K102452 · Minimally Invasive Devices, LLC · OCT · Dec 23, 2010 · Gastroenterology, Urology
Device Facts
Record ID
K102452
Device Name
DEFOGGING AND CLEANING SOLUTION- FLO-X
Applicant
Minimally Invasive Devices, LLC
Product Code
OCT · Gastroenterology, Urology
Decision Date
Dec 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Defogging and Cleaning Solution is a single-use laparoscopic accessory device intended to facilitate intra-operative defogging and cleaning of the laparoscope lens thereby maintaining visualization of the surgical site.
Device Story
Biocompatible surfactant solution (Docusate Sodium) used as laparoscopic accessory; prevents lens fogging and cleans debris/lipids from laparoscope lens intra-operatively. Applied via flushing (e.g., using FloShield Plus System) without mechanical wiping. Surfactant reduces surface tension, creating non-scattering water film (wetting) to prevent fogging. Used in OR by surgeons/staff during minimally invasive procedures. Provides clear visualization by removing debris that saline/water cannot clear. Benefits patient by reducing procedure interruptions for lens cleaning/defogging.
Clinical Evidence
No human clinical data. Evidence based on bench testing and porcine peritoneal cavity animal study. Bench testing: Syringe extractable study (ISO 10993-18); defogging efficacy comparison vs. predicate; cleaning efficacy comparison vs. water/saline (demonstrated superiority in grease removal). Animal study: Porcine model confirmed efficacy in situ, compatibility with FloShield Plus System, and lack of systemic toxicity (acute systemic toxicity study in mice). Irradiation stability (25-40 kGy) confirmed via HPLC/pH analysis.
Technological Characteristics
Biocompatible surfactant (Docusate Sodium) solution. Clear, colorless, water-soluble. Packaged in syringe or bottle. Sterilized via E-Beam irradiation (25-40 kGy). Functions via surface tension reduction (wetting).
Indications for Use
Indicated for use during minimally invasive surgery to defog and clean the laparoscope lens to maintain visualization of the surgical site.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K973558 — VISUAL ENDOSCOPIC NON-ALCOHOL ULTRACELL SPONGE (VENUS) · Ultracell Medical Technologies, Inc. · Nov 24, 1997
K152948 — Medline Anti-Fog Solution · Medline Industries, Inc. · Feb 4, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design resembling an abstract human figure or a symbol representing health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Minimally Invasive Devices, LLC % MM & A Consulting, LLC Mark L. Friedman, Ph.D. 6329 SW 81st Road Lake Butler, FL 32054
JUL 2 7 2015
Re: K102452
Trade/Device Name: Minimally Invasive Devices Defogging and Cleaning Solution Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCT Dated (Date on orig SE ltr): December 16, 2010 Received (Date on orig SE ltr): December 20, 2010
Dear Dr. Friedman,
This letter corrects our substantially equivalent letter of December 23, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetw/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEC 2 3 2010
# Indications for Use
510(k) Number (if known): _ K 102452
Device Name: Minimally Invasive Devices Defogging and Cleaning Solution
Indications for Use:
The Defogging and Cleaning Solution is a single-use laparoscopic accessory device intended to facilitate intra-operative defogging and cleaning of the laparoscope lens thereby maintaining visualization of the surgical site.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Page 1 of 1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rs. Rha 1 no 17
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k102412
{3}------------------------------------------------
Minimally Invasive Devices, Inc.
# 510(k) Summary
KIO 24.52
DEC 2 3 2010 Applicant Name Minimally Invasive Devices, Inc. and Address 1275 Kinnear Road Columbus, Ohio 43212 Contact Person: Caroline Crisafulli Date of Summary: December 16, 2010 Name of Device: Trade/Proprietary Name: Flo-X Common Name: Defogging and Cleaning Solution Establishment Name: Minimally Invasive Devices, Inc. 1275 Kinnear Road Columbus, Ohio 43212 3007362639 Establishment Number: Proposed Classification: KOG Class II 21 CFR Part 876.1500, Endoscope and/or Accessories. General and Plastic Surgery Panel: Predicate Devices:
| K944249 | Covidien Auto-Suture Anti-Fog Solution (Dexide FRED) |
|---------|------------------------------------------------------|
| K982465 | DeRoyal Industries, Inc. DeFogger |
| K080613 | Minimally Invasive Devices, Inc. FloShieldTM System |
#### Intended Use:
The Defogging and Cleaning Solution is a single-use laparoscopic accessory device intended to facilitate intra-operative defogging and cleaning of the laparoscope lens thereby maintaining visualization of the surgical site.
