Lightin System, Lightin, Lightin Generator

K232177 · Solmedix Co., Ltd. · SBS · Jul 10, 2024 · Ear, Nose, Throat

Device Facts

Record IDK232177
Device NameLightin System, Lightin, Lightin Generator
ApplicantSolmedix Co., Ltd.
Product CodeSBS · Ear, Nose, Throat
Decision DateJul 10, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4420
Device ClassClass 1

Indications for Use

Lightin is a needle for injection laryngoplasty with an optical fiber. Lightin Generator is an AC-powered light source including Light Guide which is fiber optic cable that transmits light at the tip of fiber optics embedded within Lightin. Lightin System is used for injection laryngoplasty to address glottic incompetence to indicate the injection site with visible light.

Device Story

Lightin System comprises a light-guided needle (Lightin) and an AC-powered LED light source (Lightin Generator). The system is used by clinicians during injection laryngoplasty procedures. The Lightin Generator transmits light through a fiber optic cable (Light Guide) to a PMMA fiber embedded within the needle. The light emitted from the needle tip provides visual guidance to the clinician to identify the precise injection site within the vocal fold. This visual feedback assists the clinician in accurately placing injection material to treat glottic incompetence. The device is disposable (needle) and reusable (generator). It does not deliver drugs. The system benefits patients by improving the accuracy of laryngeal injections.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -11, -23, USP <151>), electrical safety/EMC (IEC 60601-1, IEC 60601-1-2), and non-clinical performance testing (ISO 80369-7, ISO 80369-20, ISO 7864, ISO 9626). Performance validation confirmed the system successfully directs injection material to the target area and evaluated tissue heating concerns. Photobiological safety was verified per IEC 62471.

Technological Characteristics

Needle: 304 stainless steel, 23G/25G, disposable, EO sterilized. Fiber: PMMA with fluorinated polymer cladding, 0.125mm OD. Generator: AC-powered LED light source, output ≤ 8mW. Connectivity: Light Guide cable. Standards: ISO 10993, IEC 60601-1, IEC 60601-1-2, ISO 80369-7, ISO 7864, ISO 9626, IEC 62471.

Indications for Use

Indicated for use in injection laryngoplasty to address glottic incompetence by providing visible light to indicate the injection site.

