K161147 · Endocure Technologies, Inc. · FBN · Jun 7, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K161147
Device Name
EndoCure Model EUR078A
Applicant
Endocure Technologies, Inc.
Product Code
FBN · Gastroenterology, Urology
Decision Date
Jun 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EndoCure Model EUR078A has been designed for endoscopic observation and diagnosis within both the urinary tract (ureter, renal pelvis, and renal calyx) and biliary tract (common bile duct and hepatic duct). Therapeutic endoscopic procedures using various kinds of accessories are also possible, and include the ability to extract or fragment stones within these organs.
Device Story
EndoCure Model EUR078A is a flexible fiberoptic ureteroscope used for visualization and therapeutic intervention in the urinary and biliary tracts. The device is inserted into the patient to examine the ureters, renal pelvis, renal calyx, common bile duct, or hepatic duct. It features an extra channel for inserting accessories, such as a basket for stone removal or a flexible fiber for laser lithotripsy. The device is operated by a physician in a clinical setting. It is not compatible with electro-surgical (cautery) devices due to its metal distal tip. The device provides direct visualization to assist the physician in diagnosing conditions and performing stone management, potentially benefiting the patient by enabling minimally invasive stone treatment.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, -10, -11), electrical safety (IEC 60601-1, -1-2, -2-18), and performance testing (FTIR, SEM, angulation flex, and optical performance per ISO 8600-3 and ISO 8600-4).
Technological Characteristics
Flexible fiberoptic endoscope with 90-degree field of view and 1-50mm depth of view. Features a 7.3 French distal end, 8.4 French insertion tube, and 1.2mm working channel. Bending angulation is 270 degrees up/down. Materials evaluated for biocompatibility per ISO 10993. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Optical system uses a fiberoptic bundle.
Indications for Use
Indicated for endoscopic observation, diagnosis, and therapeutic procedures (stone extraction/fragmentation) in the urinary tract (ureter, renal pelvis, renal calyx) and biliary tract (common bile duct, hepatic duct).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K963855 — 12 URETEROSCOPES/URETERO-RENOSCOPES · Richard Wolf Medical Instruments Corp. · Apr 3, 1997
K200369 — OES Elite Ureteroscopes and Accessories · Olympus Winter & Ibe GmbH · Aug 28, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2016
EndoCure Technologies, Inc. % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, MD 21114
Re: K161147
> Trade/Device Name: EndoCure Model EUR078A Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FBN Dated: April 20, 2016 Received: April 22, 2016
Dear E.J. Smith,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
Page 2 - E.J. Smith
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a large, bold, sans-serif font. The text is black against a white background, making it highly legible. The name appears to be a title or heading, possibly indicating the subject of a document or presentation.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161147
Device Name EndoCure Model EUR078A
#### Indications for Use (Describe)
The EndoCure Model EUR078A has been designed for endoscopic observation and diagnosis within both the urinary tract (ureter, renal pelvis, and renal calyx) and biliary tract (common bile duct and hepatic duct). Therapeutic endoscopic procedures using various kinds of accessories are also possible, and include the ability to extract or fragment stones within these organs.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
31
# 510(k) Summary 807.92C
#### SPONSOR
# 807.92(a)(1)
| Company name: | EndoCure Technologies, Inc. |
|---------------------------|---------------------------------|
| Company Address: | 5801 Ammendale Road, Suite |
| | Beltsville, MD 20705 |
| Telephone Number: | 301-937-8848 |
| Contact Person: | Charles J. Neff (Vice President |
| Summary Preparation Date: | Operations) April 19, 2016 |
DEVICE NAME
807.92(A)(2)
| Trade Name: | EndoCure Model EUR078A |
|----------------------|----------------------------|
| Common/Usual Name: | Flexible Endoscope |
| Classification Name: | Endoscopes and Accessories |
| Regulation Number: | 21 CFR 876.1500 |
| Product Code: | FBN |
| Device Class: | Class II |
#### PREDICATE DEVICE
# 807.92(a)(3)
| Company | Product | 510(k) Number |
|---------------|--------------------------------------------|---------------|
| Olympus, Inc. | Ureterorenofiberscope/Choledochofiberscope | K912120 |
#### DEVICE DESCRIPTION
# 807.92(a)(4)
EndoCure Model EUR078A is used during the examination / procedure as an ureteroscope. The ureteroscope is an instrument for examining the inside of the urinary tract. The ureteroscope is used to see beyond the bladder into the ureters, the tubes that carry urine from the kidneys to the bladder. Through the ureteroscope, the doctor can see a stone in the ureter and then remove it with a small basket at the end of a wire inserted through an extra channel in the
{4}------------------------------------------------
ureteroscope. Another way to treat a stone through an ureteroscope is to extend a flexible fiber through the scope up to the stone and then, with a laser beam projected through the fiber, break the stone into smaller pieces that can then pass out of the body in the urine. The EndoCure Model EUR078A ureteroscope is not designed for or intended to be used with electro-surgical (cautery) devices due to the metal distal tip.