#### Device Description:
The Minimally Invasive Defogging and Cleaning Solution is a biocompatible surfactant (Docusate Sodium) when used as indicated. It functions in the same manner as the predicate devices. It is intended to be used to prevent fogging of the lens of devices (laparoscope) that provide visualization during minimally invasive surgeries. Additionally, the solution can be used to clean the end of the lens of devices (laparoscope) that provide visualization during minimally invasive surgeries (i.e. to rinse lens intra-operatively by rinsing any debris from the lens which becomes attached to the lens as a result of the procedure.)"
### Summary of Technological Characteristics:
The Defogging and Cleaning Solution is comprised of a surfactant that provides the same physical characteristics as the listed predicate devices. The Defogging and Cleaning Solution is a clear/colorless, water soluble solution that is packaged in a syringe or bottle for easy use. The
{4}------------------------------------------------
product is irradiated to ensure sterility. The defogging (anti-fogging) action of the Minimally Invasive Devices Defogging and Cleaning Solution works as a defogger in the same fashion of the predicate devices. Typically anti-fog treatments work by minimizing surface tension. resulting in a non-scattering film of water instead of single droplets, an effect called welting. Surfactants are one compound that are used in the predicate devices to "wet" the surface of the lens
#### Summary of Performance Testing:
Animal and bench studies (in simulated peritoneal cavity environments) have shown that the Minimally Invasive Devices Defogging and Cleaning Solution effectively washes debris and any oil, including lipid-based debris from the lens that saline or water will are unable to wash away. The cleaning action is a simple flush across the lens (no mechanical mechanism such as wiping is needed) when needed, the surfactant properties effectively rinse/wash the debris from the lens. The surfactant will then easily flow off the lens resulting in a refreshed and clear view of the operative site without interrupting the procedure. The testing performed included the following:
| Test Title | Summarized Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe Extractable Study | Testing in compliance with ISO 10993-18 was performed to demonstrate that<br>the container did not have any extractable compound that would leach into the<br>Defogging and Cleaning solution. |
| Evaluation of acute systemic toxicity of the<br>Defogging and Cleaning Solution following a<br>single intra-peritoneal injection into mice. | There was no mortality or evidence of systemic toxicity from the test article<br>injected into mice. |
| Verification that the Minimally Invasive<br>Devices Defogging and Cleaning Solution is<br>effective in defogging a laparoscope lens. | This testing demonstrated that the Defogging and Cleaning Solution was<br>effective as a defogging agent when the solution was applied to the<br>laparoscope lens with no mechanical action such as blotting or use of a<br>saturated sponge |
| Ability to Defog the Laparoscope Lens:<br>Comparison between the Minimally Invasive<br>Devices Defogging and Cleaning Solution<br>and a Predicate Device | The Defogging and Cleaning Solution and the predicate device both<br>demonstrated that they were effective and equivalent in defogging the<br>laparoscope lens. |
| Ability to Clean the Laparoscope Lens:<br>Comparison between the Minimally Invasive<br>Devices Defogging and Cleaning Solution,<br>Water and Saline | The results demonstrate that the Defogging and Cleaning Solution effectively<br>removed the grease from the laparoscope lens when applied with flushing via<br>the FloShield™ Plus System. In comparison with water and saline, the water<br>was the least effective at cleaning the lens, saline was slightly better, but the<br>Defogging and Cleaning Solution was superior to water and saline in clearing<br>grease from the lens. |
| Effects of Irradiation on the Defogging and<br>Cleaning Solution | Analysis of the Defogging and Cleaning Solution pre and post E-Beam<br>irradiation to 25-40 kiloGray (kGy) demonstrates that the solution is not<br>affected by irradiation. The Docusate component is stable and the solution<br>properties are unchanged. pH, concentration and analysis by HPLC were<br>used to make the assessments. |
| Animal Testing of the Defogging and | The Defogging and Cleaning Solution was effective in keeping the<br>laparoscope lens defogged when applied externally and was found to be<br>equivalent to the predicate devices in keeping the laparoscope lens defogged<br>when applied externally. |
| Cleaning Solution (Porcine Peritoneal Cavity) | The Minimally Invasive Device Defogging and Cleaning Solution did not<br>interfere with the function of the FloShield™ Plus System, worked in<br>conjunction with the FloShield™ Plus System and was effective in<br>removing/cleaning debris/blood/fat from the lens in situ when applied in 1-2 cc<br>aliquots through the FloShield™ Plus System. |
#### Conclusion:
The Minimally Invasive Devices Defogging and Cleaning Solution (Flo-X) is equivalent to the predicate devices in its ability to defog and keep a laparoscope lens defog during minimally invasive surgery and additionally aids in cleaning the laparoscope lens of debris that might otherwise accumulate on the lens during the course of the surgery. The Defogging and Cleaning Solution is at least as safe and effective as predicate devices.
2/2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.