← Product Code [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF) · K120289

# PEERSCOPE SYSTEM (K120289)

_Peermedical · FDF · Sep 28, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K120289

## Device Facts

- **Applicant:** Peermedical
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Sep 28, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients. The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.

## Device Story

PeerScope System is a flexible colonoscopy system for diagnostic visualization and therapeutic intervention in the lower digestive tract. Input consists of optical images captured by the endoscope; system includes camera heads, light source, and video processing unit. Device provides 160° standard front field of view and 300° wide field of view. Used in healthcare facilities/hospitals by physicians. Output is a video feed displayed for the clinician to guide navigation and therapeutic procedures. System facilitates clinical decision-making by providing enhanced visualization of the colon. Benefits include improved diagnostic capability via wide-angle imaging.

## Clinical Evidence

Clinical validation performed at Elisha Medical Center, Israel, on 50 adult patients of diverse gender, ethnicity, and age. Procedures performed by US, European, and Israeli physicians. No adverse events observed. Bench validation conducted at three US medical centers under simulated clinical conditions to evaluate 300° field of view. All testing passed criteria.

## Technological Characteristics

Flexible endoscope system; 160° standard/300° wide field of view. Components: camera heads, MCU, light source. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 10993 (biocompatibility), ISO 8600 (endoscope specs), ASTM E 1837. Factory-adjusted brightness control. Video output: Composite and Y/C channels. Reprocessing validated per May 2011 FDA draft guidance.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- ProtectiScope CS ([K081004](/device/K081004.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP
2 8 2012

## 510(k) Summary of Safety and Effectiveness (in accordance to 21 CFR 807.92)

#### Submitter Details

PeerMedical Ltd. 2 Hatochen St., PO 3161 Business and Industrial Park (North), Caesarea, 38900, Israel Tel: +972 4 6327731, Fax: +972 4 6327734 Date summary prepared: June 2012 Submission Contact: Dan Laor, 6 Sireni St., Haifa, 32972, Israel.

#### Device Details

Proprietary Name: Common Name: Classification Name Product Code: Subsequent Product Code: Committee/Panel: Device Class:

PeerScope System Colonoscope and accessories, flexible / rigid Endoscope accessories, 21 CFR 876.1500 FDF KOG Gastroenterology/Urology II

Reason for 510(k) Submission: New Device

## Identification of Legally Marketed Predicate Devices

ProtectiScope CS -K081004 - Manufactured by Stryker GI

#### Device Description

The PeerScope system consists of endoscopic Main Control Unit (MCU), and of the PeerScope CS colonoscope. The MCU controls the endoscope. As other endoscopic legally marketed systems, it includes video system, light source, and interface to other ancillary equipment. The device labeled for use in healthcare facility/hospital for endoscopy and endoscopic treatment within the lower digestive tract. The operation principles of the PeerScope System are similar to those of other legally marketed standard colonoscopy systems. The PeerScope system model provides 160° standard front field of view and a 300° wide field of view.

#### Intended use and indications for Use

The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients. The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.

{1}------------------------------------------------

## Technological characteristics

The technological characteristics of PeerScope system were compared to the characteristics of the predicate device and were found very similar, in terms of: endoscope geometry, illumination type, energy source, lens irrigation, air flow and air pressure. Several differences were observed and are listed below:

K1202

| Category                                             | ProtectiScope<br>CS K081004                        | PeerScope System                          | Impact of the differences on<br>device performance                                                                                              |
|------------------------------------------------------|----------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Brightness Control                                   | Automatic or<br>Manual                             | Factory Adjusted                          | Factory Adjusted method provides<br>reliable design without<br>compromising performance.                                                        |
| Output Video<br>signals                              | VBS 2 channels<br>Y/C 2 channels<br>DVI 2 channels | Composite 3<br>channels<br>Y/C 3 channels | Both devices utilized industry<br>accepted standard Output Video<br>signals.                                                                    |
| Foot Switch                                          | Included                                           | Not Included                              | The foot switch is used to operate<br>the ProtectiScope CS unique<br>feature which is not included in the<br>PeerScope System specifications.   |
| Water Flow rate for<br>Auxiliary Jet water<br>supply | 20 cc/sec                                          | 4.0 cc/sec                                | The design of the reduced jet water<br>flow has been verified & validated.<br>The results demonstrated that the jet<br>irrigation is efficient. |
| Disposable Sleeve                                    | Included                                           | Not Included                              | The disposable sleeve is a unique<br>feature which is not included in<br>the PeerScope System<br>specifications.                                |

Based on the results of verification, validation and performance testing, the impact of the above differences is insignificant in terms of the device safety and effectiveness for its intended use.

