← Product Code [FCW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCW) · K993300

# COLDLIGHTSOURCE PL3000 (K993300)

_Fiber Imaging Technologies, Inc. · FCW · Nov 5, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCW/K993300

## Device Facts

- **Applicant:** Fiber Imaging Technologies, Inc.
- **Product Code:** [FCW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCW.md)
- **Decision Date:** Nov 5, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

To provide illumination through lightconducting devices such as fiberoptic cables.

## Device Story

The Photonics Coldlightsource PL3000 is a fiberoptic light source used to provide illumination for medical procedures. It functions by generating light from a 150-watt halogen lamp, which is then transmitted through fiberoptic cables to the surgical or examination site. The device features adjustable light intensity control. It is intended for use in clinical settings by healthcare professionals. The device operates on 115/230VAC power and is designed to support endoscopic or other light-conducting medical applications, aiding visualization during procedures.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Halogen lamp light source; 150W lamp wattage; 115/230VAC 60Hz power; adjustable light intensity control; fiberoptic cable connection interface. Conforms to ENEC 11, EN60601, EN55011, and UL 2601 standards.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Luxtec Model 3150 Light Source ([K890716](/device/K890716.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NUV - 5 1999

K993300

## Appendix D 510 (k) Summary

September 23, 1999

l. Submitted for Photonics Optics GESMBH by the initial importer and distributor, Fiber Imaging Technologies, Inc., 326 Clark Street, Worcester, Massachusetts, 01606. Submission correspondent and contact: Mr. Joseph Ress, Medical Device Requlatory Consultants, 45 Pontiac Road, Newton, MA 02168, whose telephone and FAX number is (617) 965-7714.

2. Trade Name of the Device: Coldsource PL3000

Common Name: Light source

Classification Name: Endoscope and Accessories, Product code: KOG (per 876.1500)

K890716, Luxtec Model 3150 Light Source 3. Predicate Device:

4. Description of the Device: The Photonics Coldlightsource is a lightsource for fiberoptic cables, with a 230/240 PL3000 volt version which conforms to ENEC 11, EN60601, EN55011, and UL 2601. The 120 volt version is manufactured to the same requirements of these standards, and has been submitted for approval. Approval is anticipated in the immediate future.

5. Intended Use: To provide illumination through lightconducting devices such as fiberoptic cables.

6. Comparison to Predicate Device:

| Feature                     | Photonics<br>Model PL3000 | Luxtec<br>Model 3150S |
|-----------------------------|---------------------------|-----------------------|
| Power requirements          | 115/230VAC 60 Hz          | 115VAC 60 Hz          |
| Power consumption           | approx 200 W              | approx 200 W          |
| Lamp type                   | Halogen                   | Halogen               |
| Lamp Wattage                | 150 Watts                 | 150 Watts             |
| Color Temperature           | 3400° Kelvin              | 3250° Kelvin          |
| Light Intensity Control     | Yes                       | Yes                   |
| Fiberoptic cable connection | yes                       | yes                   |

7. Conclusions: Photonics Optics GMBH believes that based upon the foreqoinq, the Model PL3000 is substantially equivalent to the predicate device (K890716) marketed by Luxtec Model 3150S.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing to the right, arranged in a stacked formation. The profiles are depicted in a simple, curved line style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the profiles.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 5 1999

Fiber Imaging Technologies, Inc. c/o Mr. Joseph Ress Medical Device Regulatory Consultants 45 Pontiac Road Newton, MA 02168

Re: K993300 Photonics Coldlightsource PL3000 Dated: September 24, 1999 Received: October 1, 1999 Regulatory Class: II 21 CFR §876.1500/Procode: 78 FCW

Dear Mr. Ress:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiiance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

Page 1 of l

510(K) Number: к 993300

Device Name: Photonics Coldlightsource PL3000

Indications for Use:

The Photonics Coldlightsource PL3000 is indicated for use to provide illumination through light-conducting devices such as fiberoptic cables.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )

Concurrance of CDRH, Office of Device Evaluation (ODE)

Perscription Use yes

OR

Over-The-Counter Use NO

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Device

510(k) Number K993300

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCW/K993300](https://fda.innolitics.com/submissions/GU/subpart-b%E2%80%94diagnostic-devices/FCW/K993300)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).