← Product Code [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF) · K213773

# Insufflation Retention Device (K213773)

_Bpendo, LLC · FDF · Jul 21, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K213773

## Device Facts

- **Applicant:** Bpendo, LLC
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Jul 21, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Insufflation Retention Device (IRD) is an accessory to an endoscope. The IRD is a single balloon accessory intended for use with any standard endoscope that has a distal tip outer diameter of 9.2-13.2 mm. The device is indicated to assist with optical visualization in the large intestine during endoscopic procedures.

## Device Story

The Insufflation Retention Device (IRD) is a non-sterile, single-use colonoscope accessory designed to assist with optical visualization in the large intestine. The device features a split section for placement over a colonoscope shaft and includes a single urethane balloon. In a professional healthcare setting, a clinician manually inflates the balloon using a non-sterile syringe after positioning the device at the anus. The inflated balloon creates an adjustable seal that retains insufflation air introduced during the colonoscopy, thereby improving visibility within the large intestine. The device allows for free movement and articulation of the colonoscope without compromising its function. By maintaining insufflation, the IRD facilitates easier advancement of the colonoscope and enhances the clinician's ability to visualize the intestinal mucosa, potentially improving diagnostic outcomes.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (cytotoxicity, sensitization, irritation/reactivity per ISO 10993-1) and performance testing (device placement, insertion force, pressure leak, scope diameter/articulation/movement, balloon diameter/burst, device security, shelf life, package integrity, and glove usability).

## Technological Characteristics

Constructed from polymeric materials (urethane balloon). Single-balloon design for manual inflation via syringe. Compatible with endoscopes having 9.2-13.2 mm distal tip diameters. Non-sterile. Mechanical operation; no energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- DiLumen Endolumenal Interventional Platform ([K162428](/device/K162428.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 21, 2022

Bpendo, LLC % James Fentress Director, R&D and Regulatory Policy Gilero, LLC 635 Davis Drive STE 100 Morrisville, NC 27560

Re: K213773

> Trade/Device Name: Insufflation Retention Device Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FDF Dated: June 14, 2022 Received: June 21, 2022

Dear James Fentress:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

{1}------------------------------------------------

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213773

Device Name Insufflation Retention Device

Indications for Use (Describe)

The Insufflation Retention Device (IRD) is an accessory to an endoscope. The IRD is a single balloon accessory intended for use with any standard endoscope that has a distal tip outer diameter of 9.2-13.2 mm. The device is indicated to assist with optical visualization in the large intestine during endoscopic procedures.

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">_</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for BPENDO. The logo consists of a stylized hourglass symbol on the left, followed by the text "BPENDO" in a bold, sans-serif font. Below the text, there is a smaller line of text that reads "Insufflation Retention Device."

K213773 Page 1 of 5 510(k) Notification BPENDO

Insufflation Retention Device (IRD)

# 510(k) Summary

| Company Name:    |
|------------------|
| Company Address: |

Company Phone:

Official Contact: Phone: E-mail:

BPENDO, LLC 216 Foreman Circle Norman, OK 73069 +1 (405) 974-0776

Jim Fentress +1 (919) 595-8236 jfentress@gilero.com

Submission Date:

November 18, 2021

### Device Identification:

Trade Name: Common Name: Device Class: Regulation Number: Regulation Name: Product Code: Review Panel:

Insufflation Retention Device (IRD) Colonoscope Accessory 2 876.1500 Endoscope and Accessories FDF Gastroenterology/Urology

## Predicate Device:

| Manufacturer: | LUMENDI, LLC                                |
|---------------|---------------------------------------------|
| Trade Name:   | DiLumen Endolumenal Interventional Platform |
| 510(k):       | K162428                                     |

## Device Description:

The IRD is a non-sterile accessory for a colonoscope which provides an adjustable seal between a colonoscope and a patient's anus. This seal retains introduced insufflation air introduced during a colonoscopy. The retained insufflation air provides increased visibility and facilitates the advancement of the colonoscope. The IRD contains a split section which allows it to be placed over the continuous shaft of the colonoscope. The adjustable seal is inflated and deflated manually with a non-sterile syringe.

