Vena MicroAngioscope™ System

K251767 · Vena Medical Holdings Corp · LYK · Oct 1, 2025 · Gastroenterology, Urology

Device Facts

Record IDK251767
Device NameVena MicroAngioscope™ System
ApplicantVena Medical Holdings Corp
Product CodeLYK · Gastroenterology, Urology
Decision DateOct 1, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Vena MicroAngioscope™ System is indicated for the observation of the interior of blood vessels of the peripheral vasculature.

Device Story

Vena MicroAngioscope™ System consists of a sterile, single-use imaging angioscope (MA) and a reusable Camera Control Unit (CCU). The MA, featuring a distal radiopaque marker and integrated LED illumination, is inserted into peripheral blood vessels—often with a parent catheter or balloon irrigation—to provide real-time, high-resolution video. The CCU processes signals from the MA, providing power and outputting live video via HDMI to external monitors for physician visualization during vascular procedures. Users can capture still images or record video via CCU controls. The system aids clinicians in direct visualization of vascular structures and pathologies to guide precise interventions. The CCU is designed for use in the angiosuite, where it is draped to maintain sterility.

Clinical Evidence

No clinical or animal testing was performed. Substantial equivalence is supported by non-clinical bench testing, including verification of imaging performance (resolution, latency, field of view), mechanical integrity (tensile, torque, kink resistance), biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1 series), and software validation (IEC 62304).

Technological Characteristics

Materials: medical-grade plastics, glass, silicon, epoxy, stainless steel. Sensing: high-resolution video endoscope with LED illumination. Form factor: 160 cm length, 3F distal/2.7F proximal diameter. Connectivity: HDMI output for external monitors, USB for storage. Sterilization: Ethylene Oxide (EO) to 10^-6 SAL. Software: IEC 62304 compliant.

