G-EYE System

K220158 · Smart Medical Systems , Ltd. · FDF · Apr 11, 2022 · Gastroenterology, Urology

Device Facts

Record IDK220158
Device NameG-EYE System
ApplicantSmart Medical Systems , Ltd.
Product CodeFDF · Gastroenterology, Urology
Decision DateApr 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Intended Use

The G-EYE® Colonoscope is intended to be used with a compatible video processor (including light source), documentation equipment, monitor, endotherapy device such as a biopsy forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower gastrointestinal tract including the anus, rectum, sigmoid colon, colon and ileocecal valve. It incorporates a balloon at the distal end of the colonoscope to maintain central positioning within the lumen and help control the endoscope's view field and/or endoscope's positioning.

Device Story

G-EYE System consists of a remanufactured standard colonoscope with an integrated, reprocessable balloon and the NaviAid SPARKC inflation pump. Endoscopist advances colonoscope with balloon deflated; at the cecum, balloon is inflated to engage colon walls. During withdrawal, inflated balloon mechanically straightens colonic folds and flexures, centers optics within lumen, and minimizes uncontrolled bowel slippage. System provides improved visualization of colonic mucosa. Used in clinical settings by endoscopists. Output is visual endoscopic imagery viewed on a monitor. Benefits include enhanced mucosal visualization and controlled endoscope positioning.

Clinical Evidence

Five clinical studies involving approximately 4,500 subjects support safety and performance. Studies demonstrated the device functioned as intended with no serious device-related adverse events. Results confirm mechanical straightening of colonic flexures facilitates visualization of the colonic mucosa surface area.

Technological Characteristics

System includes a reprocessable balloon integrated into a remanufactured standard colonoscope (11.0mm-13.2mm diameter) and a dedicated SPARKC inflation pump. Biocompatibility per ISO 10993-1. Electrical safety/EMC per IEC 60601-1 and IEC 60601-1-2. Reprocessing follows original manufacturer instructions with supplemental balloon-specific protocols.

