# ENDOGUIDE (K063654)

_Kms Medical, LLC · FDF · Jan 31, 2007 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K063654

## Device Facts

- **Applicant:** Kms Medical, LLC
- **Product Code:** [FDF](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF.md)
- **Decision Date:** Jan 31, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The EndoGuide is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and removal of multiple polyps and/or foreign bodies.

## Device Story

EndoGuide is an overtube device designed for use with a pediatric colonoscope. It is back-loaded onto the colonoscope shaft prior to insertion. The device features a dual-mode operation controlled by an external vacuum source connected to the endoscopy suite's vacuum line. In 'active' mode (vacuum on), the device becomes rigid, facilitating colonoscope passage and resisting looping. In 'passive' mode (vacuum off), the device remains flexible, allowing for deeper advancement into the colon. Operated by physicians in an endoscopy suite, the device assists in maintaining access and stability during procedures, potentially reducing procedural difficulty and improving patient outcomes during polyp or foreign body removal.

## Clinical Evidence

Bench testing only. In-vitro testing was conducted to confirm performance characteristics and demonstrate substantial equivalence to the predicate device.

## Technological Characteristics

The device is an overtube for pediatric colonoscopes. It utilizes a vacuum-actuated mechanism to transition between a flexible state and a rigid state. It connects to standard hospital vacuum lines. The device is designed for use in an endoscopy suite environment.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- US Endoscopy Group, Guardus™ Overtube ([K040836](/device/K040836.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063654

## SECTION 5 510(k) SUMMARY

| 510(k) Owner: KMS Medical LLC |
|-------------------------------|
| 7290 SW 42nd Street           |
| Miami, FL                     |
| 33155                         |
| Tel: 305-266-3388             |
| Fax: 305-267-7589             |

JAN 3 1 2007

| Trade Name:                | EndoGuide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:               | EndoGuide Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name:       | Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Regulation: | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Class:                     | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code:              | KOG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device:          | US Endoscopy Group, Guardus™ Overtube, K040836                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description:        | The EndoGuide fits over a pediatric colonoscope and can<br>be converted from a flexible to a rigid configuration on<br>demand. The EndoGuide is back loaded onto the shaft of<br>the colonoscope before the instrument is inserted into the<br>patient. The device was designed to be activated by a<br>vacuum source and needs to be connected to the existing<br>vacuum line in the endoscopy suite. Application of<br>vacuum causes the device to become rigid. In this "active"<br>mode (vacuum on) the rigid insertion tube facilitates the<br>passing of the colonoscope thru the insertion tube and<br>resists colonoscope looping. In the "passive" mode<br>(vacuum off) the insertion tube is very flexible and can be<br>advanced deeper into the colon. |
| Intended Use:              | The EndoGuide is intended to be used with an endoscope to<br>facilitate intubation, change of endoscopes, and removal of<br>multiple polyps and/or foreign bodies.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device Comparison:         | The proposed device is similar to the US Endoscopy<br>Guardus™ Overtube. Both devices have identical<br>indications for use. The EndoGuide differs in that it is<br>capable of becoming rigid when required by the physician.<br>When rigid the device facilitates the passing of the<br>endoscope thru the insertion tube and resists endoscopic                                                                                                                                                                                                                                                                                                                                                                                                                  |

·

{1}------------------------------------------------

looping. But when in its flexible state, the insertion tube is can be advanced into the colon.

Performance Data: To support the substantial equivalence determination, a summary of the in-vitro testing to confirm the performance characteristics in comparison with the predicate device has been submitted.

Summary Prepared by:

Mario Arbesu Director of Product Assurance, KMS Medical LLC

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized feathers and wings.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Mr. Mario Arbesu Director of Product Assurance KMS Medical LLC 7290 S.W. 42nd Street MIAMI FL 33155

JAN 3 1 2007

Re: K063654

Trade/Device Name: EndoGuide Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF and KOG Dated: December 6, 2006 Received: December 13, 2006

Dear Mr. Arbesu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket, Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo. The word "Centennial" is below the letters "FDA". There are three stars below the word "Centennial".

Protecting and Promoting Public Health

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## SECTION 4 INDICATIONS FOR USE

X063654

510(k) Number (if known): TBD

Device Name: EndoGuide

Indications for Use:

The EndoGuide is intended to be used with an endoscope to facilitate intubation, change of endoscopes, and removal of multiple polyps and/or foreign bodies.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                  |         |
|--------------------------------------|---------|
| Division of Reproductive, Abdominal, |         |
| and Radiological Devices             |         |
| 510(k) Number                        | K063654 |

KMS Medical EndoGuide 510(k) Submission

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K063654](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FDF/K063654)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).