Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100)

K241532 · Shanghai SeeGen Photoelectric Technology Co., Ltd. · FET · Oct 22, 2024 · Gastroenterology, Urology

Device Facts

Record IDK241532
Device NameFlexible Video-Choledo-Cysto-Ureteroscope System (PL-2100)
ApplicantShanghai SeeGen Photoelectric Technology Co., Ltd.
Product CodeFET · Gastroenterology, Urology
Decision DateOct 22, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

Device Story

Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100) is a medical electronic optical instrument for visualizing human cavities (bladder, ureter, biliary, pancreatic ducts). Device consists of a flexible insertion tube with mechanical bending system and a distal CMOS digital imaging sensor. Input: optical images from human cavity; transformation: lens system converts light to electrical signals, processed by an imaging processor board; output: video displayed on a monitor for physician observation. Used in medical facilities by physicians. Output assists in real-time diagnostic and therapeutic decision-making during endoscopic procedures. Benefits include minimally invasive visualization and access to internal anatomy for diagnosis and treatment.

Clinical Evidence

No clinical data included. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993), sterilization validation, EO residual testing, transportation testing, photobiological safety, and optical/color performance metrics (MTF, SNR, dynamic range, color reducibility).

Technological Characteristics

System includes flexible endoscope probe with CMOS sensor and imaging processor. Features LED 2W light source (30,000 hrs life). Imaging output via CVBS, DVI, S-VIDEO; includes 10.1-inch integrated screen. Connectivity includes USB 2.0 (FAT32). Biocompatibility per ISO 10993; electrical safety per IEC 60601-1/2-18; EMC per IEC 60601-1-2. Sterilization via EO.

