This instrument is indicated for use with a compatible trumpet valve during endoscopic surgery for facilitating the removal of gallstones.
Device Story
Stone extractor is a manual surgical instrument used during endoscopic surgery to facilitate gallstone removal. Device consists of a 13.0" tube with a 0.6" scoop at the distal end and a nominal diameter of 0.392". Fabricated from 304 stainless steel, the device features a polished proximal section and a matte-finished distal section. Operated by a surgeon during endoscopic procedures, the device is inserted through a cannula to capture and remove stones. The device is reusable and sold non-sterile, requiring sterilization by the user. It is designed to be used in conjunction with a compatible trumpet valve. Proper handling is required during insertion and removal through hinged valves to prevent mechanical damage or snagging.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Material: 304 stainless steel. Dimensions: 13.0" length, 0.392" diameter, 0.6" scoop. Surface finish: polished proximal, matte distal. Reusable. Sold non-sterile. Mechanical device; no energy source or software.
Indications for Use
Indicated for use during endoscopic surgery for gallstone removal. Contraindicated when physician judgment deems use contrary to patient interest.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
American Hydro-Surgical Instruments, Inc. stone extractor (no 510(k) number provided)
Related Devices
K230598 — Stone Extraction Baskets · Jiangsu Vedkang Medical Science and Technology Co., Ltd. · Sep 12, 2023
K171969 — Web II Memory Extraction Basket or Memory II Double Lumen Extraction Basket, Memory 5 Fr. Soft Wire Baskets or Memory Helical Stone Extractor, Memory Eight Wire Baskets or Memory Hard Wire Baskets, Fusion Wire Guided Extraction Basket, Non-Lithotripsy Extraction Basket · Wilson-Cook Medical, Inc. · Mar 28, 2018
K071066 — SPYCATCH STONE RETRIEVAL BASKET · Boston Scientific Corp · Jun 29, 2007
K243807 — Single Use Retrieval Basket V FG-V421PR/V422PR/V431P/V432P · Olympus Medical Systems Corp. · Mar 18, 2025
K170811 — Single Use Retrieval Nitinol Basket V · Olympus Medical Systems Corp. · Nov 27, 2017
Submission Summary (Full Text)
{0}
PHX Technologies Corporation
K971123
Post Office Box 1059, Lewisville, Texas 75067 (817)387-5696 FAX (817)382-0577
# 510(K) Summary
Date: 21 March 1997
JUN 27 1997
Trade Name: Stone Extractor
Common Name: Stone Extractor
Classification Name: Unknown
Device Description: The stone extractor has a nominal diameter of 0.392", a tube length of 13.0" and has a scoop 0.6" long. The surface of the stone extractor is polished on the proximal 3.0" and matte finished on the remaining 10.0". The stone extractor is fabricated from 304 stainless steel.
Intended Use: This stone extractor is indicated for use in laproscopic surgery for the removal of gall stones.
Substantial Equivalence: This stone extractor, Submitted Device is, substantially equivalent to the stone extractor, Predicate Device, currently being sold in the United States by American Hydro-Surgical Instruments, Inc., 430 Commerce Drive, Delray, Florida 33445 To the best of our knowledge, this predicate devices is being "legally" marketed.
Comparison to Predicate Device:
| Attribute | Predicate Device | Submitted Device |
| --- | --- | --- |
| Material of construction, Stone extractor | Stainless steel | Stainless steel |
| Material of construction, handle | Aluminum | Stainless steel |
| Stone extractor, nominal diameter, inches | 0.392" | 0.392 |
| Stone extractor, nominal length, inches | 13.0" | 13.0" |
| Stone extractor, nominal length, Scoop, inches | 0.6" | 0.6" |
| Reuseability | Reuseable | Reuseable |
| Sterility | Sold non-sterile | Sold non-sterile |
Submitter's Name: PHX Technologies Corporation
Submitter's Address: 1032 Shady Oaks Drive, No. 100, Denton, TX 76205
Submitter's Phone #: (817) 387-5696
Submitter's FAX: (817) 382-0577
Submitter's Contact Person: James F. Chapel
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 27 1997
Mr. James F. Chapel
President
PHX Technologies Corporation
P.O. Box 1059
Lewisville, Texas 75067
Re: K971123
Stone Extractor
Dated: May 13, 1997
Received: May 19, 1997
Regulatory class: II
21 CFR §876.1500/Product code: 78 KOG
Dear Mr. Chapel:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
Page 1 of 1
510(k) Number (if known): K971123
Device Name: Billiary Stone Extractor
Indications For Use:
Indications: This instrument is indicated for use with a compatible trumpet valve during endoscopic surgery for facilitating the removal of gallstones.
Contraindications: This instrument is contraindicated for use when, in the judgment of the physician, it's use would be contrary to the best interest of the patient.
Precautions: During surgery this instrument must be handled with care to avoid misuse and possible damage. Use care when passing this instrument through a cannula. When inserting the instrument into a cannula that has a hinged valve, make sure that the valve is fully open. When removing the instrument from a cannula that has a hinged valve, make sure that the valve is fully open and pull the instrument straight out, as lateral or side movement may cause the tip to hang on the valve.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K971123
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.