The UV5000 Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics.
Device Story
The UV5000 Handle is a reusable, battery-operated handheld component of the Uro-G Cystoscope System. It integrates an LCD monitor directly into the handle, eliminating the need for external cables to a separate monitor. The device accepts video input from a disposable steerable endoscopic cannula; processes and displays real-time anatomical images to the physician during diagnostic procedures in hospitals or physician offices. The physician operates the device manually, using buttons on the handle to adjust LED illumination and capture still images or video. The output provides direct visualization of tissue structures, aiding clinical decision-making during endoscopic examinations. The device is cleaned and disinfected between uses.
Clinical Evidence
Bench testing only. No clinical or animal studies were conducted. Performance testing included field of view accuracy, depth of field, image distortion, resolution, uniformity, color performance, electrical safety, and EMC testing. All tests passed.
Technological Characteristics
Reusable handheld unit with integrated 5-inch LCD monitor (1024x768 resolution). Powered by a 21700 Lithium-Ion rechargeable battery. Connectivity: Standalone. Materials: Not specified (non-patient contacting). Sterilization: Not sterile; cleaned/disinfected via surface wipes. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18. Software: Version 1.0, Moderate Level of Concern.
Indications for Use
Indicated for use with UroViu Cannula (Uro-G HD, Uro-G, Uro-V, Uro-N, Hystero-V) to permit visualization of the urethra, bladder, cervical canal, and uterine cavity for diagnostic purposes.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K232837 — Uro-G HD Flexible Cystoscope · Uroviu Corporation · Jun 7, 2024
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 11, 2025
UroViu Corporation
Thomas Lawson, Ph.D.
Director, Regulatory Affairs
4546 El Camino Real, Suite 214
Los Altos, CA 94022
Re: K250797
Trade/Device Name: UV5000 Handle
Regulation Number: 21 CFR§ 876.1500
Regulation Name: Endoscope and Accessories
Regulatory Class: II
Product Code: FAJ
Dated: March 14, 2025
Received: July 30, 2025
Dear Thomas Lawson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250797 - Thomas Lawson, Ph.D.
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250797 - Thomas Lawson, Ph.D.
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark R. Kreitz -S
for Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology and Urology Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) | |
| K250797 | |
| Device Name UV5000 | |
| Indications for Use (Describe) | |
| The UV5000 Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
UroViu Corporation
K250797 Page 1of 10
Traditional 510(k) Notification UV5000 Handle
# 510(k) SUMMARY
General Information
| Submitter | UroViu Corporation |
| --- | --- |
| Address | 4546 El Camino Real
Suite 214
Los Altos, CA 94022 |
| Correspondence Person | Thomas Lawson, PhD
Director, Regulatory Affairs |
| Contact Information | Email: thom@uroviu.com
Phone: 510-206-1794 |
| Date Prepared | September 9, 2025 |
Proposed Device
| Trade Name | UV5000 Handle |
| --- | --- |
| Common Name | Endoscope and accessories |
| Regulation Number and Classification Name | 21 CFR§876.1500
Cystoscope and Accessories, Flexible/Rigid |
| Product Code | FAJ |
| Regulatory Class | II |
| Note: This is the first 510(k) submission for this device. | |
Predicate Device
| Trade Name | Uro-G HD Flexible Cystoscope |
| --- | --- |
| Common Name | Endoscope and accessories |
| Premarket Notification | K232837 |
| Regulation Number and Classification Name | 21 CFR§876.1500
Endoscope and Accessories |
| Product Code | FAJ |
| Regulatory Class | II |
| Note: This predicate device has not been subject to a design-related recall. | |
{5}
UroViu Corporation
K250797 Page 2 of 10
Traditional 510(k) Notification
UV5000 Handle
## Device Description
The Uro-G Cystoscope System is comprised of a cannula and handle that is intended to allow visualization of the urethra, bladder, cervical canal, and uterine cavity for the purpose of performing diagnostics. This 510(k) discloses the modifications made to the handle of the Uro-G HD Cystoscope System (K232837), UV5000 Handle. The Uro-G HD Flexible Cystoscope System is a handheld, battery-operated portable cystoscope consisting of a disposable steerable endoscopic cannula with a 2.2 mm channel for fluid infusion and instrument advancement, and a reusable handle with a video monitor.
## Indications for Use
The indication for use for the UV5000 Handle is:
The UV5000 Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics.
## Comparison to the Predicate Device
The focus of this Traditional 510(k) is the handle component of the Uro-G HD Flexible Cystoscope, which was reviewed in K232837 as the 4500 Handle. The UV5000 Handle is substantially equivalent to the U4500 Handle based upon the following equivalences:
1. The intended use of both devices is the same.
2. Both devices provide the means for the physician to view the urethra and bladder via the video monitor that is integrated into the body of the handle.