### Indications for Use:

The Insufflation Retention Device (IRD) is an accessory to an endoscope. The IRD is a single balloon accessory intended for use with any standard endoscope that has a distal tip outer diameter of 9.2-13.2 mm. The device is indicated to assist with optical visualization in the large intestine during endoscopic procedures.

### Technological Characteristics and Substantial Equivalence:

The following chart presents an overview of comparisons between the subject device (IRD) and the predicate device (DiLumen Endolumenal Interventional Platform):

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for BPENDO. The logo is in blue and consists of a stylized hourglass shape on the left, followed by the text "BPENDO" in a sans-serif font. Below the text "BPENDO" is the text "Insufflation Retention Device" in a smaller font.

510(k) Notification

| Device Attribute                        | SUBJECT: [BPENDO]<br>Insufflation Retention Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | PREDICATE: [Lumendi]<br>DiLumen Endolumenal Interventional<br>Platform                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class                            | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Classification<br>Name           | Colonoscope and Accessories,<br>Flexible/rigid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Colonoscope and Accessories,<br>Flexible/rigid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulation Number                       | 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code                            | FDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | FDF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use and<br>Intended Use | The Insufflation Retention Device (IRD)<br>is an accessory to an endoscope. The IRD<br>is a single balloon accessory intended for<br>use with any standard endoscope that has<br>a distal tip outer diameter of 9.2-13.2 mm.<br>The device is indicated to assist with<br>optical visualization in the large intestine<br>during endoscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                  | The Lumendi DiLumen is an accessory to<br>an endoscope.  The DiLumen dual balloon<br>accessory is intended for use with any<br>standard endoscope that has a distal tip<br>outer diameter of 12.5-14.3 mm and a<br>working length of 1680mm or greater.<br>The device is indicated to ensure<br>complete positioning of an endoscope in<br>the large intestine, and assist with optical<br>visualization, diagnosis, and endoscopic<br>treatment                                                                                                                                                                                                                                                                       |
| Intended Users                          | Licensed healthcare practitioners<br>(physicians, nurses, and healthcare<br>professionals (HCPs)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Licensed healthcare practitioners                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Intended Use<br>Environment             | Professional healthcare facilities                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Professional healthcare facilities<br>(Presumed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Technology and Design                   | The IRD is constructed from polymeric<br>materials and is designed to access the<br>large intestine through the anus as a<br>colonoscope accessory placed over the<br>colonoscope.<br>The IRD contains a urethane balloon<br>which can be manually inflated and<br>deflated by the user. This balloon, as part<br>of the overall device is inserted into the<br>large intestine through the anus in a<br>deflated state<br>When temporarily inflated by the user, the<br>balloon allows increased visibility of the<br>large intestine through the retention of<br>insufflation gas introduced normally as<br>part of the colonoscopy procedure.<br>The IRD allows free movement of the<br>colonoscope, without compromising the<br>colonoscope function | The DiLumen device is constructed from<br>polymeric materials and is designed to<br>access the large intestine through the anus<br>as a colonoscope accessory placed over<br>the colonoscope.<br>The DiLumen devices contains two<br>urethane balloons which can be manually<br>inflated and deflated by the user. These<br>balloons, as part of the overall device is<br>inserted into the large intestine through<br>the anus in deflated states.<br>When temporarily inflated by the user, the<br>balloons allow increased visibility of the<br>balloon-bounded region of the large<br>intestine<br>The DiLumen device allows free<br>movement of the colonoscope, without<br>compromising the colonoscope function |
| Device Attribute                        | SUBJECT: [BPENDO]<br>Insufflation Retention Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | PREDICATE: [Lumendi]<br>DiLumen Endolumenal Interventional<br>Platform                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Principles of Operation                 | The IRD fits over a colonoscope. The device contains one deflated balloon. After the device has been positioned appropriately at the terminal end of the large intestine (anus), the balloon is manually inflated with air using a syringe. Balloon inflation allows introduced insufflation to provide greater visibility in the large intestine.                                                                                                                                                                                                                                                                                                                                                                                                       | The DiLumen device fits over a colonoscope. The device contains two deflated balloons. When positioned appropriately within the large intestines, the balloons are inflated with air using an inflation handle with a squeeze bulb to inflate the balloons within the large intestines. Balloon inflation allows greater visibility in the bounded region in the large intestine.                                                                                                                                                                                                                                                                                                                                      |
| Biocompatibility                        | Acceptable biological risk established by demonstrating that the device meets ISO 10993. See Section 15 - Biocompatibility.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Acceptable biological risk established by demonstrating that the device meets ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sterilization                           | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Prescription                            | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Rx only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for BPENDO. The logo consists of a stylized hourglass symbol on the left, followed by the text "BPENDO" in a bold, sans-serif font. Below the text, there is a smaller line of text that reads "Insufflation Retention Device". The text and hourglass symbol are all in a light blue color.

