Viction Disposable Irrigation and suction System (VC-FLA0514; VC-FLA0523; VC-FLA0532; VC-FLB0514; VC-FLB0523; VC-FLB0532; VC-FLC0514; VC-FLC0523; VC-FLC0532; VC-FLD0514; VC-FLD0523; VC-FLD0532)

K250130 · Hangzhou Viction Medical Technology Co.,Ltd · GCJ · Aug 6, 2025 · Gastroenterology, Urology

Device Facts

Record IDK250130
Device NameViction Disposable Irrigation and suction System (VC-FLA0514; VC-FLA0523; VC-FLA0532; VC-FLB0514; VC-FLB0523; VC-FLB0532; VC-FLC0514; VC-FLC0523; VC-FLC0532; VC-FLD0514; VC-FLD0523; VC-FLD0532)
ApplicantHangzhou Viction Medical Technology Co.,Ltd
Product CodeGCJ · Gastroenterology, Urology
Decision DateAug 6, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Viction Disposable Irrigation and Suction system is intended to provide irrigation and suction functions during general surgery and laparoscopic surgery. It is designed to deliver irrigation fluids to surgical sites and to remove fluid waste and tissue debris.

Device Story

Disposable manual surgical instrument for irrigation and suction during general and laparoscopic procedures. Device consists of stainless steel inner/outer tubes, hand-piece with on-off control button, and conical connectors. Irrigation functions via gravity-fed fluid flow; suction connects to hospital negative pressure source via Yankee joint. Operated by surgeons in clinical settings to clear surgical sites of fluid and debris, improving visibility. Single-use, sterile, ethylene oxide-sterilized device.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including fluid flow rate verification (irrigation ≥160 ml/min, suction ≥600 ml/min), negative pressure resistance (up to 60 kPa), sealing integrity, coating adhesion (ASTM D3359-23), and biocompatibility testing (ISO 10993-1: cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity).

Technological Characteristics

Materials: Stainless steel with PTFE coating, ABS, PVC, Silicone. Principle: Gravity-fed irrigation; hospital-supplied negative pressure suction. Dimensions: 5mm diameter; 14cm, 23cm, or 32cm lengths. Sterilization: Ethylene oxide (ISO 11135), residue ≤10μg/g. Performance: Irrigation flow ≥160 ml/min, suction flow ≥600 ml/min, max negative pressure 60 kPa. Coating adhesion per ASTM D3359-23.

