550 Device Name: Megachannel Endoscopic Overtube Indications for Use: The Megachannel Endoscopic Overtube is indicated for use in conjunction with an endoscope for tissue or foreign body manipulations and/or where multiple removal and reinsertions of the endoscope are required. Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Helmut Renn (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _ Page 1 of _1 February 27, 2008 510(k) Notification for Megachannel Endoscopic Overtube CONFIDENTIAL .
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...