510(k) Notification

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a logo for BP ENDO Insufflation Retention Device. The logo features a stylized hourglass symbol on the left, followed by the text "BPENDO" in a bold, sans-serif font. Below the text, there is a smaller line of text that reads "Insufflation Retention Device".

K213773 Page 4 of 5

510(k) Notification BPENDO Insufflation Retention Device (IRD)

## Substantial Equivalence:

The Insufflation Retention Device is substantially equivalent to the predicate: LUMENDI. LLC DiLumen Endolumenal Interventional Platform. The subject device and the predicate device have similar indications for use and intended use. Both devices are accessories for colonoscopes which utilize manual, clinician inflation of elastomeric (polyurethane) balloons to stabilize the device in the large intestines. Both devices are used to assist the visibility for clinicians using the endoscope. Both devices accomplish this improved visibility without compromising the use of the endoscope. Both devices are single-use and non-sterile.

Comparison of the three primary technological differences between the subject IRD and the predicate DiLumen devices do not introduce new questions about the safety or efficacy of the subject device. First, while the DiLumen device utilizes two independently inflatable balloons to stabilize the endoscope compared to the single balloon used in the subject IRD device, the balloons serve similar function, and require similar evaluation to verify function. Second, the IRD does not require a balloon extension/retraction mechanism included in the DiLumen device. The omission of these features does not introduce new or different risks. Finally, the subject IRD and the predicate DiLumen devices utilized different materials for their construction, but the subject device was appropriately evaluated for biological safety consistent with its patient contact and duration of use.

### Discussion of Non-clinical Tests:

The following non-clinical tests were conducted to demonstrate substantial equivalence to the predicate device.

### Biocompatibility:

The IRD, like the predicate device, was evaluated for biocompatibility appropriate to the contact characterization (type and duration), in accordance with the requirements of ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, and the FDA Guidance for Industry - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Specific testing included:

- Cytotoxicity
- Sensitization
- Irritation or Intracutaneous Reactivity ●

#### Performance Testing:

The following performance data were provided in support of this Premarket Notification:

- Device Placement ●
- Insertion Force ●
- Pressure Leak
- Scope diameter testing ●
- Balloon diameter testing
- Scope movement
- Scope articulation ●
- Device Removal

{7}------------------------------------------------

K213773 Page 5 of 5

510(k) Notification BPENDO Insufflation Retention Device (IRD)

Image /page/7/Picture/2 description: The image is a logo for BPENDO Insufflation Retention Device. The logo features a stylized hourglass symbol to the left of the text "BPENDO" in a bold, sans-serif font. Below the company name is the text "Insufflation Retention Device" in a smaller, sans-serif font.

- Device security ●
- Device Shelf life ●
- Multiple use testing
- Package integrity
- Glove usability ●
- Use time
- Attachment force
- Balloon feedback ●
- Connector leakage
- Balloon burst ●

## Conclusion

The information in this submission supports the safety and efficacy of the subject device for its intended use and demonstrates substantial equivalence with the predicate device. The IRD's differences in external materials, technology and operation from the predicate device do not raise new questions about safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K213773](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K213773)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
