ELLIPSE 4
Device Facts
| Record ID | K964522 |
|---|---|
| Device Name | ELLIPSE 4 |
| Applicant | Andromeda Medizinische Systeme GmbH |
| Product Code | FEN · Gastroenterology, Urology |
| Decision Date | Mar 21, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1620 |
| Device Class | Class 2 |
Indications for Use
The Ellipse 4™ is intended to assist in the diagnosis of incontinence and/or voiding disorders.
Device Story
Ellipse 4™ is a compact urodynamic system for clinical use; registers pressure, uroflow, and EMG signals. Inputs: microtip catheters (MTCs) for pressure; surface electrodes for EMG. Signals undergo A/D conversion; displayed on LCD; printed via thermal printer; stored in memory. Data transferable to PC via network. Assists clinicians in diagnosing incontinence and voiding disorders by providing objective physiological measurements; enables comprehensive urodynamic investigations.
Clinical Evidence
Bench testing confirmed accuracy of transducers, pump, EMG, pressure, pressure flow, and urethral pressure profile. Clinical experience includes over 10,000 urodynamic investigations with no reported complications.
Technological Characteristics
Coated aluminum housing; microtip catheters (MTCs) for pressure sensing; surface electrodes for EMG; A/D signal conversion; LCD display; thermal printer; network connectivity for PC download; non-energy delivering.
Indications for Use
Indicated for patients requiring urodynamic evaluation, including flowmetry, cystometry, pressure flow, urethral pressure profile, and EMG testing for incontinence or voiding disorders.
Regulatory Classification
Identification
A urodynamics measurement system is a device used to measure volume and pressure in the urinary bladder when it is filled through a catheter with carbon dioxide or water. The device controls the supply of carbon dioxide or water and may also record the electrical activity of the muscles associated with urination. The device system may include transducers, electronic signal conditioning and display equipment, a catheter withdrawal device to enable a urethral pressure profile to be obtained, and special catheters for urethral profilometry and electrodes for electromyography. This generic type of device includes the cystometric gas (carbon dioxide) device, the cystometric hydrualic device, and the electrical recording cystometer, but excludes any device that uses air to fill the bladder.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Related Devices
- K960503 — DANTEC DUET · Dantec Medical, Inc. · Sep 16, 1996
- K212830 — CT3000Pro · Srs Medical · Dec 21, 2021
- K983325 — IN-PROBE II URODYNAMIC SYSTEM · Influence, Inc. · Nov 24, 1998
- K243052 — Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)) · Bright Uro, Inc. · Mar 24, 2025
- K992715 — NEURO URODYNAMIC SUITE, MODEL 9032E0101 · Medtronic Functional Diagnostics A/S · Feb 28, 2000