Fusion Cytology Brush, CytoMax II Double Lumen Cytology Brush

K181317 · Wilson-Cook Medical, Inc. · FDX · Feb 8, 2019 · Gastroenterology, Urology

Device Facts

Record IDK181317
Device NameFusion Cytology Brush, CytoMax II Double Lumen Cytology Brush
ApplicantWilson-Cook Medical, Inc.
Product CodeFDX · Gastroenterology, Urology
Decision DateFeb 8, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Used for collection of cells in the biliary system.

Device Story

Endoscopic cytology brush used for biliary cell collection; device passed through endoscope over prepositioned wire guide to target location. Consists of nylon brush on stainless-steel drive wire with stainless-steel tip, double lumen catheter with radiopaque markers, wire guide access hub, and pinvise handle. Physician operates handle by pushing forward to extend brush and pulling backward to retract. Enables tissue sampling for diagnostic purposes; benefits patient by facilitating minimally invasive biliary cytology.

Clinical Evidence

Bench testing only. Testing included shelf life, packaging validation, tensile strength (device and handle), force to retract brush, and wire guide compatibility. Biocompatibility testing performed per ISO 10993-1 for surface devices with limited mucosal membrane contact.

Technological Characteristics

Nylon brush, stainless-steel drive wire, stainless-steel tip, double lumen catheter. Features radiopaque markers, wire guide access hub, detachable extension line with luer lock, and pinvise handle. Mechanical actuation. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for the collection of cells in the biliary system.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. February 8, 2019 Wilson-Cook Medical, Inc. /Cook Endoscopy Theresa de Prat Regulatory Affairs Specialist II 4900 Bethania Station Road Winston-Salem, NC 27105 Re: K181317 Trade/Device Name: Fusion Cytology Brush, CytoMax II Double Lumen Cytology Brush Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: FDX Dated: January 10, 2019 Received: January 11, 2019 Dear Theresa de Prat: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ 2 statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Angel A. Solergarcia -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181317 Device Name Fusion Cytology Brush CytoMax II Double Lumen Cytology Brush Indications for Use (Describe) Used for collection of cells in the biliary system. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 009. 510k Summary {4}------------------------------------------------ Page 1 of 3 #### 510(k) Summary | Name: | Wilson-Cook Medical, Inc. /Cook Endoscopy | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------| | Address: | 4900 Bethania Station Road<br>Winston-Salem, North Carolina 27105 | | Phone: | (336) 744-0157 | | Fax: | (336) 201-5994 | | Contact: | Theresa de Prat, Regulatory Affairs Specialist II | | Date: | May 18, 2018 | | Trade Names: | Fusion® Cytology Brush<br>CytoMax II® Double Lumen Cytology Brush | | Common Name: | Endoscopic Cytology Brush | | Classification Name: | Endoscope and accessories 21 CFR §876.1500, FDX, Class II | | Predicate Devices: | Wilson-Cook CytoMax II® Double Lumen Cytology Brush<br>Wilson-Cook Fusion® Cytology Brushes cleared on December 29, 2017 under K171573. | Used for collection of cells in the biliary system. Intended Use: ### Device Description: The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brush (subject devices) represent modifications made to the Wilson-Cook Double Lumen Biliary Cytology Brush and Fusion® Cytology Brush (predicate devices) cleared to market via 510(k) K171573 by Wilson-Cook Medical, Inc. The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brushes consist of a nylon brush on a stainless-steel drive wire with a stainless-steel tip, a double lumen catheter with ink markings, radiopaque markers, wire guide access via a wire guide hub or an IDE port, a {5}------------------------------------------------ detachable extension line with a leur lock for optional flushing of the wire guide hub and a pinvise handle. The Fusion® Cytology Brush and CytoMax II® Double Lumen Cytology Brushes are used by passing the device through an endoscope over a prepositioned wire guide to a target location. The cytology brush is located at the distal end (patient contacting) of the device with the pin vise handle located at the proximal end (non-patient contacting). The handle is actuated by pushing the pin vise handle forward to extend the cytology brush and then pulling backward to retract the brush. ## Substantial Equivalence: Modifications were made to the currently cleared Wilson-Cook CytoMax II® Double Lumen Cytology Brush and Wilson-Cook Fusion® Cytology Brushes (K171573). These changes include: catheter diameter and length, wire guide size compatibility changes including wire guide access port and recommended endoscope channel sizes. ## Performance Data: Performance testing consisting of non-clinical bench testing that demonstrates the subject devices met the performance requirements to fulfill their intended use. This testing testing provides reasonable assurance that the subject devices will function as intended. The subject devices do not raise new questions of safety or effectiveness as compared to the predicate devices. ## Summary of non-clinical testing: The following non-clinical testing was conducted to demonstrate the performance of the subject devices and confirms that the subject devices perform as intended. - . Shelf Life Testing - Packaging Validation . - . Tensile Strength Testing - . Tensile Strength of Handle - Force to Retract Brush . {6}------------------------------------------------ K181317 Page 3 of 3 - . Wire Guide Compatibility Biocompatibility testing was performed in accordance with FDA guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and ISO 10993-1:2009 Biological evaluation of medical device: Evaluation and testing within a risk management process for surface devices with limited mucosal membrane contact. ### Conclusion: We believe that the subject devices are substantially equivalent to the predicate devices with respect to intended use, key operating principles, materials of construction, and technological characteristics. We consider the risks associated with the modifications to the subject devices to have been adequately addressed through our Design Control processes and do not affect the safety or effectiveness of the devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...