← Product Code [GCT](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCT) · K013880

# DAYLITE XENON LIGHT SOURCES (K013880)

_Designs For Vision, Inc. · GCT · Dec 7, 2001 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCT/K013880

## Device Facts

- **Applicant:** Designs For Vision, Inc.
- **Product Code:** [GCT](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCT.md)
- **Decision Date:** Dec 7, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Designs for Vision Daylite Light Sources are indicated for use in surgery and medical applications where high intensity illumination is required.

## Device Story

Daylite XeNon Light Sources provide high-intensity illumination for surgical procedures; system includes xenon light source, fiberoptic cables, and fiberoptic headsets. Headsets are coaxial, bifurcated, or focusable. Light source features universal chuck for cable attachment, 180-watt or 300-watt power output, and continuous illumination level adjustment providing 6000°K color temperature. Used in surgical suites by clinicians to provide illumination for surgical headlights. Device enhances visibility during medical applications.

## Clinical Evidence

Bench testing only; electrical safety characteristics demonstrated.

## Technological Characteristics

Xenon light source; 180W or 300W power output; 6000°K color temperature; fiberoptic delivery system; universal chuck for cable attachment; continuous illumination adjustment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Walter Lorenz Surgical Inc. Lumenon Xenon Light Source ([K992050](/device/K992050.md))
- Welch Allyn, Inc. Xenon 300 Light Source ([K980281](/device/K980281.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K013880
Page 1 of 2

## DEC 0 7 2001

# 510(k) Summary for Designs for Vision, Inc. Daylite Light Sources

#### SPONSOR 1.

Designs for Vision, Inc. 760 Koehler Avenue Ronkonkoma, NY 11779

Amy Grumet-Avallone Contact Person: (631) 585-3300 Telephone:

November 1, 2001 Date Prepared:

#### DEVICE NAME 2.

Daylite XeNon Light Sources Proprietary Name: Common/Usual Name: Surgical Lights Surgical Lights/Routine Fiberoptic Lights Classification Name:

#### PREDICATE DEVICES 3.

Walter Lorenz Surgical Inc. Lumenon Xenon Light Source K992050 K980281 Welch Allyn, Inc. Xenon 300 Light Source

#### DEVICE DESCRIPTION 4.

The Daylite XeNon Light Sources are comprised of high intensity xenon light sources, fiberoptic cables and fiberoptic headsets. The Designs for Vision Daylite Light Sources are supplied with headsets that were first placed into service prior to 1976 and are preamendment devices. The Designs for Vision headsets have been marketed since the early 1970's with a long history of safe use in the surgical suite. The headsets are either coaxial, bifurcated, or focusable designs.

{1}------------------------------------------------

K013880
Page 2 of 2

The light source includes a universal chuck for fiberoptic cable attachment. The universal chuck accepts various sizes of light cables. The XeNon Light sources provide either 180-watt or 300-watt power output and contain a continuous illumination level adjustment which provides 6000°K color temperature light.

#### ર. INTENDED USE

·The Designs for Vision Daylite Light Sources are indicated for use in surgery and medical applications where high intensity illumination is required.

### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.

The Daylite Xenon Light Source devices have the same intended use as the predicate device and similar technological characteristics. They all consist of Xenon Light Sources supplying fiberoptic illuminators with illumination for headlights.

#### PERFORMANCE TESTING 7.

Testing has been performed which demonstrates the electrical safety characteristics of the Designs for Vision, Inc. Daylite XeNon Light Sources.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure embracing a bird, both rendered in a simple, flowing line style.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 0 7 2001

Designs for Vision, Inc. c/o Mr. Donald James Sherratt Medical Stream Director Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K013880

Trade/Device Name: DayLite XeNon Light Sources Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Codes: 78 GCT Dated: November 21, 2001 Received: November 23, 2001

Dear Mr. Sherratt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed noutication. The I Dri miaing of basion for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your ae of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on and please note the regulation entitled, "Misbranding Office of Compilance at (301) 67 - 105 (21 CFR Part 807.97). Other general information on by reletence to premation (on the Act may be obtained from the Division of Small Manufacturers, your responsibilities and Consumer & For toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K013880

510(k) Number (if known):

Device Name: Designs for Vision, Inc. Daylite XeNon Light Sources

Indications For Use:

The Daylite Xenon Light Source is indicated for use in surgery and medical applications where high intensity illumination is required

(Please do not write below this line - continue on another page if necessary)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel le. Benjam

Prescription Use V (Per 21 CFR 801.109) OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCT/K013880](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/GCT/K013880)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).