← Product Code [FDS](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS) · K242779

# Hood (DH-083ST) (K242779)

_Fujifilm Corporation · FDS · Oct 11, 2024 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K242779

## Device Facts

- **Applicant:** Fujifilm Corporation
- **Product Code:** [FDS](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS.md)
- **Decision Date:** Oct 11, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

This hood is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

## Device Story

Hood (DH-083ST) is a mechanical accessory for flexible endoscopes; attaches to distal end of endoscope; secured with sterile medical tape. Device maintains field of view during endoscopic procedures like mucosal resection. Operated by physicians in clinical settings. Device inserted through mouth; removed after procedure; single-use; disposed of per local regulations. Benefits include improved visualization during endoscopic interventions.

## Clinical Evidence

Bench testing only. Performance testing conducted to verify dimensional specifications (outer diameter, inner diameter, total length, distance from tip) and compatibility with endoscopes per ISO 8600-1:2015. All test objectives passed.

## Technological Characteristics

Mechanical accessory; dimensions vary from predicate; compatible with specific endoscopes listed in operation manuals; no change in materials; single-use; non-sterile (implied by use of sterile medical tape for attachment).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Hood models DH-096ST ([K232314](/device/K232314.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

October 11, 2024

Fujifilm Corporation % Chaitrali Kulkarni Sr. Regulatory Affairs Specialist Fujifilm Healthcare Americas Corporation 81 Hartwell Ave. Suite 300 Lexington, Massachusetts 02421

Re: K242779

Trade/Device Name: Hood (DH-083ST) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FDS Dated: September 13, 2024 Received: September 13, 2024

Dear Chaitrali Kulkarni:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242779

Device Name

Hood (DH-083ST)

Indications for Use (Describe)

This hood is intended to be used in combination with compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

Type of Use (Select one or both, as applicable)

> | Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

|                                    |                                                          |                                                                        | 510(k) Summary                                              |              |                     | Prepared on: 2024-09-13 |                     |
|------------------------------------|----------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------------|--------------|---------------------|-------------------------|---------------------|
| Contact Details                    |                                                          |                                                                        |                                                             |              |                     |                         | 21 CFR 807.92(a)(1) |
| Applicant Name                     |                                                          | Fujifilm Corporation                                                   |                                                             |              |                     |                         |                     |
| Applicant Address                  |                                                          | 798 MIYANODAI KAISEI-MACHI ASHIGARAKAMI-GUN KANAGAWA<br>258-8538 Japan |                                                             |              |                     |                         |                     |
| Applicant Contact Telephone        |                                                          | 704-517-4886                                                           |                                                             |              |                     |                         |                     |
| Applicant Contact                  |                                                          | Ms. Chaitrali Kulkarni                                                 |                                                             |              |                     |                         |                     |
| Applicant Contact Email            |                                                          | hcusregulatoryaffairs@fujifilm.com                                     |                                                             |              |                     |                         |                     |
| Correspondent Name                 |                                                          |                                                                        | Fujifilm Healthcare Americas Corporation                    |              |                     |                         |                     |
| Correspondent Address              |                                                          |                                                                        | 81 Hartwell Ave. Suite 300 Lexington MA 02421 United States |              |                     |                         |                     |
| Correspondent Contact Telephone    |                                                          |                                                                        | 704-517-4886                                                |              |                     |                         |                     |
| Correspondent Contact              |                                                          | Ms. Chaitrali Kulkarni                                                 |                                                             |              |                     |                         |                     |
| Correspondent Contact Email        |                                                          | hcusregulatoryaffairs@fujifilm.com                                     |                                                             |              |                     |                         |                     |
| Device Name                        |                                                          |                                                                        |                                                             |              |                     |                         | 21 CFR 807.92(a)(2) |
| Device Trade Name                  |                                                          | Hood (DH-083ST)                                                        |                                                             |              |                     |                         |                     |
| Common Name                        |                                                          | Endoscope and accessories                                              |                                                             |              |                     |                         |                     |
| Classification Name                |                                                          | Gastroscope And Accessories, Flexible/Rigid                            |                                                             |              |                     |                         |                     |
| Regulation Number                  |                                                          | 876.1500                                                               |                                                             |              |                     |                         |                     |
| Product Code(s)                    |                                                          | FDS                                                                    |                                                             |              |                     |                         |                     |
| Legally Marketed Predicate Devices |                                                          |                                                                        |                                                             |              |                     |                         | 21 CFR 807.92(a)(3) |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) |                                                                        |                                                             | Product Code |                     |                         |                     |
| K232314                            | Hood models DH-096ST                                     |                                                                        |                                                             | FDS          |                     |                         |                     |
| Device Description Summary         |                                                          |                                                                        |                                                             |              | 21 CFR 807.92(a)(4) |                         |                     |

### Device Description Summary

A.Intended Use

This hood is intended to be used in compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

B. Principles of Operation

Align the objective lens of endoscope with the drain of the hood and attach the hood to the distal end of endoscope by pressing the hood until it stops. Securely attach the endoscope with sterile medical tape.

Insert the endoscope equipped with the hood through the mouth. Perform the intended treatment. After completion of the examination, slowly withdraw the endoscope along with the attached hood. Peel the tape off the endoscope completely and remove

{5}------------------------------------------------

K242779 Page 2 of 2

the hood from the endoscope. Dispose of the hood and tape according to local laws and requlations.

## Intended Use/Indications for Use

This hood is intended to be used in compatible endoscopes to maintain the field of view during endoscopic procedures such as mucosal resection.

## Indications for Use Comparison

The intended use of the predicate model and the proposed model is the same.

## Technological Comparison

The differences in the sizes of the proposed hood model and the predicate hood model does not affect the safety of the proposed model because the hood has been tested for its compatible endoscopes and the compatible endoscope is listed on the Operation Manuals. There is no change to the materials of the hoods.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The changes to the proposed device model are in the dimeter, maximum diameter of attaching endoscope, distance from the tip, total length and combination endoscope. To determine if the changes would affect the safety of the subject device, performance testing was conducted, and the passed all test objectives. It was determined that the new hood model is substantially equivalent to the predicate devices.

Endoscope specific testing was conducted according to ISO 8600-1: 2015

The proposed device met performance specifications in the following additional testing:

- · Outer diameter: Measure the diameter of Maximum diameter portion by a scale.
- · Maximum diameter of attaching endoscope: Measure the diameter of Maximum diameter portion by a scale.
- · Total length: Measure the total length by a scale.
- · Inner diameter of distal end: Measure the inner diameter of distal end by a scale.
- · Distance from the tip: Measure the distance from the tip by a scale.

### 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

### 21 CFR 807.92(a)(6)

21 CFR 807.92(a)(5)

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K242779](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDS/K242779)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
