← Product Code [FEQ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ) · K161482

# EndoChoice Water Bottle Cap Irrigation System (K161482)

_Endochoice, Inc. · FEQ · Jun 27, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K161482

## Device Facts

- **Applicant:** Endochoice, Inc.
- **Product Code:** [FEQ](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ.md)
- **Decision Date:** Jun 27, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The water bottle cap irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

## Device Story

Sterile, 24-hour multi-patient use irrigation system; connects sterile water bottles to irrigation pumps and endoscopes. Components include polycarbonate cap, medical-grade PVC tubing, and stainless steel weight. Flexible pump tubing inserts into irrigation pump; rigid tubing connects to endoscope via Luer connector. System incorporates one-way check valves in tubing and endoscope connectors to prevent backflow and cross-contamination. Used in clinical settings during GI endoscopy; operated by medical staff. Provides irrigation to clear field of view; enhances visualization for physician during procedure.

## Clinical Evidence

Bench testing only. Included functional performance testing post-aging and biocompatibility testing in conformance with ISO 10993-1. All tests passed.

## Technological Characteristics

Materials: Polycarbonate (PC 110, Makrolon 2458), PVC (APEX, 3304A), Stainless Steel (Grade 304). Components: Cap, tubing, weight, one-way check valves. Sterilization: Ethylene Oxide. Connectivity: Mechanical interface to standard 33mm water bottles and irrigation pumps. Form factor: Disposable tubing set.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EndoChoice Water Bottle Cap Irrigation System ([K133747](/device/K133747.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2016

EndoChoice, Inc. Daniel Hoefer Regulatory Affairs Manager 11810 Wills Rd. Alpharetta, GA 30009

Re: K161482

Trade/Device Name: EndoChoice Water Bottle Cap Irrigation System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FEQ Dated: May 26, 2016 Received: May 31, 2016

Dear Daniel Hoefer,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K161482

#### Device Name

EndoChoice Water Bottle Cap Irrigation System

Indications for Use (Describe)

The water bottle cap irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water supply during GI endoscopic procedures with an irrigation pump.

| Type of Use (Select one or both, as applicable)                                |
|--------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 201.2 and Part)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 201.6</label>     |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# EndoChoice Water Bottle Cap Irrigation System

### 1. Company Identification

#### Applicant:

EndoChoice, Inc. 11810 Wills Road Alpharetta, GA 30009 Establishment Registration: 3007591333

### 2. Contact Person

Daniel Hoefer Sr. Regulatory Affairs Manager EndoChoice, Inc.

#### 3. Device Name

| Trade name:          | EndoChoice Water Bottle Cap Irrigation System |
|----------------------|-----------------------------------------------|
| Common/Usual Name:   | Water Bottle Cap Irrigation System            |
| Classification name: | FEQ; Endoscope and Accessories                |

### 4. Device Classification

| Classification | Endoscope and accessories, 21CFR 876.1500 |
|----------------|-------------------------------------------|
| Product Code:  | FEQ                                       |
| Committee:     | Gastroenterology/Urology                  |

### 5. Intended Use

The water bottle cap irrigation system (tubing and accessories to accommodate various endoscopes and irrigation pumps) is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

### 6. Device Description

The EndoChoice Water Bottle Cap Irrigation System is a sterile, consumable, 24-hour multi-patient use device designed to fit commercially available sterile water bottles and irrigation pumps for providing sterile water during endoscopic procedures.

The EndoChoice water bottle cap system is comprised of a polycarbonate standard water bottle cap, medical grade PVC tubing that extends from the endoscope connector into the water bottle, and a medical grade stainless steel cylinder that serves 

{4}------------------------------------------------

as both the water intake portion of the disposable water bottle cap and a weight to hold the tubing down while in the water bottle.

The bottle tubing is added into the sterile water bottle and the irrigation system is fastened via the tubing system cap.

Flexible pump tubing connected to the top of the water bottle cap runs to the tubing connector reducer. This flexible tubing is the part of the device that is placed into the irrigation pump. A tubing connector reducer joins the flexible tubing and rigid tubing. The rigid tubing runs to the tubing Luer connector. The tubing Luer connector has the first one-way check valve which prevents potential backflow of fluid from the endoscope into the tubing set. The separately packaged, single-use endoscope connectors (used to connect the tubing set to the auxiliary water channel) also contain one-way valves used to prevent potential backflow.

As with the predicate, the separately packaged, single-use connectors are used to connect to different endoscope models, and all include the second one-way valve to prevent contamination due to potential backflow.

Between patients, the endoscope connectors must be changed to ensure that there is no risk of cross contamination. EndoChoice water bottle irrigation tubing fits standard threaded lid disposable sterile water bottles of approximately 33 mm diameter.

The water bottle cap system is supplied sterile and can be used up to 24 hours.

### 7. Substantial Equivalence

The modified EndoChoice Water Bottle Cap Irrigation System is equivalent to the legally marketed unmodified EndoChoice Water Bottle Cap Irrigation System (K133747). A feature comparison of the two devices is shown below.

Based on the intended use, technological characteristics and overall performance of the devices, EndoChoice, Inc. believes that the modifications to the EndoChoice Water Bottle Cap Irrigation System leave the device equivalent to the unmodified predicate and that the differences between the devices do not raise new issues of safety or effectiveness.

