ENDOSCOPIC LIGHT SOURCE, MODEL XL300/L5

K021717 · World of Medicine Lemke GmbH · GCT · Aug 20, 2002 · Gastroenterology, Urology

Device Facts

Record IDK021717
Device NameENDOSCOPIC LIGHT SOURCE, MODEL XL300/L5
ApplicantWorld of Medicine Lemke GmbH
Product CodeGCT · Gastroenterology, Urology
Decision DateAug 20, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Endoscopic Light Source XL300/L5 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

Device Story

The Endoscopic Light Source XL300/L5 is a medical illumination device used in clinical settings during endoscopic diagnostic and surgical procedures. It utilizes a 300 W xenon lamp to generate light, which is transmitted through a connected fiber optic cable to provide illumination within body cavities, hollow organs, and canals. The device features manual brightness adjustment. It is operated by healthcare professionals in a clinical environment. The output is visible light, which enables the physician to visualize internal anatomy during endoscopic procedures, thereby facilitating diagnostic or surgical intervention.

Clinical Evidence

Bench testing only. The device complies with IEC 60601-1, IEC 60601-1-2, and UL2601-1 standards, and conforms to the Medical Device Directive 93/42/EEC.

Technological Characteristics

300 W xenon lamp; 5600 °K color temperature; ~500-hour lamp life; manual brightness control; fiber optic cable interface. Complies with IEC 60601-1, IEC 60601-1-2, and UL2601-1.

Indications for Use

Indicated for use with fiber optic endoscopes to provide illumination of body cavities, hollow organs, and canals during endoscopic diagnostic and surgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K02/7/7 pg ' of r ## 510(K) SUMMARY Endoscopic Light Source XL300/L5 #### I. Submitter: WORLD OF MEDICINE Lemke GmbH Danziger Strasse 21 82194 Gröbenzell Germany #### II. Device Names: - Classification Name: 1. - Common or Usual Name: 2. - Proprietary Name: 3. Accessory to an Endoscope Endoscopic Light Source Endoscopic Light Source XL300/L5 #### III. Classification: Class II. This device is described in 21 C.F.R. § 876.1500. The product code for the device is GCT. #### IV. Predicate Device: - Karl Storz Xenon 300 Light Source for Non-Flash Applications . (K962595) manufactured by Karl Storz Endoscopy-America, Inc. #### V. Intended Use: The Endoscopic Light Source XL300/L5 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures. #### VI. Device Description: The Endoscopic Light Source XL300/L5 uses a 300 W xenon lamp to provide illumination during endoscopic diagnostic and surgical procedures through a fiber optic cable, which is connected to the device. Brightness can be adjusted manually. The color temperature of the xenon lamp is approximately 5600 °K and the lamp life is approximately 500 hours. #### VII. Substantial Equivalence: The Endoscopic Light Source XL300/L5 described in this notification is similar in design and technological characteristics to the Karl Storz Xenon 300 Light Source for Non-Flash Applications (K962595) manufactured by Karl Storz Endoscopy-America, Inc. {1}------------------------------------------------ Koz1717 pg. 2 of 2 ## 510(K) SUMMARY Page -2-1-2- Both the Endoscopic Light Source XL300/L5 and the predicate device is intended to provide illumination of body cavities, hollow organs and canals during endoscopic diagnostic and surgical procedures. In addition, both devices are designed to be used with a 300 W Xenon lamp. The differences between the Endoscopic Light Source XL300/L5 and predicate device are minor and raise no new questions of safety and effectiveness. Accordingly, WORLD OF MEDICINE Lemke GmbH believes that the Endoscopic Light Source XL300/L5 is substantially equivalent to the predicate device currently on the market. ### VIII. Performance Data: The Endoscopic Light Source XL300/L5 will comply with the International Standard IEC 60601-1, IEC 60601-1-2 and will conform to the Medical Device Directive 93/42/EEC. In addition, the device will meet the requirements of the Underwriters Laboratories Standard UL2601-1. Signed: S. Roal Susanne Raab Official Correspondent {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of what appears to be an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 2 0 2002 WORLD OF MEDICINE LEMKE GmbH c/o Ms. Susanne Raab Official Correspondent 91 Trowbridge Street CAMBRIDGE MA 02138 Re: K021717 Trade/Device Name: Endoscopic Light Source XL300/L5 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 GCT Dated: May 11, 2002 Received: May 23, 2002 Dear Ms. Raab: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## STATEMENT OF INDICATIONS FOR USE APPLICANT: WORLD OF MEDICINE Lemke GmbH 510(K) NUMBER (if known): 021717 DEVICE NAME: Endoscopic Light Source XL300/L5 INDICATIONS FOR USE: The Endoscopic Light Source XL300/L5 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures. (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 C.F.R. § 801.109) (Optional Format 1-2-96) Prescription Use (Per 21 CFR 801.109) (Per 21 CFR 801.109) Nancy C. Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...