{6}
UroViu Corporation
K250797 Page 3 of 10
Traditional 510(k) Notification
UV5000 Handle
Table 1 shows the comparison of the UV5000 Handle (the subject of this submission) to the predicate device, the 4500 Handle of the Uro-G HD Flexible Cystoscope. The similarities between the two devices satisfy the criteria for a $510(\mathrm{k})$ notice.
Table 1. Comparison of the UV5000 Handle to the predicate device, the 4500 Handle of the Uro-G HD Flexible Cystoscope.
| | Subject Device UV5000 Handle UroViu Corp. (This submission) | Predicate Device Uro-G HD Flexible Cystoscope UroViu Corp. K232837 | Equivalence |
| --- | --- | --- | --- |
| Device Overview | | | |
| Device Class | II | II | Same |
| FDA Product Code | FAJ | FAJ | Same |
| Product Classification | 21 CFR 876.1500 | 21 CFR 876.1500 | Same |
| Intended Use | Used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics | Used to permit viewing of the urethra and bladder or cervical canal and uterine cavity for the purpose of performing diagnostics | Same |
| Indication for Use | The UV5000 Handle is to be used with UroViu Cannula Uro-G HD, Uro-G, Uro-V, Uro-N, and Hystero-V and is intended to be used to permit viewing of the urethra and bladder or cervical canal and uterine cavity | The Uro-G HD Flexible Cystoscope is intended for endoscopic diagnosis and infusion of irrigating fluids within the bladder and urethra | Same |
{7}
K250797 Page 4 of 10
UroViu Corporation
Traditional 510(k) Notification UV5000 Handle
| | for the purpose of performing diagnostics. | | |
| --- | --- | --- | --- |
| Contraindications | As described in each UroViu cannula instructions for use | As described in each UroViu cannula instructions for use | Same |
| | | | |
| Technical Characteristics | | | |
| Device Identity | UV5000 | Uro-G HD Flexible Cystoscope (4500 handle) | N/A |
| Components | • 1 piece handle
• 1 piece charging cradle
• 1 piece USB cable | • 1 piece handle
• 1 piece charging cradle
1 piece USB cable | Same |
| Length | 178 mm | 176 mm | Equivalent
The 2 mm difference in length does not introduce any new risk |
| Diameter | 38 mm | 36 mm | Equivalent
The 2 mm increase in diameter is to accommodate a different size battery but this does not introduce any new risk |
| Weight | 345 g | 264 g | Equivalent
The extra weight is mainly due to the increased size of the monitor and the larger capacity |
{8}
UroViu Corporation
K250797 Page 5 of 10
Traditional 510(k) Notification
UV5000 Handle
| | | | battery and does not introduce any new risk |
| --- | --- | --- | --- |
| Monitor | 5 inches diagonal
125.4 mm horizontal
91.4 mm vertical | 4.3 inches diagonal
122.2 mm horizontal
73.9 mm vertical | Equivalent
The slight larger monitor makes it easier for the user to view the image and does not introduce any new risk |
| Input format | RGB | RGB | Same |
| Effective pixels | 1280 x 720 | 1280 x 720 | Same |
| LCD resolution | 1024 x 768 | 800 x 480 | Equivalent
Increased resolution does not introduce any new risk |
| LCD display resolution | 0.8 million pixels | 0.4 million pixels | Equivalent
Increased resolution does not introduce any new risk |
| Power supply | 21700 Lithium Ion rechargeable battery | 18650 Lithium Ion rechargeable battery | Equivalent
The higher power supply necessary for the increased power consumption but it does not introduce any new risk in that it is compliant with IEC 60601-1 & 60601-2-18 |
| Voltage of fully | 4.2 V | 3.7 V | Equivalent |
{9}
UroViu Corporation
K250797 Page 6 of 10
Traditional 510(k) Notification
UV5000 Handle
| charged battery | | | The 0.5 V increase in power supply is necessary for the increased power consumption but it does not introduce any new risk in that it is compliant with IEC 60601-1 & 60601-2-18 |
| --- | --- | --- | --- |
| Time to obtain full charge | 4 hours | 4 hours | Same |
| Operating time after fully charged | 2 hours | 2 hours | Same |
| Battery capacity | 5000 mA | 3100 mA | Equivalent The higher capacity is necessary for the increased power consumption but it does not introduce any new risk in that it is compliant with IEC 60601-1 & IEC 60601-2-18 |
| Biocompatibility | N/A
Does not contact the patient & the user wears gloves while using the device | N/A
Does not contact the patient & the user wears gloves while using the device | Same |
| Operational Characteristics | | | |
| Electrical Safety | IEC 60601-1 & IEC 60601- | IEC 60601-1 & | Same |
{10}
UroViu Corporation
K250797 Page 7 of 10
Traditional 510(k) Notification UV5000 Handle
| | 2-18 compliant | IEC 60601-2-18 compliant | |
| --- | --- | --- | --- |
| EMC | IEC 60601-2 compliant | IEC 60601-2 compliant | Same |
| Adjust brightness of illumination of the LEDs | Adjust by depressing a button on the handle to change settings | Adjust by depressing a button on the handle to change settings | Same |
| Images or video capture | Capture still images (photo) or video during a procedure by depressing the camera button on the handle | Capture still images (photo) or video during a procedure by depressing the camera button on the handle | Same |
| Cleaning, disinfecting, and sterilizing | The handle is not sterile.
The handle is cleaned and disinfected following directions in the instructions for use (IFU):
• Wipe the surface with a Super Sani-Cloth or CaviWipe towelette
• Surface to remain visibly wet for 3 minutes
• After 3 minutes, allow the handle to air dry for 5 minutes
Wet a sterile pad with sterile water and wipe all exposed surfaces of the handle. | The handle is not sterile.
The handle is cleaned and disinfected following directions in the instructions for use (IFU):
• Wipe the surface with a Super Sani-Cloth or CaviWipe towelette
• Surface to remain visibly wet for 3 minutes
• After 3 minutes, allow the handle to air dry for 5 minutes
Wet a sterile pad with sterile water and wipe all | Same
Same |
{11}
UroViu Corporation
K250797 Page 8 of 10
Traditional 510(k) Notification
UV5000 Handle
| | | exposed surfaces of the handle. | |
| --- | --- | --- | --- |
| Frequency of use | Reuseable | Reusable | Same |
| Duration of use | < 24 hours | < 24 hours | Same |
| Location of use | Hospitals and physician offices | Hospitals and physician offices | Same |
| Mechanism of Action | | | |
| Operating Principles | Rather than run a cable from the cystoscope to a separate video monitor as is required for all other cystoscopes, the LCD monitor is integrated into the body of the handle of the UroViu cystoscope system so that the user can see images of the tissue and anatomical structures directly and in real time as the cystoscope is being advanced to and within the bladder. | Rather than run a cable from the cystoscope to a separate video monitor as is required for all other cystoscopes, the LCD monitor is integrated into the body of the handle of the UroViu cystoscope system so that the user can see images of the tissue and anatomical structures directly and in real time as the cystoscope is being advanced to and within the bladder. | Same |
| Performance Testing | | | |
| Bench testing for Image Quality | Pass | Pass | Same |
| Battery change and discharge test | Pass | Pass | Same |
| 3 Year service life | Pass | Pass | Same |
| Software (SW) | | | |
| SW version | 1.0 | 4.1 | This SW version was developed separately but it used version 4.1 |
{12}
UroViu Corporation
K250797 Page 9 of 10
Traditional 510(k) Notification
UV5000 Handle
| | | | for the 4500 Handle as reference. |
| --- | --- | --- | --- |
| Level of Concern | Moderate | Moderate | Same |
| General functions of the software | • Initialize and setup video processing software
• Initialize LCD display control
• Initialize user interface (menus, touch screen)
• Initialize recording, storage, and playback of video and still images | • Initialize and setup video processing software
• Initialize LCD display control
• Initialize user interface (menus, touch screen)
• Initialize recording, storage, and playback of video and still images | Same |
## Performance Data
All necessary performance testing as conducted:
- Field of view accuracy
- Direction of view
- Depth of Field
- Noise and Dynamic Range
- Image distortion
- Image resolution
- Image uniformity
- Color performance
- Electrical safety and electromagnetic compatibility testing
The device was found to meet the compliance requirements for electrical safety as specified in ISO 60601-1 including provisions for EMC safety in ISO 60601-1-2 and IEC 60601-2-18 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment, including thermal safety. Mechanical characteristics were also tested, with successful results.
{13}
UroViu Corporation
K250797 Page 10 of 10
Traditional 510(k) Notification
UV5000 Handle
## Preclinical (Animal) Studies
Bench testing was sufficient to demonstrate performance of the device. No preclinical testing of the subject device was necessary.
## Clinical Studies
Bench testing was sufficient to demonstrate performance of the device. No clinical testing of the subject device was necessary.
## Conclusion
The information submitted in this premarket notification confirms that the minor differences between the UV5000 Handle and the predicate raise no new questions of safety and effectiveness. Therefore, the UV5000 Handle is substantially equivalent to the identified predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.