Indications for Use

Indicated for use in general and laparoscopic surgery to deliver irrigation fluids to surgical sites and remove fluid waste and tissue debris.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION August 6, 2025 Hangzhou Viction Medical Technology Co.,Ltd Zhicong Luo General Manager First Floor ,No.1 Building, No.95 Binwen Road, Binjiang District Hangzhou, Zhejiang 310051 China Re: K250130 Trade/Device Name: Viction Disposable Irrigation and suction System (VC-FLA0514; VC-FLA0523; VC-FLA0532; VC-FLB0514; VC-FLB0523; VC-FLB0532; VC-FLC0514; VC-FLC0523; VC-FLC0532; VC-FLD0514; VC-FLD0523; VC-FLD0532) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ Dated: July 7, 2025 Received: July 7, 2025 Dear Zhicong Luo: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1} K250130 - Zhicong Luo Page 2 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical- {2} K250130 - Zhicong Luo Page 3 devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Julie A. Morabito -S Julie Morabito, Ph.D. Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250130-S001 Device Name Viction Disposable Irrigation and suction System (VC-FLA0514;VC-FLA0523;VC-FLA0532;VC-FLB0514;VC-FLB0523;VC-FLB0532;VC-FLC0514;VC-FLC0523;VC-FLC0532;VC-FLD0514;VC-FLD0523;VC-FLD0532) Indications for Use (Describe) The Viction Disposable Irrigation and Suction system is intended to provide irrigation and suction functions during general surgery and laparoscopic surgery. It is designed to deliver irrigation fluids to surgical sites and to remove fluid waste and tissue debris. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} 510 (K) Summary K250130 This summary of 510(K)safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92. Type of submission : Traditional The date the summary was prepared: July 4, 2025 I. SUBMITTER Hangzhou Viction Medical Technology Co., Ltd First floor, No.1 Building, No.95 Binwen Road, Binjiang District, Hangzhou City, Zhejiang Province, China. MD310051 Contact Person: Yu Gu Phone: +86-057188305089 Email: guyu@victionmed.com Date Prepared: February 11, 2025 II. Device Name of Device : Viction Disposable Irrigation and Suction system, (model VC-FLA0514; VC-FLA0523; VC-FLA0532; VC-FLB0514; VC-FLB0523; VC-FLB0532; VC-FLC0514; VC-FLC0523; VC-FLC0532; VC-FLD0514; VC-FLD0523; VC-FLD0532) Common or Usual Name: Suction Irrigation Classification Name: Endoscope And Accessories (21 CFR876.1500) Regulatory Class: 2 Product Code: GCJ III. PREDICATE DEVICE Single Use Surgical Suction Irrigator (K161421), manufactured by Jiangsu Haize Medical Scientific Development Co., Ltd. Located in WiXi city, JiangSu Province, china. This predicate has not been subjected to a design-related recall. No reference devices were used in this submission. IV. DEVICE DESCRIPTION The Viction disposable Irrigation and suction system are constitutes with Irrigation & suction device and a double tubes (optional accessories). The Irrigation & suction device including an inner/outer stainless steel tubes, On-off Control Button, Conical connectors and Hand-piece. Double tubes constitutes with Irrigation {5} tube, suction tube, and it connects with Yankee joint. Insert needle and Robert Clip. The Viction Disposable Suction Irrigation system is used to deliver sterile irrigation fluids to surgical sites during laparoscopic and endoscopic procedures. The Irrigation & suction device including an inner/outer stainless steel tubes, On-off Control Button, Conical connectors and Hand-piece. On-off control button on handle-piece can control fluid circuit on and off. It is designed to deliver sterile irrigation fluids to surgical sites during surgery procedures. Double tubes constitutes with Irrigation tube, suction tube, and it connects with Yankee joint, Insert needle and Robert Clip. The suction tube is connected to the negative pressure device of the hospital through the Yankee joint. The waste liquid is discharged through a suction tube under the negative pressure environment. Which to improve the surgeon's visibility. There are 12 models of the product, models VC-FLA0514; VC-FLA0523; VC-FLA0532; VC-FLB0514; VC-FLB0523; VC-FLB0532; VC-FLC0514; VC-FLC0523; VC-FLC0532; VC-FLD0514; VC-FLD0523; VC-FLD0532. All product's Structure are the same, the length of the irrigation suction tubes is the main difference. It is a single use, disposable device and is sold sterile. ## V. INDICATION FOR USE The Viction Disposable Irrigation and Suction system is intended to provide irrigation and suction functions during general surgery and laparoscopic surgery. It is designed to deliver irrigation fluids to surgical sites and to remove fluid waste and tissue debris. ## Technological characteristics summary The irrigation function of the disposable irrigation and suction system is achieved through the gravity of the liquid itself. The flushing fluid flows downward under its own gravity through the irrigation suction probe, forming a flushing water flow with a certain pressure. The suction tube is connected to the negative pressure device of the hospital through the Yankee joint. The waste liquid is discharged through a suction tube under the negative pressure environment. Which to improve the surgeon's visibility. All components should be securely connected: the irrigation flow rate should be no less than 160 ml/min; the suction tube flow rate should be no less than 600 ml/min; {6} the product is sterilized with ethylene oxide and should be sterile, with ethylene oxide residue ≤10μg/g; the maximum negative pressure the product can withstand is 60 kPa. This product is not a therapeutic or diagnostic medical device, the product has no diagnostic, therapeutic, preventive, mitigation or cure of disease, no monitoring function. ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS ## WITH THE PREDICATED DEVICE The Viction Disposable Suction Irrigation system is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance. **Predicate device name**: Single Use Surgical Suction Irrigator **510(K) number**: K161421 **Common Name**: Suction Irrigation **Manufacturer**: JiangSu Medical Scientific Development CO., LTD. The following table shows similarities and differences of indications for use between our device and the predicate devices. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness. | Item | Predicate device | Proposed device | Comparison | | --- | --- | --- | --- | | Trade Name | Single Use Surgical Suction Irrigator | Viction Disposable Irrigation and suction system | / | | 510(K)Number | K161421 | K250130 | / | | Classification regulation | 21 CFR 876.1500 | 21 CFR 876.1500 | Equivalence | | Classification and Code | ClassII, GCJ | ClassII, GCJ | Equivalence | | Device Classification Name | Laparoscope, General & Plastic Surgery | Laparoscope, General & Plastic Surgery | Equivalence | {7} | Indications for Use | The Single Use Surgical Suction Irrigator is indicated to provide suction and irrigation functions in general surgery and laparoscopic surgery to help remove dropsy, he matocele and tissue debr is by vacuum aspiration and deliver irrigation fluids to surgical site. | The Viction Disposable Irrigation and Suction system is intended to provide irrigation and suction functions during general surgery and laparoscopic surgery. It is designed to deliver irrigation fluids to surgical sites and to remove fluid waste and tissue debris. | Similar | | --- | --- | --- | --- | | Suction Irrigation probe specification (Diameter×Length) | JM-CX1A JM-CX4, JM-CX4B (45cm×5mm) | 5mm×32cm; 5mm×23cm; 5mm×14cm; | Similar | | Patient Contacting Material | Stainless steel ABS PVC Silicone | Stainless steel with PTEF coating. ABS PVC Silicone | Similar | | Biocompatibility | Cytotoxicity Test: Intracutaneous Reactivity Test; Acutesystemic Toxicity Test; pyrogen testing; | Cytotoxicity Test: Intracutaneous Reactivity Test; Acutesystemic Toxicity Test; pyrogen testing; | Similar | | Sterilization | EO Sterilized/ISO 11135 | EO Sterilized/ISO 11135 | Equivalence | | Disposable | Yes | Yes | Equivalence | | Additional accessories | Yes | Yes | Equivalence | ## Conclusion Based on the comparison of indications of use and other product performances. The Viction Disposable Suction Irrigation system manufactured by “Hangzhou Viction Medical Technology Co., Ltd.” is substantial equivalent to its predicate devices. {8} # VII. PERFORMANCE DATA Non-clinical tests were conducted to verify that the subject device met all design specifications to be considered substantially equivalent to the predicate device. ## Electrical safety and electromagnetic compatibility Not applicable to this submission ## Biocompatibility testing The biocompatibility evaluation for the Viction disposable Irrigation and suction system device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. More details of biocompatibility tests are summarized in section 15. which include the following tests: - Acute Systemic Toxicity - Skin Sensitization Test - Intracutaneous Reactivity Test - Pyrogen Test - In Vitro Cytotoxicity Test ## Bench testing The tests listed below evaluated the performance of the subject device. - Fluid flow rate of Irrigation tubes and suction tubes. The test was conducted to ensure the flow rate of the irrigation tubes and suction tubes met the predefined acceptance criteria. - Negative pressure resistance of suction. - Sealing of irrigation tube. - coating adhesion. The test was conducted to ensure the test results complied with standard 3B Per ASTM D3359-23. - Particulate contamination index. ## Packaging and Sterilization - Ethylene Oxide Sterilization Validation - Packaging Validation ## Clinical Testing Not applicable to this submission All the test results meets the predefined acceptance criteria and complied with the design specification of the subject device throughout the use life. The results of the {9} non-clinical testing demonstrate that the victim disposable irrigation and suction system is as safe and effective as the predicate device. ## VIII. CONCLUSIONS Based upon the intended use and known technical information provided in this pre-market notification, the Victim Disposable Irrigation and Suction system has been shown to be substantially equivalent to currently marketed predicate device K161421.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...