MYE · System, Electrogastrography (Egg)

Gastroenterology, Urology · 21 CFR 876.1735 · Class 2

Overview

Product CodeMYE
Device NameSystem, Electrogastrography (Egg)
Regulation21 CFR 876.1735
Device ClassClass 2
Review PanelGastroenterology, Urology

Identification

An electrogastrography system (EGG) is a device used to measure gastric myoelectrical activity as an aid in the diagnosis of gastric motility disorders. The device system includes the external recorder, amplifier, skin electrodes, strip chart, cables, analytical software, and other accessories.

Classification Rationale

Class II (Special Controls). The special controls are as follows:

Special Controls

*Classification.* Class II (Special Controls). The special controls are as follows:(1) The sale, distribution and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter. (2) The labeling must include specific instructions: (i) To describe proper patient set-up prior to the start of the test, including the proper placement of electrodes; (ii) To describe how background data should be gathered and used to eliminate artifact in the data signal; (iii) To describe the test protocol (including the measurement of baseline data) that may be followed to obtain the EGG signal; and (iv) To explain how data results may be interpreted. (3) The device design should ensure that the EGG signal is distinguishable from background noise that may interfere with the true gastric myoelectric signal. (4) Data should be collected to demonstrate that the device has adequate precision and the EGG signal is reproducible and is interpretable.

Recent Cleared Devices (8 of 8)

RecordDevice NameApplicantDecision DateDecision
K252504Gastric AlimetryAlimetry , Ltd.Dec 5, 2025SESE
K240946Gastric AlimetryAlimetry , Ltd.Jul 3, 2024SESE
K232925Gastric Alimetry SystemAlimetry , Ltd.Nov 17, 2023SESE
K223398Gastric Alimetry SystemAlimetry , Ltd.Apr 27, 2023SESE
K213924Gastric Alimetry SystemAlimetry , Ltd.Jun 3, 2022SESE
K212954G-Tech Wireless Patch System (WPS)G-Tech MedicalJan 10, 2022SESE
K014269POLYGRAM NET ELECTROGASTROGRAPHY APPLICATION SOFTWARE, MODEL 9043S0151; POLYGRAF ID, MODEL 9043G0102Medtronic Functional Diagnostics A/SNov 21, 2002SESE
DEN990005THE 3PM EGG MACHINE3pm Co., Inc.Aug 20, 1999DENG

Top Applicants

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...