#### DEVICE INDICATIONS FOR USE
The EndoCure Model EUR078A has been designed for endoscopic observation and diagnosis within both the urinary tract (ureter, renal pelvis, and renal calyx) and biliary tract (common bile duct and hepatic duct). Therapeutic endoscopic procedures using various kinds of accessories are also possible, and include the ability to extract or fragment stones within these organs.
## COMPARISON OF TECHNICAL CHARACTERISTICS
#### 807.92(A)(6)
807.92(A)(5)
| Parameters | EndoCure Model<br>EUR078A | Olympus Corporation<br>URF-P2<br>Ureterorenofiberscope/<br>Choledochofiberscope | Similarities and<br>Differences |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| 510(k) Number | K141866 | K912120 | N/A |
| Intended Use | The EndoCure Model<br>EUR078A has been<br>designed for<br>endoscopic observation<br>and diagnosis within<br>both the urinary tract<br>(ureter, renal pelvis,<br>and renal calyx) and<br>biliary tract (common<br>bile duct and hepatic<br>duct). Therapeutic<br>endoscopic procedures<br>using various kinds of<br>accessories are also<br>possible, and include<br>the ability to extract or<br>fragment stones within<br>these organs. | The Olympus URF Type<br>P2 has been designed<br>for endoscopic<br>observation and<br>diagnosis within both<br>the urinary tract (ureter,<br>renal pelvis, and renal<br>calyx) and biliary tract<br>(common bile duct and<br>hepatic duct).<br>Therapeutic endoscopic<br>procedures using<br>various kinds of<br>accessories are also<br>possible, and include<br>the ability to extract or<br>fragment stones within<br>these organs. | Substantially<br>equivalent<br>indications for use |
| Field of View | 90 degree | 90 degree | Identical |
| Depth of View | 1-50mm | 1-50mm | Identical |
| Optical System | Fiberoptic Bundle | Fiberoptic Bundle | Identical |
| Distal End Outer | 7.3 French (2.43mm) | 9.3 French (3.1mm) | Difference in distal |
{5}------------------------------------------------
| Diameter | | | end outer diameter |
|------------------------|--------------------|--------------------|-----------------------------------|
| Insertion Tube O.D. | 8.4 French (2.8mm) | 9.9 French (3.3mm) | Difference in insertion tube O.D. |
| Bending Angulation | Up:270/Down:270 | Up:180/Down:180 | Difference in bending angulation |
| Channel Inner Diameter | 1.2mm | 1.2mm | Identical |
| Working Length | 700mm | 700mm | Identical |
# Nonclinical Testing
# Safety Testing
- ISO 10993-5 Biological Evaluation of Medical Devices Part 5: Tests for In Vitro . Cytotoxicity (ISO MEM Elution Assay with L-929 Mouse Fibroblast Cells)
- ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization (Intracutaneous Reactivity Test) (ISO Guinea Pig Maximization Sensitization Test)
- ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization (Acute Systemic Injection Test)
- . ISO 10993-11: Biological Evaluation of Medical devices – Part 11: Test for Systemic Toxicity
- IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) 0r IEC 60601-. 1: 2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2007 Third Edition, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility – Requirements and tests
- IEC 60601-2-18 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of endoscopic equipment
# Performance testing
- . FTIR and SEM Analysis
- . Angulation Flex Bench test
- . ISO 8600-3 Optics and Optical Instruments – Medical endoscopes and endoscopic accessories – Part 3: Determination of field of view and direction of view of endoscopes with optics.
{6}------------------------------------------------
- ISO 8600-4 Optics and Optical Instruments Medical endoscopes and . endoscopic accessories – Part 4: Determination of maximum width of insertion portion.
## CONCLUSION
# 807.92(B)(3)
The EndoCure Model EUR078A is substantially equivalent to the predicate device in intended use, materials and design. Safety and performance testing to ISO 8600-3 and ISO 8600-4 and biocompatibility testing to ISO 10993 Standards concluded that the device does not introduce significant questions of safety and efficacy and is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.