← Product Code [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF) · K181084

# PENTAX Video Colonoscopes (EC Family) (K181084)

_Pentax of America, Inc. · FDF · Jan 18, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K181084

## Device Facts

- **Applicant:** Pentax of America, Inc.
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Jan 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument is intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

## Device Story

PENTAX Video Colonoscopes (EC Family) are flexible endoscopes used for visualization and surgical intervention in the lower digestive tract. The device consists of an insertion portion (with distal end, bending section, objective lens, light guide, and instrument channel), a control body (with angulation levers, suction controls, and remote buttons), and a PVE connector. The device connects to a video processor and light source. The distal end illuminates the body cavity via a light-carrying bundle; a CCD sensor captures reflected light, transmitting image data to the processor for display on a monitor. Physicians operate the device to perform diagnostic or therapeutic procedures. The device is reusable and requires manual pre-cleaning, cleaning, and high-level disinfection or sterilization between uses. This submission specifically updates the manual pre-cleaning and cleaning instructions to improve clarity and streamline the reprocessing workflow.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing, specifically reprocessing validation data including soil accumulation and cleaning efficiency studies to support the revised manual cleaning instructions.

## Technological Characteristics

Flexible video colonoscope; reusable semi-critical device. Components: insertion portion, control body, PVE connector. Imaging: CCD sensor at distal end. Illumination: light guide bundle connected to external light source. Connectivity: electrical contacts for video processor interface. Reprocessing: manual pre-cleaning and cleaning followed by high-level disinfection or sterilization.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- PENTAX Medical Video Colonoscopes (EC Family) ([K131855](/device/K131855.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

January 18, 2019

PENTAX of America, Inc. James W. Monroe Director, Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K181084

> Trade/Device Name: PENTAX Video Colonoscopes (EC Family) Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDF Dated: December 18, 2018 Received: December 19, 2018

Dear James W. Monroe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Daniel G. Walter Jr -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K181084

Device Name

PENTAX Medical Video Colonoscopes (EC Family)

Indications for Use (Describe)

This instrument is intended to be used with a PENTAX Video Processor (including light source), documentation equipment, Monitor, Endotherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">❌</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### PENTAX Medical Video Colonoscopes (EC Family) Optimized IFU Traditional 510(k) Submission

#### 510(k) Summary

#### SUBMITTER I.

PENTAX of America, Inc., HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782 Phone: 201-571-2300 Ext 2129 Fax: 201-571-2340

Contact: James W. Monroe Date Prepared: April 14, 2018

#### II. DEVICE

The purpose of this 510(k) is to obtain clearance for revised reprocessing instructions which include optimized manual pre-cleaning and cleaning instructions for the PENTAX Medical Video Colonoscopes (EC Family), its components, and its accessories. There are no other changes to the reprocessing instructions for use, nor are there any changes to the design, intended use, or indications for use of the EC Family of endoscopes.

| Device Name          | PENTAX Video Colonoscopes (EC Family)       |
|----------------------|---------------------------------------------|
| 510(k) Number        | K131855                                     |
| Common Name          | Colonoscope and Accessories, Flexible/Rigid |
| Classification Name  | Endoscope and accessories                   |
| Regulation No.       | 876.1500                                    |
| Device Class         | 2                                           |
| Product Code         | FDF                                         |
| Classification Panel | Gastroenterology/ Urology                   |

Table 5.1: Regulatory Classification of PENTAX Medical Video Colonoscopes.

### III. PREDICATE DEVICES

The predicate device for this submission is also the subject of this submission, the PENTAX Medical Video Colonoscopes (EC Family) (K131855). The predicate device is identical to the device subject to this 510(k); the only change relates to the manual pre-cleaning and cleaning portions of the reprocessing instructions to streamline the instructions and improve clarity. No changes are being made to the high-level disinfection instructions or any other portion of the instructions for use.

