VISION-SCIENCES MODEL S-V200 COLOR MVIDEO FLEXIBLE SIGMOIDOSCOPE WITH ENDOSHEATH SYSTEM
Device Facts
| Record ID | K964326 |
|---|---|
| Device Name | VISION-SCIENCES MODEL S-V200 COLOR MVIDEO FLEXIBLE SIGMOIDOSCOPE WITH ENDOSHEATH SYSTEM |
| Applicant | Vision-Sciences, Inc. |
| Product Code | FAM · Gastroenterology, Urology |
| Decision Date | Jan 3, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The proposed VSI Model S-V200 Color Video Sigmoidoscope is intended for use to examine the lower bowel tract, and, using additional accessories, to perform various diagnostic and therapeutic procedures.
Device Story
Reusable video sigmoidoscope with D-shaped cross-section; incorporates fiberoptic illumination and imaging bundles. Utilizes sterile, disposable EndoSheath covering containing air, water, and suction/biopsy channels. Sheath disposed after each procedure. Used by clinicians in clinical settings for lower bowel tract visualization and intervention. Output is real-time video feed for physician observation; facilitates diagnostic/therapeutic decision-making. Disposable sheath design eliminates need for high-level disinfection of scope between patients, reducing cross-contamination risk.
Clinical Evidence
Bench testing only. Performance testing included scope angulation/cycle testing and scope resolution testing, both performed with and without the EndoSheath. No clinical data or biocompatibility data presented.
Technological Characteristics
Reusable video sigmoidoscope; D-shaped cross-section; fiberoptic illumination and imaging bundles. Sterile, disposable protective EndoSheath provides air, water, and suction/biopsy channels. No software or electronic processing described.
Indications for Use
Indicated for examination of the lower bowel tract and performance of diagnostic and therapeutic procedures in patients requiring sigmoidoscopy.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- Vision-Sciences Model S-V100 Video Sigmoidoscope with SS-V32 EndoSheath System (K932843)
- Model VS-200 Video Sigmoidoscope (K954704)
Related Devices
- K964100 — VISION-SCIENCES MODEL SS-U32 ENDO SHEATH FOR USE WITH THE VSI S-F100 FLEXIBLE FIBEROPTIC SIGMOIDOSCOPE AND VSI S-V100 · Vision-Sciences, Inc. · Dec 16, 1996
- K964131 — ENDOSCOPE · Akos Biomedical, Inc. · Jun 27, 1997
- K963344 — DISPOSABLE SS-F32 ENDOSHEATH FOR USE WITH THE S-F100 FLEXIBLE FIBEROPTIC SIGMOIDOSCOPE · Vision-Sciences, Inc. · Nov 21, 1996
- K032688 — COLONOSIGHT MODEL 510B · Sightline Technologies , Ltd. · Mar 5, 2004
- K230332 — Ambu® aScope Colon; Ambu® aBox 2 · Ambu A/S · Sep 15, 2023