Indications for Use

Indicated for use in endoscopy and endoscopic surgery within the lower gastrointestinal tract (anus, rectum, sigmoid colon, colon, ileocecal valve) to maintain central positioning and control view field/positioning via an integrated balloon.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 11, 2022 Smart Medical Systems Ltd. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, PA 19103 Re: K220158 Trade/Device Name: G-EYE System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDF Dated: January 19, 2022 Received: January 19, 2022 Dear Janice Hogan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K220158 Device Name G-EYE® System Indications for Use (Describe) The G-EYE® colonoscope is intended to be used with a compatible video processor (including light source), documentation equipment, monitor, endotherapy device such as a biopsy forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower gastrointestinal tract including the anus, rectum, sigmoid colon, colon and ileocecal valve. It incorporates a balloon at the colonoscope to maintain central positioning within the lumen and help control the endoscope's view field and/or endoscope positioning. Type of Use (Select one or both, as applicable) ال Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY # SMART Medical Systems Ltd.'s G-EYE® System # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Adva Yoselzon VP of Regulatory & Clinical Affairs Smart Medical Systems Ltd. 5 Hanofar, Raanana 4366404 Tel: +972-9-7444321 Fax: +972-9-7444543 Mobile: +972-52-8911112 Email: ayoselzon@smartmedsys.com Date Prepared: January 19, 2021 #### Name of Device G-EYE® System #### Common or Usual Name Endoscope and accessories, Gastroenterology-Urology ### Classification 21 CFR 876.1500, Class II, product code FDF # Predicate Devices G-EYE® System (Predicate Device K202469) G-EYE® System (Reference Device K192588) # Intended Use / Indications for Use The G-EYE® Colonoscope is intended to be used with a compatible video processor (including light source), documentation equipment, monitor, endotherapy device such as a biopsy forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower gastrointestinal tract including the anus, rectum, sigmoid colon, colon and ileocecal valve. It incorporates a balloon at the distal end of the colonoscope to maintain central positioning within the lumen and help control the endoscope's view field and/or endoscope's positioning. # Device Description The G-EYE® System comprises the G-EYE® Colonoscope and the NaviAid™ SPARKC ("SPARKC") inflation pump. The G-EYE® Colonoscope includes a reprocessable, reusable balloon (the G-EYE® balloon) that is attached to a standard, commercially available colonoscope which is remanufactured by Smart Medical to include the balloon at its bending section. The installation of the balloon onto the standard colonoscope is considered {4}------------------------------------------------ remanufacturing of the standard colonoscope by Smart Medical, which assumes responsibility for the final device including the G-EYE® balloon and the remanufactured colonoscope. The G-EYE® balloon is inflated by the dedicated SPARKC inflation system, which enables automatic control of the required balloon pressure in one of several user-selected, systemcontrolled pressure levels such that the balloon conforms to the colonic lumen diameter. The G-EYE® Colonoscope is advanced by the endoscopist with the balloon deflated, which essentially preserves the standard colonoscope's diameter and insertion technique. At the cecum, the balloon is inflated to engage the colon walls, and the colonoscope is then withdrawn by the endoscopist using standard withdrawal technique. Withdrawal of the inflated balloon mechanically straightens the colonic folds and anatomic flexures, centers the colonoscope optics within the colonic lumen, and minimizes uncontrolled bowel slippage during withdrawal, therefore maintaining the endoscope's field of view # Technological Characteristics The G-EYE® System is nearly identical to the previously cleared version of the device ("Predicate Device") (K202469). Both the subject device and Predicate Device have the same indications for use and principles of operation and substantially similar technological characteristics and materials. The subject device and Predicate Device are both multi-use and intended to be reprocessed. Similar to the predicate, the G-EYE® Colonoscope is reprocessed in accordance with the reprocessing instructions of the original endoscope manufacturer with supplemental instructions related to the G-EYE® balloon. The current submission solely expands the range of endoscope models. This difference does not raise different questions of safety and effectiveness, as confirmed by the company's reprocessing validation testing. The materials of the G-EYE® Colonoscope are substantially similar to those of the Predicate Device, and any differences in the materials of the G-EYE® Colonoscope and the predicate device do not raise different questions of safety or effectiveness. This was confirmed through the company's biocompatibility testing and performance testing. Both the G-EYE® System and the Predicate Device feature a standard colonoscope. All compatible colonoscopes are 510(k) cleared and have an insertion tube diameter of 11.0mm - 13.2mm. The primary difference between the G-EYE® Colonoscope of the G-EYE® System and the G-EYE® colonoscope of the Predicate Device is the underlying standard colonoscope model, as noted above. As all standard colonoscopes underlying the G-EYE® Colonoscopes are 510(k) cleared colonoscopes, the minor differences among the standard colonoscopes included in the G-EYE® System and in the Predicate Device do not raise new questions of safety and effectiveness, and the difference does not adversely impact performance, as confirmed by the company's performance testing. # Performance Data In support of this 510(k) premarket notification, Smart Medical conducted nonclinical testing to evaluate the performance of the G-EYE® System for its intended use. The performance testing conducted includes the following: - . Reprocessing validation testing consistent with FDA's quidance in Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. {5}------------------------------------------------ - Biocompatibility of patient-contacting components has been established in accordance with ISO 10993-1. - Software verification and validation was performed, and demonstrated that the software performs as intended. - . Electrical safety and electromagnetic compatibility testing was conducted and results were passing in accordance with IEC 60601-1 and IEC 60601-1-2. - . Verification bench testing of the SPARKC and G-EYE® Colonoscope which included operational and functional testing. Each of these tests met its acceptance criteria, supporting substantial equivalence to the predicate. Although not required to demonstrate substantial equivalence, clinical data is available from five clinical studies that further support the safety and performance of the device for its intended use. In total, the studies included approximately 4500 subjects. All five studies demonstrated that the device functioned as intended, with no serious device-related adverse events. In addition, the studies demonstrated that the device served its intended function, and that the mechanical straightening of the colonic flexures provided by the G-EYE® System facilitates visualization of the colonic mucosa surface area. # Substantial Equivalence The G-EYE® System and its Predicate Device have the same intended use, same indications, same principles of operation, and similar technological characteristics. The minor technological differences do not present different questions of safety or effectiveness than the Predicate Device, and accepted scientific methods exist to evaluate the performance of the G-EYE® System compared to its Predicate Device. Performance data has demonstrated that the G-EYE® System is as safe and effective as its predicate. Thus, the G-EYE® System is substantially equivalent to the previously cleared G-EYE® System (K202469).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...