{4}------------------------------------------------

#### PENTAX Medical Video Colonoscopes (EC Family) Optimized IFU Traditional 510(k) Submission

#### IV. DEVICE DESCRIPTION

The subject device is identical to the predicate device. There are no changes to the design, specifications, or technological characteristics of the PENTAX Medical Video Colonoscopes as described in K131855. The associated model numbers for the PENTAX Medical Video Colonoscopes are provided in Table 5.2.

| PENTAX Video Colonoscopes (EC Family) |           |            |            |           |
|---------------------------------------|-----------|------------|------------|-----------|
| EC-3890TLK                            | EC-3490Li | EC-3490TLi | EC-3890LK  | EC34-i10L |
| EC-2990Li                             | EC-3890Li | EC-3490LK  | EC-3890LZi | EC38-i10L |

Table 5.2: Model Numbers Associated with the PENTAX Medical Video Colonoscopes

The PENTAX Medical Video Colonoscopes (EC Family) are used with Video Processors. They are composed of three main components: an Insertion Portion, Control Body, and PVE Connector.

The Insertion Portion is inserted into the body cavity of the patient. The Insertion Portion includes the Distal End and Bending Section. The Objective Lens, Light Guide, and Instrument Channel are located on the Distal End of the Insertion Portion. The Distal End also contains an Air/Water Nozzle and Water Jet Nozzle. This Air/Water Feeding System is used to deliver the air and water to the Objective Lens.

The Control Body is held by the user's hand. The Control Body includes the Angulation Control lever used to operate the endoscope angulation, a Suction Cylinder and Suction Nipple for suctioning fluid and air in the body cavity, Remote Buttons used to operate the function of video processor, and an Instrument Channel Inlet used to insert endotherapy devices, such as biopsy forceps, into the body cavity.

The PVE Connector is connected to the Video Processor via an Electrical Contacts. The PVE Connector includes the Electrical Contacts and Light Guide Plug. The Light Guide Plug is connected to the Light Source inside the Video Processor. The Light Guide of the Distal End is used to illuminate the body cavity by light that is carried through the Light Carrying Bundle. The Light Carrying Bundle guides the light Guide Plug that is connected to the Light Source. The CCD built into the Distal End receives reflected light (image data) from the body cavity, and sends the image data to the Video Processor through the video cable. The image data are converted into the image signal by the Video Processor and the image inside the body cavity is displayed on the Monitor.

The PENTAX Medical Video Colonoscopes are reusable semi-critical devices. Since they are packaged non-sterile, they must be high-level disinfected or sterilized before initial use. Prior to each subsequent procedure, they must be subjected to an appropriate cleaning and either high-level disinfection or sterilization processes.

{5}------------------------------------------------

### V. INDICATIONS FOR USE

There are no changes to the Indications for Use for the subject device. The sole purpose of this 510(k) is to modify the manual pre-cleaning and cleaning portions of the instructions to streamline them and improve clarity. No changes are being made to the high-level disinfection instructions.

This instrument is intended to be used with a PENTAX Video processor (including light source), documentation equipment, Monitor, Endotherapy Devices such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgerv within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

## VI. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE

The subject device is identical to the predicate device. The subject device has the same fundamental technology and operating principles of the predicate device, including the same intended use and design technological characteristics, such as Insertion Portion, Control Body and fiberoptics illumination. There are no differences in specifications, including, but not limited to, depth of field, distal end width, insertion tube width, instrument channel width, and total length. The sole difference between the subject and predicate device is that the reprocessing instructions have been modified. This modification does not impact the intended use and does not raise different questions of safety and effectiveness.

### VII.PERFORMANCE DATA

The optimization of the reprocessing instructions for the PENTAX Medical Video Colonoscopes do not affect the biocompatibility, electrical safety, electromagnetic compatibility, software verification and validation, or performance testing for the scopes.

Reprocessing validation data is provided to support the equivalence of the subject and predicate devices, including soil accumulation and cleaning efficiency studies.

### VIII. CONCLUSION

The subject device is identical to the predicate device. The subject device has the same intended use and technological characteristics as the predicate device. There are no changes to the design of the subject device.

The data submitted support the revised reprocessing instructions for the PENTAX Medical Video Colonoscopes and demonstrate that the endoscopes can be reprocessed in a safe and effective manner.

{6}------------------------------------------------

#### PENTAX Medical Video Colonoscopes (EC Family) Optimized IFU Traditional 510(k) Submission

The data provided in this 510(k) Premarket Notification support the equivalence of the subject and predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K181084](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K181084)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