{5}------------------------------------------------

|                                                                    | Comparison Table                                                                                                                                                                                                                                                                               |                                                                                                                      |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
|                                                                    | EndoChoice Water Bottle Cap<br>Irrigation System<br>(Unmodified)                                                                                                                                                                                                                               | EndoChoice Water Bottle<br>Cap Irrigation System<br>(Modified)                                                       |
| 510(k) number                                                      | K133747                                                                                                                                                                                                                                                                                        | Pending                                                                                                              |
| Compatibility with currently available<br>endoscopes               | Olympus 140, 160, 180,<br>Endoscopes                                                                                                                                                                                                                                                           | Olympus 140, 160, 180,<br>And Fuse Endoscopes                                                                        |
| Sterilization Method                                               | Ethylene Oxide                                                                                                                                                                                                                                                                                 | Unchanged                                                                                                            |
| Compatibility with commercially available<br>sterile water bottles | Compatible with Standard 33 mm<br>water bottle caps                                                                                                                                                                                                                                            | Unchanged                                                                                                            |
| Compatibility with Olympus OFP-2<br>irrigation pumps               | N/A                                                                                                                                                                                                                                                                                            | Compatible                                                                                                           |
| Indications for use statement                                      | The water bottle cap irrigation<br>system (tubing and accessories<br>to accommodate various<br>endoscopes and irrigation<br>pumps) is intended to provide<br>irrigation via sterile water<br>supply during GI endoscopic<br>procedures when used in<br>conjunction with an irrigation<br>pump. | Unchanged                                                                                                            |
| Contraindications                                                  | Those specific to any endoscopic<br>procedure                                                                                                                                                                                                                                                  | Unchanged                                                                                                            |
| Materials (patient contacting)                                     |                                                                                                                                                                                                                                                                                                |                                                                                                                      |
| Flexible Pump Tubing                                               | PVC - Flexchem                                                                                                                                                                                                                                                                                 | PVC - APEX, change in<br>PVC formulation, equivalent<br>to previous formulation                                      |
| Flexible Pump Tubing dimensions                                    | Outer diameter 6.4 mm, inner<br>diameter 3.2mm                                                                                                                                                                                                                                                 | Outer diameter 9.7 mm, inner<br>diameter 6.5 mm                                                                      |
| Compression Luer Connector / Reducer                               | Polycarbonate, Makrolon 2558                                                                                                                                                                                                                                                                   | Polycarbonate, PC 110,<br>equivalent to previous<br>formulation                                                      |
| Rigid Tubing                                                       | PVC, MD80-GS-PVC                                                                                                                                                                                                                                                                               | Unchanged                                                                                                            |
| Tubing Connector Reducer                                           | Polycarbonate, Makrolon 2558                                                                                                                                                                                                                                                                   | Polycarbonate, Makrolon<br>2458, change in formulation,<br>equivalent to previous<br>formulation                     |
| Bottle Tubing                                                      | PVC 5420                                                                                                                                                                                                                                                                                       | PVC 3304A, change in PVC<br>formulation, equivalent to<br>previous formulation                                       |
| Bottle Tubing Weight                                               | Stainless Steel Grade 303                                                                                                                                                                                                                                                                      | Stainless Steel Grade 304,<br>change in stainless steel<br>formulation, equivalent to<br>previous formulation        |
| Male Luer Base                                                     | PC110                                                                                                                                                                                                                                                                                          | Polycarbonate Makrolon<br>2458, change in formulation,<br>equivalent to previous<br>formulation                      |
| Female Luer Connector                                              | PC110                                                                                                                                                                                                                                                                                          | Unchanged                                                                                                            |
|                                                                    | EndoChoice Water Bottle Cap Irrigation System (Unmodified)                                                                                                                                                                                                                                     | EndoChoice Water Bottle Cap Irrigation System (Modified)                                                             |
| Materials (no patient contact) / Use                               |                                                                                                                                                                                                                                                                                                |                                                                                                                      |
| Packaging                                                          | Individually packaged in Tyvek Peel Pouch                                                                                                                                                                                                                                                      | Unchanged                                                                                                            |
| Tyvek package                                                      | Dimension: 320 mm x 150 mm, part# 1059B                                                                                                                                                                                                                                                        | Dimension: 340 mm x 240 mm, part# 1073B, dimensional change and Tyvek type change are equivalent to previous package |
| Use                                                                | Disposable, 24-hour multi-patient use                                                                                                                                                                                                                                                          | Unchanged                                                                                                            |

{6}------------------------------------------------

#### 8. Non-clinical testing

Non-clinical testing has been performed on the device. Specifically, the following has been completed on the accessory water bottle, cap, and tubing:

- Benchtop functional performance testing, post-aging ●
- Biocompatibility testing in conformance with ISO 10993-1 ●

All test results passed, demonstrating that the device is safe and effective in comparison with the predicate device.

#### 9. Conclusion

The modified EndoChoice Water Bottle Cap Irrigation System is equivalent to the legally marketed predicate device. It is the same or equivalent in terms of design, intended use, materials, and labeling.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K161482](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FEQ/K161482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
