ES-3800, VIDEO SIGMOIDOSCOPE
Device Facts
| Record ID | K951577 |
|---|---|
| Device Name | ES-3800, VIDEO SIGMOIDOSCOPE |
| Applicant | Pentax Precision Instrument Corp. |
| Product Code | FAM · Gastroenterology, Urology |
| Decision Date | Mar 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Indications for Use
The ES-3800, Video Sigmoidoscope, is intended to provide optical visualization of, and therapeutic access to, the Lower Gastrointestinal Tract (extent of exam, sigmoid colon). The Lower Gastrointestinal Tract includes, but is not restricted to, the organs; tissues; and subsystems: Large Bowel. The instrument is introduced per rectally when indications consistent with the requirement for the procedure are observed in Adult and Pediatric patient populations.
Device Story
Flexible video sigmoidoscope; used with external video processor. Input: CCD-captured optical images of lower GI tract illuminated by xenon lamp/RGB color wheel system. Operation: CCD collects image data per color strobe; processor compiles full-color frames. Features: flexible insertion tube; control body for angulation, air/water/CO2 delivery, suction; working channel for biopsy forceps. Usage: clinical setting; operated by physicians. Output: video signal for display on monitor. Benefit: enables visualization and therapeutic access to large bowel for diagnosis/treatment.
Clinical Evidence
No clinical data; substantial equivalence based on bench testing and technical specifications.
Technological Characteristics
Flexible endoscope with CCD sensor; light-carrying bundles; 300W xenon lamp source; RGB color wheel filtering; manual angulation controls; air/water/CO2/suction channels; working channel for accessories.
Indications for Use
Indicated for optical visualization and therapeutic access to the lower gastrointestinal tract (sigmoid colon) in adult and pediatric patients requiring sigmoidoscopy.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- FS-34, Fiber Sigmoidoscope (K792341)
- VB-1530/ 1830, Video Bronchoscope (K934920)
- EPM-3300, Video Processor (K934918)
Related Devices
- K961567 — ES-3840, VIDEO SIGMOIDOSCOPE · Pentax Precision Instrument Corp. · Jun 24, 1996
- K961570 — KEC-3840L/F, VIDEO COLONOSCOPE (*L OR F IN MODEL NUMBER DENOTES LENGTH) · Pentax Precision Instrument Corp. · Jun 24, 1996
- K951579 — EC-3800TL, VIDEO COLONOSCOPE · Pentax Precision Instrument Corp. · Mar 21, 1996
- K951574 — EC-3800L, VIDEO COLONOSCOPE · Pentax Precision Instrument Corp. · Mar 21, 1996
- K961563 — EC-384OTL, VIDEO COLONSCOPE · Pentax Precision Instrument Corp. · Jun 19, 1996