← Product Code [FFX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFX) · K183072

# EndoFLIP System (K183072)

_Crospon, Ltd. · FFX · Feb 15, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFX/K183072

## Device Facts

- **Applicant:** Crospon, Ltd.
- **Product Code:** [FFX](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFX.md)
- **Decision Date:** Feb 15, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1725
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The Endoflip System is indicated for use in a clinical setting to measure and dimensions in the esophagus, pylorus, and anal sphincters in adults and to measure pressure and dimensions in patients from 5 years of age. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.

## Device Story

EndoFLIP System measures pressure and dimensions within GI tract; utilizes single-use catheter with balloon distension; used in clinical settings by physicians. System provides real-time assessment of esophageal, pyloric, and anal sphincter function; aids in evaluating motility disorders. In pediatric patients, device used during myotomy surgeries to assess myotomy adequacy and in Eosinophilic esophagitis (EoE) to measure esophageal distensibility. Output displayed to clinician for intraoperative or diagnostic decision-making; benefits include objective assessment of GI anatomy and function.

## Clinical Evidence

Clinical evaluation based on literature review of pediatric use. Evidence demonstrates safety and clinical utility in pediatric achalasia patients during myotomy (real-time assessment of myotomy adequacy) and in children with Eosinophilic esophagitis (EoE) to assess esophageal distensibility.

## Technological Characteristics

System consists of a measuring console and single-use catheter. Provides programmed balloon distension. No changes in materials, sensing principles, or energy sources compared to predicate K160725.

## Regulatory Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

## Predicate Devices

- Endoflip System ([K160725](/device/K160725.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.

February 15, 2019

Crospon Ltd % Avishag Metzer Regulatory Affairs Program Manager Medtronic 2 Hacarmet St. New Industrial Park POB 258 Yogneam 20962 ISRAEL

Re: K183072

Trade/Device Name: EndoFLIP System Regulation Number: 21 CFR 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: Class II Product Code: FFX Dated: October 31, 2018 Received: November 5, 2018

Dear Avishag Metzer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183072

Device Name Endoflip System

#### Indications for Use (Describe)

The Endoflip System is indicated for use in a clinical setting to measure and dimensions in the esophagus, pylorus, and anal sphincters in adults and to measure pressure and dimensions in patients from 5 years of age. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### I. SUBMITTER

| Submitter Name and Address:        | Crospon Ltd. (Medtronic)                                                   |
|------------------------------------|----------------------------------------------------------------------------|
|                                    | Galway Business Park                                                       |
|                                    | Dangan                                                                     |
|                                    | Galway, Ireland                                                            |
| Contact Person:                    | Avishag Metzer                                                             |
|                                    | Regulatory Affairs Program Manager                                         |
|                                    | Email: Avishag.Metzer@medtronic.com                                        |
| Phone Number:                      | 972 (4) 9097983                                                            |
| Fax Number:                        | 972 (73) 2501545                                                           |
| Establishment Registration Number: | 3006897778                                                                 |
| Date Prepared:                     | February 15, 2019                                                          |
| II. DEVICE                         |                                                                            |
| Device Trade Name(s):              | Endoflip System                                                            |
| Device Common Name:                | Gastrointestinal motility monitoring system                                |
| Classification:                    | Regulation: 21 CFR 876.1725 Gastrointestinal motility<br>monitoring system |
|                                    | Product Code: FFX: System, Gastrointestinal Motility<br>(Electrical)       |

# III. PREDICATE DEVICE(S)

Endoflip System K160725

# IV. DEVICE DESCRIPTION

There is no change between the Predicate device (K160725) and Endoflip System.

The Endoflip System comprises a measuring system and a single use catheter to provide programmed balloon distensions.

{4}------------------------------------------------

The Endoflip System and catheter are identical to the predicate (K160725). The proposed modification to the Endoflip System is to add a specific age range that includes pediatrics to the indications for use statement.

## V. INDICATIONS FOR USE

The Endoflip System is indicated for use in a clinical setting to measure pressure and dimensions in the esophagus, pylorus, and anal sphincters in adults and to measure pressure and dimensions in the esophagus, in patients from 5 years of age. It is intended to be used as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.

# VI. TECHNOLOGICAL CHARACHTERISTICS

There is no change in the technological characteristics of Endoflip System from the cleared predicate device (K160725).

# VII. PERFORMANCE DATA

## Clinical Evaluation

The body of evidence presented demonstrates the safety of the Endoflip device and highlights the potential clinical benefit of utilization in the pediatric population. Specifically, Endoflip was used in pediatric achalasia patients during myotomy surgeries enabling real time assessment of the adequacy of the myotomy. Other application described in the literature review involves the use of Endoflip in children with Eosinophilic esophagitis (EoE) to assess esophageal distensibility.

The proposed change in this submission does not raise new performance or safety issues.

## VIII. CONCLUSION

Based on the clinical evaluation presented, the company believes that the Endoflip System and the predicate device are substantially equivalent and do not raise new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFX/K183072](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FFX/K183072)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
