PH ANALYSIS MODULE
Device Facts
| Record ID | K961346 |
|---|---|
| Device Name | PH ANALYSIS MODULE |
| Applicant | Synectics-Dantec |
| Product Code | FFT · Gastroenterology, Urology |
| Decision Date | Jan 29, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1400 |
| Device Class | Class 1 |
| Attributes | Software as a Medical Device, Pediatric |
Indications for Use
The pH Analysis Module is a software program that has been designed to analyze intragastroesophageal pH data recorded from the gastrointestinal tract in pediatric and adult populations. The program is to be used on a personal computer analyzing patient data in the hospital environment under supervision of a trained physician. The analyzed data can be viewed on the computer screen or printed out on a separate paper.
Device Story
pH Analysis Module; software add-on for Polygram Software for Windows. Inputs: recorded intragastroesophageal pH data. Processing: automated analysis of pH tracings; calculation of parameters including duration of recording, number of periods of interest, time below pH threshold, symptom correlation, and scoring. Output: analysis report containing patient demographics, procedure summary, pH tracing, period table, and physician signature section. Used in hospital environment; operated by trained physician. Benefit: assists physician in diagnosis and post-treatment evaluation of gastrointestinal pH data.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based analysis module; runs on personal computer; Windows-based environment; integrates with Polygram Software for Windows platform.
Indications for Use
Indicated for analysis of intragastroesophageal pH data recorded from the gastrointestinal tract in pediatric and adult populations.
Regulatory Classification
Identification
A stomach pH electrode is a device used to measure intragastric and intraesophageal pH (hydrogen ion concentration). The pH electrode is at the end of a flexible lead which may be inserted into the esophagus or stomach through the patient's mouth. The device may include an integral gastrointestinal tube.
Predicate Devices
- EsopHogram Software (K859361)
Reference Devices
- Polygram Software for Windows (K946322)
Related Devices
- K961070 — ESOPHAGEAL MANOMETRY ANALYSIS MODULE · Synectics Medical, Inc. · Jun 11, 1996
- K011471 — BILITEC 2000, POLYGRAM '98 PH TESTING APPLICATION (INCL.BILE), OPTICAL FIBER PROBER · Medtronic Functional Diagnostics A/S · Nov 8, 2002
- K981733 — DIGITRAPPER DELTA, MODEL 9043G0201 AND POLYGRAM '98 PH EXTENSION MODEL 9043S0111 · Medtronic Functional Diagnostics A/S · Feb 4, 1999
- K060166 — OHMEGA · Medical Measurement Systems B.V. · Apr 21, 2006
- K102543 — BRAVO PH MONITORING SYSTEM AND ACCESSORIES · Given Imaging , Ltd. · Dec 1, 2010