Indications for Use

Indicated for patients requiring endoscopic examination of the urinary tract, percutaneous examination of the kidney interior, or surgical examination of bile ducts, including diagnostic and therapeutic procedures performed with additional accessories.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. October 22, 2024 Shanghai SeeGen Photoelectric Technology Co., Ltd. Yihua Ma RA Supervisor 3 Floor, Building No.1, 4299 JinDu Road, Minhang District Shanghai. 201108 CHINA Re: K241532 Trade/Device Name: Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET, KQM Received: September 11, 2024 Dear Yihua Ma: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2}------------------------------------------------ Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, **Mark R. Kreitz -S** for Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K241532 Device Name Flexible Video-Choledo-Cysto-Ureteroscope System (PL-2100) #### Indications for Use (Describe) The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. # 5.1 Submitter | Submitted by: | Shanghai SeeGen Photoelectric Technology Co., Ltd. | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Address: 3 Floor, Building No.1, 4299 JinDu Road, Minhang<br>District, Shanghai, China | | Contact<br>Person: | Yihua Ma<br>RA Supervisor | | | Shanghai SeeGen Photoelectric Technology Co., Ltd.<br>Address: 3 Floor, Building No.1, 4299 JinDu Road, Minhang<br>District, Shanghai, China<br>Phone: 0086-18616909737<br>Email: mayihua@seegen.com.cn | | Date<br>Prepared: | May 30, 2024 | ### 5.2 Device | Device Name: | Flexible Video-Choledo-Cysto-Ureteroscope<br>System | |----------------------|--------------------------------------------------------------------------------------| | Common Name: | Choledochoscope and Accessories, Flexible/Rigid;<br>Camera, Surgical and Accessories | | Regulatory Class: | Class II | | Regulation Number: | 21 CFR 876.1500 | | Classification Name: | Endoscopic Video Imaging System/Component, GU | | Product Code: | KQM,FET | ### 5.3 Predicate Device | Device Name: | Flexible Video-Choledo-Cysto-Ureteroscope<br>System, K211686 | |--------------------|-----------------------------------------------------------------------------------------------------------| | Common Name: | Choledochoscope and Accessories, Flexible/Rigid;<br>Camera, Surgical and Accessories;<br>LED light Source | | Regulatory Class: | Class II | | Regulation Number: | 21 CFR 876.1500 | | Regulation Name: | Endoscope and Accessories | | Product Code: | FGB, FBN, FET, FAJ, FGA | {5}------------------------------------------------ #### 5.4 Device Description Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface. Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends. The product is equipped with tiny size digital imaging parts -- photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis. #### 5.5 Indication for Use: Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for The endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy. ### 5.6 Substantial Equivalence and Technological Characteristics {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for SEEGEN. The logo is red and black, with the word "SEEGEN" in large, bold letters. Below the word "SEEGEN" is the phrase "vision borderless" in smaller letters. To the right of the word "SEEGEN" are some Chinese characters. | Item | Imaging Processor System(Including Light Source)<br>(Proposed Device),Model:PL-2100 | Imaging Processor System(Including Light Source),<br>K211686(Primary Predicate Device), Model:PL-1000 | Comment | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Indication for Use | Signals from the endoscope are converted to the image and<br>displayed on an inspection monitor. | Signals from the endoscope are converted to the image<br>and displayed on an inspection monitor. | Same | | Target patients: | Patients who are considered suitable for the application of<br>this product by the physician | Patients who are considered suitable for the application of<br>this product by the physician | Same | | User<br>qualifications | Physicians (experts who have been approved by the<br>endoscopic medical safety administrator at each medical<br>facility. If the eligibility requirements are defined by an<br>official body such as a government entity and/or an<br>academic society, follow such requirements.) Specific<br>training to use this product is not required. | Physicians (experts who have been approved by the<br>endoscopic medical safety administrator at each medical<br>facility. If the eligibility requirements are defined by an<br>official body such as a government entity and/or an<br>academic society, follow such requirements.) Specific<br>training to use this product is not required. | Same | | Place of use | Medical facility | Medical facility | Same | | Power input | 100-240V~50/60Hz48VA | 100-240V~50/60Hz45VA | Similar | | Imaging output | CVBS&DVI&S-VIDEO | HDMI | Different | | Light Source | LED 2W | LED 2W | Same | | Lamp Life | 30,000 hrs | 30,000 hrs | Same | | Screen | 10.1 inches | N/A | Different | | USB Flash<br>Memory | USB 2.0(Max:128G;FAT32) | USB 2.0(Max:128G;FAT32) | Same | {7}------------------------------------------------ K241532 Page 4 of 5 | Image: SEEEEN 视介光电 | | | | | |--------------------|--|--|--|--| | vision borderless | | | | | | illumination | ≥2000LUX | ≥2000LUX | Same | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Resolution | d = 10mm, the MTF50 value of the SFR at the center part shall not be lower than 200lw / pH, and the MTF50 value at 70% of the field of view position shall not be lower than the MTF50 value at 80% of the center position. | d = 10mm, the MTF50 value of the SFR at the center part shall not be lower than 200lw / pH, and the MTF50 value at 70% of the field of view position shall not be lower than the MTF50 value at 80% of the center position. | Same | | Filed of<br>View | 110°±10% | 110°±10% | Same | | Observe<br>the depth<br>of field | 5-50mm<br>MTF50 ≥ 200 (LW / PH) | 5-50mm<br>MTF50 ≥ 200 (LW / PH) | Same | | Direction of View | 0°±2° | 0°±2° | Same | | TV<br>distortion | < 15% | < 15% | Same | | SNR and<br>Dynamic<br>Range | SNR≥10<br>DR≥90 | SNR≥10<br>DR≥90 | Same | | Color<br>reducibility<br>y test | ΔE≤23<br>ΔC≤23 | ΔE≤23<br>ΔC≤23 | Same | | Shading | Lightness non-uniformity DL≥70<br>Luminance non-uniformity Dγ≥95 | Lightness non-uniformity DL≥70<br>Luminance non-uniformity Dγ≥95 | Same | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for SEEGEN, a company specializing in vision technology. The logo features the company name in a stylized, bold font, with the tagline "vision borderless" underneath. To the right of the company name are Chinese characters, likely representing the company's name or a related concept in Chinese. The overall design is modern and sleek, reflecting the company's focus on innovation and cutting-edge technology in the field of vision. #### 5.7 Substantial Equivalence The new image processor,(PL-2100), are used as predicate device compared to proposed device Flexible Video-Choledochoscope System manufactured by Shanghai SeeGen Photoelectric Technology Co., Ltd. #### 5.8 Non-clinical Performance Data The Flexible Video-Choledo-Cysto-Ureteroscope System has been successfully tested for its functions. Safety testing was performed including electrical safety IEC 60601-1 and IEC 60601-2-18, electromagnetic compatibility per IEC 60601-1-2 and biocompatibility of the patient contact materials per ISO 10993. Additional validations were conducted for the sterilization process, EO residual, transportation and Photobiological safety, optical and colour performance test. #### 5.9 Clinical Test Data No Clinical Study is included in this submission. #### 5.10 Conclusion In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Shanghai SeeGen Photoelectric Technology Co., Ltd. has demonstrated that proposed device Flexible Video-Choledo-Cysto-Ureteroscope System is substantially equivalent to Boston Scientific Corporation's currently marketed Flexible Video-Choledo-Cysto-Ureteroscope System, K211686.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...