### Performance data

#### Bench data:

The device safety and performance were verified by tests by PeerMedical and accredited third party laboratories. List of standards was used / relied upon for testing:

IEC 60601-1: 1988 + Am. #1 (1991), #2 (1995) IEC 60601-2-18:1996 /EN 60601-2-18:1996 2nd Ed. +Am.1 (2000) IEC 60601-1-2:2007 ISO 10993:2009 Part #1 ISO 10993:2009 Part #5 ISO 10993:2010 Part #10 ISO 10993:2007 Part #12 ISO 8600-4 First edition 1997-07-01 ISO 8600-1 Second edition 2005-05-01 ISO 8600-3 First edition 1997-07-01 Am 1 ISO 8600-6 First edition 2005-03-15 ASTM E 1837- 96 (reapproved 2007)

{2}------------------------------------------------

K120289 page 3 of 3

Reprocessing validation was carried out in accordance with " Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Draft Guidance / May 2011.

The device performance was validated by testing of a final production unit in simulated clinical conditions focusing on evaluation of the device wide 300° field of view. The procedure was performed by physicians at three US medical centers. Methods and procedures were identical to those used in clinical conditions. The results of the bench validation passed the experiment criteria. The device met its intended use and specifications. Hazardous conditions were not observed.

#### Clinical Data:

The device performance was validated by testing of a final production unit in clinical conditions. The procedure was performed at Elisha Medical Center, Haifa, Israel on a diverse adult population in terms of gender, ethnicity and age. The procedure was performed by US, European and Israeli physicians. Methods and procedures were identical to those used in USbased studies. Fifty (50) patients were enrolled in conformance with the device labeling. The results of the clinical validation passed the experiment criteria. No adverse events were observed.

The conclusions drawn from the bench and clinical tests demonstrate that the device meets its specifications, intended use and indication for use.

#### Substantial Equivalence

The above presented data demonstrate that:

- The predicate device ProtectiScope CS -K081004 is legally marketed. a.
- The PeerScope System and the predicate device have the same intended use. b.
- The PeerScope System does not raise different questions of safety and C. effectiveness.
- The data provided demonstrate equivalence and support the indications. d.

Conclusion: It is the opinion of PeerMedical that the PeerScope system is substantially equivalent to the predicate device, in terms of safety and effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the circumference of the circle. In the center of the seal is a stylized image of an eagle or other bird with its wings spread, with the body and wings formed by thick, curved lines. The image is black and white.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609. Silver Spring, MD 20993-0002

2 8 2012

PeerMedical Ltd.
% Mr. Dan Laor
Quality & Regulatory Advisory
6 Sireni
HAIFA 32972
ISRAEL

Re: K120289

Trade/Device Name: PeerScope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: August 14, 2012 Received: August 15, 2012

Dear Mr. Laor:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your bootion 310(t) presidentially equivalent (for the indications ferenced above and nave colorizes ally marketed predicate devices marketed in interstate for the stated in the encrosule) to regars mannent date of the Medical Device American Cruze commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, devices that have been rochasinou in accessariou of a premarket approval application (PMA). alla Cosmetic Hot (110) market the device, subject to the general controls provisions of the Act. The I ou may, merciolo, market the device, books of the requirements for annual registration, listing of general controls provisions of the fisc labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract lability adulteration. Trease note. CDICT does inst bevice labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (500 above) and in the major regulations affecting your device can be It may be subject to additional consols. Entrally - Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA 3 issuation of a brander complies with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA has made a decornmance and regulations administered by other Federal agencies. You must or ally Federal statutes and regulations , including, but not limited to: registration and listing comply with an the Act 3 requirements, moradaily, strenging (reporting (reporting of medical

{4}------------------------------------------------

Page 2-

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse evones) (21 OFF 003), good many and 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 1000 10 forth in the quality by brounds (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific actives for your actives/CDRH/CDRHOffices/ucm115809.htm for go to mp.nww.ida.gov/rtbound2123.cal Health's (CDRH's) Office of Compliance. Also, please the comer for DOY.cos and "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou may obtain other general incomment one Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

K120289 510(k) Number (if known):

Device Name: PeerScope System

Indications For Use:

The PeerScope System is intended for diagnostic visualization of the digestive tract. The system also provides access for therapeutic interventions using standard endoscopy tools. The PeerScope System is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve) for adult patients.

The PeerScope System consists of PeerMedical camera heads, endoscopes, video system, light source and other ancillary equipment.

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1

Page 1 of

n Sian-Off

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K120289](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K120289)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