Indications for Use

Indicated for observation of the interior of blood vessels of the peripheral vasculature in patients undergoing vascular interventions.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 1, 2025 Vena Medical Holdings Corp. Joseph De Croos Director of Regulatory Affairs 809 Wellington St N, Unit 2 Kitchener, Ontario N2H 5L6 Canada Re: K251767 Trade/Device Name: Vena MicroAngioscope™ System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: LYK, DQY Dated: June 9, 2025 Received: June 10, 2025 Dear Joseph De Croos: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251767 - Joseph De Croos Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251767 - Joseph De Croos Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Aneesh S. Deoras -S Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251767 | ? | | Please provide the device trade name(s). | | ? | | Vena MicroAngioscope™ System | | | | Please provide your Indications for Use below. | | ? | | The Vena MicroAngioscope™ System is indicated for the observation of the interior of blood vessels of the peripheral vasculature. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | Vena MicroAngioscope™ System {4} Vena Medical 809 Wellington St. N., Unit 2 Kitchener, ON N2H 5L6, Canada # 510(k) Summary ## I. SUBMITTER INFORMATION Applicant / Manufacturer Name and Address: Vena Medical Holdings Corp. 809 Wellington St N, Unit 2 Kitchener, ON N2H 5L6, Canada Sponsor Contact: Joseph De Croos, Director of Regulatory Affairs 519-897-6654 joseph@venamed.ca Date Prepared: September 30, 2025 ## II. DEVICE Trade Name: Vena MicroAngioscope™ System Common or Usual Name: Angioscope Classification Name: Endoscope and accessories Regulation Number: 21 CFR §876.1500 Regulatory Class: Class II Primary Product Code: LYK ## III. PREDICATE DEVICE Trade Name: Olympus AF Angioscope System 510(k) Number: K911278 ## IV. DEVICE DESCRIPTION The Vena MicroAngioscope System consists of two main components: the Vena MicroAngioscope™ (MA) and the Vena Camera Control Unit (CCU). The Vena MA is a full-color, forward-viewing, imaging angioscope intended to be used in conjunction with fluoroscopy to aid in vascular interventions. Designed to be used either with a parent catheter or balloon irrigation, it has a working length of 160 cm with a diameter of 3 French (0.040”) at the distal tip and 2.7 French (0.036”) at the proximal end. A radiopaque marker band is positioned at the distal tip of the device. The device features a lighting system surrounding a high-resolution video endoscope which provides real-time, high-resolution imaging inside blood vessels. This allows for direct, accurate visualization of vascular structures and pathologies, enabling precise interventions. The proximal end consists of a bifurcated joint for connection to an LED illumination source and a camera connector of the Vena CCU. This angioscope must be connected to the Vena CCU to acquire images and emit light. The Vena MicroAngioscope is the sterile, single use applied part of the Vena MicroAngioscope System. The Vena CCU is a compact, reusable imaging processor designed to interface with the MA to provide power, process images, and output live video. Housed in a durable metal enclosure, the CCU features a K251767 {5} Vena Medical 809 Wellington St. N., Unit 2 Kitchener, ON N2H 5L6, Canada front panel interface with a dedicated camera connector for enabling visualization and an LED connector for powering the integrated illumination system. The unit includes two control buttons: one for capturing still images and another for recording video. The CCU requires the Vena MicroAngioscope to be attached to the front connectors for operation. Real-time imaging is transmitted via an HDMI output, allowing seamless integration with external monitors, while a USB port enables image and video storage. To maintain sterility in the angiosuite, the CCU is designed to be fully draped during procedures. Additionally, the CCU can be grounded by connecting to the equipotential grounding pin. By providing high-quality image processing and intuitive controls, the Vena CCU enables enhanced visualization and procedural efficiency in vascular interventions. # V. INTENDED USE/INDICATIONS FOR USE The Vena MicroAngioscope™ System is indicated for the observation of the interior of blood vessels of the peripheral vasculature. # VI. TECHNOLOGICAL CHARACTERISTICS AND PRODUCT FEATURE COMPARISON | Feature | Subject Device Vena MicroAngioscope™ System | Predicate Device Olympus AF Angioscope System | Comparison | | --- | --- | --- | --- | | Regulatory Clearance | Pending | K911278 | N/A | | FDA Classification | Class II | Class II | Same | | Product Code(s) | LYK | LYK | Same | | Regulation Number | 876.1500 | 876.1500 | Same | | Regulation Name | Angioscope | Angioscope | Same | | Anatomical Locations | Peripheral | Peripheral | Same | | Indication for Use | The Vena MicroAngioscope™ System is indicated for the observation of the interior of blood vessels of the peripheral vasculature. | The endoscopic observation of the interior of blood vessels of the peripheral vasculature. | Same | | Intended Use | The Vena MicroAngioscope™ System is intended for the observation of the interior of blood vessels of the peripheral vasculature. | The endoscopic observation of the interior of blood vessels of the peripheral vasculature | Same | | Material | Commonly used medical grade plastics, glass, silicon, epoxy, stainless steel | Glass, Glass with glue (cemedyne), Stainless Steel, Rubber, Fluorinated ethylene Propylene, Glue #3 | Similar | | Tip Shape | Straight | Straight | Same | K251767 {6} Vena Medical 809 Wellington St. N., Unit 2 Kitchener, ON N2H 5L6, Canada | Feature | Subject Device Vena MicroAngioscopeTM System | Predicate Device Olympus AF Angioscope System | Comparison | | --- | --- | --- | --- | | Radiopaque Markers | Distal Tip has radiopaque marker band | N/A | Similar | | Field of View | 127 degrees | 60-75 degrees | Similar | | Direction of View | 0 degrees | 0 degrees | Same | | Working Length | 1600mm | 1000-1500mm | Similar | | Insertion Tube Diameter | 1mm | 0.5-2.8mm | Same | | Shaft Diameter | 0.9mm | 0.5-2.8mm | Same | | Angulation System | none | none-120 degree | Same | | Channel Diameter | none | none-1.0mm | Same | | Accessories Supplied | Camera Control Unit, Power Cable, HDMI Cable, Sterile Drape | OTV-S2 Medical TV System, TV adapter AR-TF2, BNC video cables | Similar | | How Supplied | Sterile, single use | Non-sterile, Reusable | Similar | | Sterilization Method | EO | EO | Same | | Sterility Assurance Level | 10 to the -6 | 10 to the -6 | Same | ## VII. PERFORMANCE DATA The Vena MA System was verified and validated in accordance with 21 CFR §820.30. The following tests were completed to demonstrate substantial equivalence and that any technological differences do not raise new or different questions of safety and effectiveness. The device successfully passed all testing, and the results demonstrate the device is safe, effective, and performs as well or better than the predicate device. ### Performance Bench Testing - CCU Unit – Compatibility/ Functionality/Recording/Capture - Color - Connector Insertion and Removal Force - Delivery and Retrieval Force - Dimensional Verification - Distortion - Depth of Field - Field of View - Direction of View - Illumination Intensity - Image Intensity Uniformity - Image Latency K251767 Page 3 of 5 {7} Vena Medical 809 Wellington St. N., Unit 2 Kitchener, ON N2H 5L6, Canada - Kink Resistance - Marker Band Position - Particulate - Photobiological Safety - Resolution - Signal to Noise Ratio - Simulated Use (Validation) - Simulated Use (Verification) - Tensile Test - Tip Flexibility - Torqueability - Visual Inspection - Water Ingress ## Packaging & Labeling Testing - Accelerated Aging - Labeling Validation - Labeling Verification - Package and Transit Testing ## Biocompatibility - Acute System Toxicity - Cytotoxicity - Pyrogenicity - Sensitization (Maximization) - Intracutaneous Reactivity - Hemolysis - Complement Activation - Standard Thrombogenicity The passing biocompatibility testing results demonstrate that the subject device meets biological safety requirements per ISO 10993-1. ## Sterilization The Vena MicroAngioscope™ is sold sterile, for single use and single patient only. The Vena MA is sterilized using 100% Ethylene Oxide (EO) gas to achieve a sterility assurance level of 10⁻⁶ in compliance with ISO 11135:2014, "Sterilization of Health-Care Products - Ethylene Oxide - Requirements for The Development, Validation and Routine Control of a Sterilization Process for Medical Devices". ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18 and IEC 60601-4-2. ## Software Verification and Validation Testing Software verification and validation testing in accordance with IEC 62304. K251767 {8} Vena Medical 809 Wellington St. N., Unit 2 Kitchener, ON N2H 5L6, Canada ## Animal Study No animal testing was required to support this 510(k) as the indications for use and technology are substantially equivalent to those of the predicate device. ## Clinical Studies No clinical testing was required to support this 510(k) as the indications for use and technology are substantially equivalent to those of the predicate device. ## VIII. CONCLUSIONS The performance tests have concluded that the Vena MicroAngioscope System, is substantially equivalent in its intended use, design, material, performance, and the underlying scientific technology used, to the predicate device. Non-clinical bench, packaging & labeling, and biocompatibility testing demonstrates that the Vena MicroAngioscope System meets specifications and performs as intended and that any technological differences do not raise new or different questions of safety and effectiveness. K251767
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...