CLARUS MODEL 2127-900 SERIES ENDOSCOPE ADATER (2127-900 SERIES)
K971584 · Clarus Medical Systems, Inc. · OCV · May 23, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K971584
Device Name
CLARUS MODEL 2127-900 SERIES ENDOSCOPE ADATER (2127-900 SERIES)
Applicant
Clarus Medical Systems, Inc.
Product Code
OCV · Gastroenterology, Urology
Decision Date
May 23, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Model 2127-900 Series Endoscope Adapter is an accessory for the Clarus Medical Systems Model 2127 Endoscope. It is intended to be reusable. It is used to temporarily clamp or hold the endoscope when mounting to an airway catheter.
Device Story
The Model 2127-900 Series Endoscope Adapter is a machined or molded plastic accessory designed to facilitate the temporary mounting of a Clarus Model 2127 Endoscope to an airway catheter. The device features a tapered hub that fits into the proximal end of an airway catheter and a central aperture sized to accommodate the endoscope shaft. It is used by clinicians to stabilize the endoscope during airway procedures. The adapter is reusable and compatible with standard disinfecting solutions, such as 2% glutaraldehyde.
Clinical Evidence
Bench testing only. Biocompatibility testing of the plastic material was performed and passed. Physical testing included exposure to 2% glutaraldehyde disinfecting solution, dimensional measurements, and workmanship examination.
Technological Characteristics
Machined or molded plastic adapter; tapered hub design; central aperture for endoscope shaft; reusable; compatible with 2% glutaraldehyde disinfection.
Indications for Use
Indicated for use as an accessory to the Clarus Model 2127 Endoscope to temporarily clamp or hold the endoscope when mounting to an airway catheter.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Reference Devices
Clarus Model 2127 Endoscope
Related Devices
K241842 — Luer-Split MAJ-2092 · Olympus Medical Systems Corporation · Mar 19, 2025
K082038 — CLARUS VIDEO AIRWAY ENDOSCOPE, MODELS 30000-V AND 30003-V · Clarus Medical, LLC · Apr 28, 2009
K251986 — AUXILIARY WATER TUBE MAJ-855 · Olympus Medical Systems Corporation · Sep 26, 2025
K132168 — BRONCHOSCOPY ELBOW · Respironics, Inc. · Nov 20, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 2 3 1997
A 4711585
tg 1 of 2
#### 510(k) Summary
## Model 2127-900 Series Endoscope Adapter
### General Information
| Classifications | Class II Accessory |
|-----------------|----------------------------------------------------------------------------------------|
| Trade Name | Clarus Model 2127-900 Series Endoscope Adapter |
| Submitted | Clarus Medical Systems, Inc.<br>1000 Boone Avenue North, #100<br>Minneapolis, MN 55427 |
| Contact | Dale Sappenfield<br>Vice President Engineering and Quality |
#### Device Description
The Model 2127-900 Series Endoscope Adapter is a machined or molded plastic piece used to hold an endoscope and airway catheter together. The endoscope adapter has a machined or molded tapered hub designed to fit firmly into the proximal end of various airway catheters. Additionally the endoscope adapter has a central hole designed to fit the outside diameter of the shaft on the appropriate legally marketed Clarus Model 2127 Endoscope.
#### Intended Use
The Model 2127-900 Series Endoscope Adapter is an accessory for the Clarus Medical Systems Model 2127 Endoscope. It is intended to be reusable. It is used to temporarily clamp or hold the endoscope when mounting to an airway catheter.
#### Testing
Biocompatibility testing was done on the material used in the endoscope adapter. The material passed biocompatibility testing and is suitable for this application.
Physical testing included disinfecting solution (2% glutaraldehyde) exposure testing, examination for workmanship, dimensional measurements and certification of material.
{1}------------------------------------------------
ATTI=
Pg 20
# Summary of Substantial Equivalence
The Model 2127-900 Series Endoscope Adapter is an accessory to a Model 2127 Clarus Endoscope. When used with the appropriate endoscope, the endoscope adapter augments the capabilities of the endoscope by acting as a clamping and holding devise to temporarily mount the endoscope to and airway catheter.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a person with outstretched arms, representing care and protection.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 3 1997
Mr. Dale Sappenfield Vice President, Engineering & Quality Clarus Medical Systems, Inc. 1000 Boone Avenue North Minneapolis, Minnesota 55427
Re: K971584 Clarus Model #2127-900 Series Endoscope Adapter Dated: May 13, 1997 Received: May 16, 1997 Regulatory class: II 21 CFR §876.1500/Product code: 78 KOG
Dear Mr. Sappenfield:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat/ and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
Device Name: CLARDS HODEL 2127- 900 SERIES GODDECOPE
Indications For Use:
HOUNT AND 450 ENDOSCOPE ACCESSORY TEST PORARILY 00 USED AIRWAY CATHETER APPRODUATE TOGETHER ENDOSECHE ADD AN
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert R. Anthony
(Division Sign-Off) Devision of Reproductive, Abdominal, ENT, and Radiological Devices
x(k) Number k-97115-1
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches