Sterile gel for human use, e.g. in medical procedures - to aid the introduction of catheters and endoscopic instruments in transurethral examinations, and in intermittent catheterization - particularly for incontinence treatment.
Device Story
Lubricano Sterile Gel is a sterile, water-soluble lubricating jelly composed of hydroxyethylcellulose, glycerol, and purified water. It is supplied in a 10 mL syringe and is free from fats, latex, disinfectants, and preservatives. The gel is applied by clinicians during transurethral medical procedures to reduce friction, ensuring easy movement of catheters and endoscopic instruments while adhering to the mucosa. By facilitating instrument insertion, the device aids in transurethral examinations and intermittent catheterization, particularly for patients managing incontinence. The device is intended for professional use in clinical settings.
Clinical Evidence
In vivo and in vitro testing established the safety profile of the device as non-toxic, non-irritating, and non-sensitizing, comparable to the predicate device.
Technological Characteristics
Sterile, water-soluble gel. Composition: hydroxyethylcellulose, glycerol, and purified water. Free from fats, latex, disinfectants, and preservatives. Supplied in a 10 mL syringe. Mechanical lubrication principle.
Indications for Use
Indicated for human use to facilitate the insertion of catheters and endoscopic instruments during transurethral examinations and intermittent catheterization, including for patients undergoing incontinence treatment.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Related Devices
K083189 — STERI-LUB LUBRICATION GEL · Avail Medical Products, Inc. · Nov 21, 2008
K101522 — ULTRA SEAL STERILE LUBRICATING JELLY · Ultra Seal Corporation · Dec 23, 2010
K112110 — STERILE LUBRICATING JELLY · Jianerkang Medical Dressing Company · Apr 10, 2012
K103718 — DYNACOR LUBRICATION GEL · Medline Industries, Inc. · May 10, 2011
K073684 — SHEFFIELD PHARMACEUTICALS (STERILE) LUBRIGEL · Sheffield Industries, Div. of Faria , Ltd. · May 22, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Ko8199D
DEC 0 5 2008
### 510(k) Summary
Submitted by Farco-Pharma GmbH Pharmazeutische Präparate Gereonsmühlengasse 1 - 11 Address: 50670 Köln. Germany
Telephone: +49 221 594061
Contact Name: Kornelia Elv-Koort, Regulatory Affairs Dept.
October 1, 2008 Date Submitted:
Lubricano® Sterile Gel - Transurethral Trade Name: Lubricating jelly for transurethral surgical instrument Common Name: Product Code / Regulation: FHX (21 C.F.R. 876.1500)
Description: Lubricano® Sterile Gel is a sterile, water-soluble gel composed of hydroxyethylcellulose, glycerol, and purified water, which is contained in a 10 mL syringe. Lubricano® Sterile Gel is free from fats, latex, disinfectants and preservatives. Lubricano® Sterile Gel ensures that catheters and instruments move easily, and it adheres well to the mucosa.
Intended Use:Sterile gel for human use, e.g. in medical procedures - to aid the introduction of catheters and endoscopic instruments in transurethral examinations, and in intermittent catheterization - particularly for incontinence treatment.
Substantial Equivalence:
Lubricano Sterile Gel is similar in intended use and technological characteristics to the predicate lubricating jelly for transurethral surgical instruments. The device is similar with respect to indications for use and physical characteristics to the predicate device in terms of 510(k) substantial equivalency.
Results of in vivo and in vitro testing establishes the safety profile of the device as non-toxic, non-irritating and non-sensitizing, which is comparable to the predicate device.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract eagle or bird-like figure, composed of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FARCO-Pharma GmbH % Mr. Seth A. Mailhot Attorney Latham & Watkins LLP 555 Eleventh Street, NW WASHINGTON DC 20004-1304
# DEC 0 5 2008
Re: K081990
Trade/Device Name: Lubricano® Sterile Gel Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FHX Dated: November 19, 2008. Received: November 20, 2008
#### Dear Mr. Mailhot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Arput Mr. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosur
{3}------------------------------------------------
Ko81990
# Indications for Use
510(k) Number (if known):
Device Name: Lubricano® Sterile Gel - Transurethral
Indications for Use:
Sterile gel for human use, e.g. in medical procedures - to aid the introduction of catheters and endoscopic instruments in transurethral examinations, and in intermittent catheterization particularly for incontinence treatment.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
4-2
Heidi Lewis
(Division Sign-OH)
1 Division of Reproductive, Abdominal & 1
**Radiological Devices**
510(k) Number
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.