K984168 · United States Endoscopy Group, Inc. · FCF · Dec 7, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K984168
Device Name
BARRETT'S ESOPHAGEAL CYTOLOGY DEVICE
Applicant
United States Endoscopy Group, Inc.
Product Code
FCF · Gastroenterology, Urology
Decision Date
Dec 7, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Barrett's Cytology Device is intended to collect cytologic samples from the esophageal mucosa without the aid of endoscopy for the purpose of cytologic analysis to determine the presence of abnormalities like Barrett's Esophagus.
Device Story
Non-endoscopic intubation catheter with integrated cytology brush; used to collect esophageal mucosal samples. Physician advances catheter into stomach; distal inflatable balloon identifies lower esophageal sphincter (LES). External slide disc marks distance from dental arches; catheter withdrawn to target sampling site. Cytology brush deployed to collect cells; brush retracted into catheter lumen to protect sample during withdrawal. Device enables lower-cost surveillance compared to endoscopic brush cytology. Physician performs procedure; samples sent for microscopic analysis to identify abnormalities. Benefits include reduced procedure cost and avoidance of endoscopy.
Clinical Evidence
No clinical data provided in the submission. The document references literature regarding balloon cytology and sponge samplers for esophageal surveillance, but the device itself relies on bench testing and adherence to design controls for safety and effectiveness.
Technological Characteristics
Catheter-based system with distal inflatable balloon for LES localization and internal cytology brush. Materials and manufacturing follow FDA Quality System Regulations, ISO 9001, and EN46001. Sterilized via ethylene oxide gas. Manual operation; no electronic components or software.
Indications for Use
Indicated for patients requiring esophageal cytologic sampling to detect abnormalities such as Barrett's Esophagus. Contraindicated in patients with liver cirrhosis, esophageal varices, active gastroduodenal bleeding, risk of perforation/aspiration, severe heart disease, hypertension, pulmonary deficiencies, topical anesthetic allergies, esophageal obstruction, or acute laryngitis.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Reference Devices
Balloon Cytology (Falk G, et al. 1997)
Balloon and Sponge Samplers (Roth M, et al. 1997)
Related Devices
K230339 — EsoCheck Cell Collection Device · Lucid Diagnostics, Inc. · Feb 24, 2023
K210137 — EsoCheck Cell Collection Device · Lucid Diagnostics, Inc. · Feb 18, 2021
K222366 — EsoCheck Cell Collection Device · Lucid Diagnostics, Inc. · Oct 26, 2022
K183262 — EsoCheck CCD Cell Collection Device · Lucid Diagnostics, Inc. · Jun 21, 2019
K203450 — EsophaCap Swallowable Cellular Retrieval Device (changed from Cell-Mata) · Capnostics, LLC · May 4, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
12/7/99
K984168
# SECTION II 510k SUMMARY OF SAFETY AND EFFECTIVENESS
The Summary of Safety and Effectiveness on the Barrett's Esophageal Cytology Device reflects data available and presented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies may require alterations of the conclusions or recommendations set forth.
## Procedure/Product Overview
The Barrett's Cytology Device is intended to collect cytologic samples from the esophageal mucosa without the aid of endoscopy for the purpose of cytologic analysis to determine the presence of abnormalities like Barrett's Esophagus. In the past the methods of obtaining cytologic samples from the gastrointestinal tract were limited to brush cytology completed under direct visualization methods via endoscopy. This method although proven has a high procedure cost associated with it. The Barrett's Cytology Device combines non endoscopic intubation catheter methods and cytology brush technology to provide a device that will lower the cost of surveillance for this patient population. The Barrett's Cytology Device is a catheter that is advanced into the esophagus much like a gastric lavage catheter. The catheter with the aid of common intubation techniques is further advanced into the stomach. The placement into the stomach is confirmed by the use of a distal inflatable balloon which is used to identify the lower esophageal sphincter (LES). Identification of the LES is required because the target sample location is typically the distal esophagus. Once the LES is identified an external slide disc on the catheter is used to mark the length of the patient's esophagus from the dental arches. At this time the balloon will be deflated and the catheter withdrawn away from the LES into the esophagus using the slide disc and catheter markings for distance reference as to how far back the catheter is to be withdrawn. When the catheter is withdrawn to the desired location to sample a cytology brush contained within the catheter is deployed out the distal end of catheter and the cytologic brushing commences. the Once the brushing of the esophagus is completed the brush is withdrawn back inside the catheter to contain the cell harvest within the
{1}------------------------------------------------
catheter lumen. The entire catheter is withdrawn from the patient and the cytology brush is again deployed to prepare microscopic slides for analysis.
### Contraindications
The use of the Barrett's Cytology Device is contraindicated and is not appropriate in cases where the patients have or present with:
- Consideration of bleeding complications in patients with liver । cirrhosis, esophageal varices, active gastroduodenal bleeding
- Consideration of the potential complications of esophageal or . gastric bleeding and or perforation and aspiration digestive fluids into the lung.
- i Consideration of stress associated with intubation procedures (gagging, discomfort) in patients with severe heart disease, hypertension, or pulmonary deficiencies.
- ー Reaction to topical anesthetics.
- -Insertion may be complicated in patients with esophageal obstruction, acute laryngitis, or general debilitation of the patient due to other causes.
- । Other contraindications as determined by physician.
## Manufacturing Overview
U.S.E. will manufacture and test the product to performance specifications based on predicate and/or substantially equivalent devices.
U.S.E. quality system is based on the requirements of The FDA Quality System Regulations and it is also certified to ISO 9001 and EN46001 and markets devices under following the requirements of The European Medical Device Directive. The U.S.E. quality system methods and procedures are utilized to assure the product conformance to design specifications. The device design, development and validation will be implemented following relevant design control procedures.
{2}------------------------------------------------
# Sterilization
This device will be offered in a sterile packaging configuration. The ethylene oxide gas process, packaging materials and methods will be subject to the validation requirements of the U.S.E. quality system.
# Bibliography
- Falk G, Chittajallu R, et al. Alimentary Tract Surveillance of Patients with Barrett's Esophagus for Dysplasia and Cancer with Balloon Cytology. Gastroenterology. 1997;112:1787-1797
- Falk G, Richter J, Reflux Disease and Barrett's Esophagus. Endoscopy. 1998; 30: 61-72
- Roth M, Liu S, et al. Cytologic Detection of Baophageal Squamous Cell Carcinoma and Precursor Lesions Using Balloon and Sponge Samplers in Asymtomatic Adults in Linxian, China. Cancer. 1997; vol.80: 2047-2059
- Rayhorn N, Intubation and Drainage. Gastroenterology Nursing A Core Curriculum.1998; Chapter 29:307-321
201
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH &" is visible above the eagle, and "HUMAN SERVICES" is visible below the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 7 1999
Mr. Dean Secrest U.S. Endoscopy Group 5976 Heisley Road Mentor, OH 44060
Re: K984168 Barrett's Esophageal Cytology Device Dated: October 7, 1999 Received: October 12, 1999 Requiatory Class: Il 21 CFR 876.1075/Procode: 78 FCF
Dear Mr. Secrest:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclasslified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements conceming your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number if Known_
K984168
Device Name: Barrett's Cytology Device
Indications for Use:
The Barrett's Cytology Device is intended to collect cytologic samples from the esophageal mucosa without the aid of endoscopy abrorralities bil abnormalities like Barrett's Esophagus.
(please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE) Prescription Use OR Over-the-Counter Use (Per 21 CFR 801.109)
David C. Severson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K984168/5
2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.