Regulatory Classification

Identification

An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. July 10, 2024 Solmedix Co., Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Ste 610 Houston, Texas 77054 Re: K232177 Trade/Device Name: Lightin System (Lightin, Lightin Generator) Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I Product Code: SBS, EQH Dated: June 12, 2024 Received: June 12, 2024 Dear Dave Kim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Joyce C. Lin -S for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices {2}------------------------------------------------ OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K232177 Device Name Lightin System (Lightin, Lightin Generator) #### Indications for Use (Describe) Lightin is a needle for injection laryngoplasty with an optical fiber. Lightin Generator is an AC-powered light source including Light Guide which is fiber optic cable that transmits light at the tip of fiber optics embedded within Lightin. Lightin System is used for injection laryngoplasty to address glottic incompetence to indicate the injection site with visible light. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------|---------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Solmedix. The logo features the word "solmedix" in gray, with a yellow accent mark on the left side of the word and a yellow trademark symbol on the right side. The font is sans-serif and the overall design is simple and modern. #706, 330, Seongarm-ro, Map 03920 Rep. of Korea Tel +82-2-717-7852 # 510(k) Summary # K232177 This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92. 1. Date: June 4. 2024 ## 2. Manufacturer and 510(k) Sponsor - . Name: Solmedix Co., Ltd. - . Address: #706, 330, Seongarm-ro, Mapo-gu, Seoul, 03920 Republic of Korea - . Contact Name: Hyeroung Lee (Ms) / RA Part Leader - Telephone number: +81-2-3153-7717 ● - . E-mail address: hrlee@solmedix.com ### 3. Submission Correspondent - Name: Dave Kim, MBA ● - . Company Name: Mtech Group LLC - Address: 7505 Fannin St. Suite 610, Houston, TX 77054, USA - Telephone number: 713-467-2607 - . E-mail address: davekim@mtechgroupllc.com ## 4. Device Identification The subject device is comprised of the following components: | Trade Name | Lightin | |-----------------------------|--------------------------| | Device Classification Name | Set, Laryngeal Injection | | Regulation Number | 21CFR§874.4420 | | Regulation Class | Class I | | Classification Product Code | SBS | | Classification Pannel | Ear Nose & Throat | | Trade Name | Lightin Generator | |-----------------------------|-----------------------------------| | Device Classification Name | Source, Carrier, Fiberoptic Light | | Regulation Number | 21CFR§874.4350 | | Regulation Class | Class I | | Classification Product Code | EQH | | Classification Pannel | Ear Nose & Throat | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "solmedix" in gray letters. The "s" and the "x" are connected to yellow accents. The yellow accent on the left is a vertical line, and the yellow accent on the right is a curved vertical line. ## 5. Legally Marketed Predicate Device Predicate Device #1 | Manufacturer | Aspen Surgical Products, Inc. | |----------------------|-------------------------------| | Trade Name | Symmetry Surgical | | 510(k) | 510K Exempt (CL 1) | | Regulation Number | 21CFR§874.4420 | | Regulation Class | Class I | | Product Code | KAA | | Classification Panel | Ear Nose & Throat | #### Predicate Device #2 | Manufacturer | AESCULAP AG | |-----------------------------|----------------------------| | Trade Name | AESCULAP® LED LIGHT SOURCE | | 510(k) | 510K Exempt (CL 1) | | Regulation Number | 21CFR&874.4350 | | Regulation Class | Class I | | Classification Product Code | EQH | | Classification Pannel | Ear Nose & Throat | #### Reference Device #1 | Manufacturer | Peregrine Surgical Ltd | |-----------------------|--------------------------------------------------| | Trade Name | 23ga and 25 ga Adjustable Chandelier Illuminator | | 510(k) | K151604 | | Regulation Number | 21CFR&876.1500 | | Regulation Class | Class II | | Product Code | MPA | | Classification Pannel | Ear Nose & Throat | Reference Device #2 | Manufacturer | Zhejiang Kindly Medical Device Co., Ltd. | |-----------------------|-------------------------------------------| | Trade Name | Sterile Hypodermic Needles for Single Use | | 510(k) | K233037 | | Regulation Number | 21 CFR&880.5570 | | Regulation Class | Class II | | Product Code | FMI | | Classification Pannel | General hospital | ## 6. Device Description Lightin System consists of Lightin and Lightin Generator. Lightin is a light-guided needle used injection laryngoplasty and Light generator delivers the light energy to Lightin by generating LED light from an AC-powered light source. Page 2 / 13 {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for "solmedix". The word "solmedix" is written in gray letters. There is a yellow bracket on the left side of the word and a yellow design on the right side of the word. #706, 330, Seongarm-ro, Mapo-qu, Seoul 03920 Rep. of Korea Tel +82-2-717-7852 Lighin consists of injection needle, Y-connector and SMA connector/cable. The light transmitted from the Lightin Generator is emitted from the tip of fiber optics embedded within Lightin. Lightin system used for injection laryngoplasty procedures to indicate the injection site with visible light. This system does not contain or deliver drugs. ### 7. Indications for Use Lightin is a needle for injection laryngoplasty with an optical fiber. Lightin Generator is an AC-powered light source including Light Guide which is fiber optic cable that transmits light at the tip of fiber optics embedded within Lightin. Lightin System is used for injection laryngoplasty to address glottic incompetence to indicate the injection site with visible light. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Solmedix, followed by their address: #706, 330, Seongarm-ro, Mapo-gu, Seoul. The image also includes the postal code, 03920, and the country, Rep. of Korea. Finally, the image shows the telephone number: +82-2-717-7852. # 8. Comparison Table of Technological Characteristics with the predicated device - A. Lightin | Attribute | Predicate Device #1 | Proposed device | Remark | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product name | Symmetry Surgical | Lightin,<br>Lightin System | | | 510(k) no | 510K Exempt (CL 1) | K232177 | | | Product Code | KAA | SBS | | | Manufacturer | Aspen Surgical Products, Inc. | Solmedix Co., Ltd | | | Indications for use | An ear, nose, and throat manual<br>surgical instrument is one of a variety<br>of devices intended for use in surgical<br>procedures to examine or treat larynx.<br>This generic type of device includes<br>laryngeal injection set. | Lightin is a needle for injection<br>laryngoplasty with an optical fiber.<br>Lightin generator is an AC-powered<br>light source including Light Guide<br>which is fiber optic cable that<br>transmits light at the tip of fiber<br>optics embedded within Lightin.<br>Lightin System is used for injection<br>laryngoplasty to address glottic<br>incompetence to indicate the<br>injection site with visible light. | Similar<br>Predicate Device #1 and Lightin is the same<br>as that used for laryngeal injection.<br>Lightin uses LED light to indicate the injection<br>site and the LED source emits light through<br>the PMMA fiber in the needle. Considering<br>the application area and the indications for<br>use, Lightin has less risk than the predicate<br>device. | | Appearance | No information | Image: A blue fiber optic cable | Similar<br>The needle tip resembles an injection needle,<br><br>Different<br>inclusion of a light-transmitting fiber | | Jacket Material | N/A | PVC | Different<br>Fiber optic cable is not included in the primary<br>predicate device (#1).<br>The subject device has fiber optics inserted in<br>the needle. The jacket materials protecting<br>the fiber optic of both devices are commonly<br>used. | | Jacket Color | N/A | Blue<br>Orange | Similar<br>Jacket is color coded for needle identification.<br>Color coding was applied to Lightin according<br>to the needle gauge standard. | | Illumination Fiber -<br>OD | N/A | 0.125mm (0.004521") | Different<br>PD #1: No optic fiber<br>Lightin uses LED light to indicate the injection<br>site and the LED source emits light through<br>the PMMA fiber in the needle. Considering<br>the application area and the indications for<br>use, Lightin has less risk than the predicate<br>device. | | Illumination Fiber<br>Material | N/A | PMMA<br>(Polymethyl methacrylate with<br>Fluorinated Polymer Cladding) | Different | | Needle size | No information | 1. 23G<br>2. 25G | | | Needle Material | No information | 304 stainless Steel | | | Reusable/<br>Disposable | No information | Disposable | | | Sterilization | No information | EO | | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for "solmedix". The word "solmedix" is written in gray letters. There is a yellow bracket on the left side of the word and a yellow bracket on the right side of the word. Tel +82-2-717-7852 Page 5 / 13 {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for "solmedix". The word "solmedix" is written in gray letters. There is a yellow bracket on the left side of the word and a yellow "J" on the right side of the word. Tel +82-2-717-7852 ### Reference Device #1: K151604 | Attribute | Reference device #1 | Proposed device | Remark | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product name | 23ga and 25ga Adjustable Chandelier<br>Illuminator | Lightin,<br>Lightin System | | | 510(k) no | K151604 | K232177 | | | Product Code | MPA | SBS | | | Manufacturer | Peregrine Surgical, Ltd | Solmedix Co., Ltd | | | Indications for use | Adjustable Chandelier Illuminator family<br>of ophthalmic illuminators is for wide<br>angle illumination during ophthalmic<br>surgery. | Lightin is a needle for injection<br>laryngoplasty with an optical<br>fiber.<br>Lightin generator is an AC-<br>powered light source including<br>Light Guide which is fiber optic<br>cable that transmits light at the tip<br>of fiber optics embedded within<br>Lightin.<br>Lightin System is used for<br>injection laryngoplasty to address<br>glottic incompetence to indicate | <b>Similar</b><br>For the reference #1, the main Indication for use is<br>for wide-angle illumination during ophthalmic<br>surgery. However, Lightin uses LED light to<br>indicate the injection laryngoplasty site. Although<br>there is a difference such as in the indication for<br>use, Reference #1 and Lightin have similar<br>technical characteristics in that the LED source<br>emits light through the PMMA fiber in the needle. | | | | the injection site with visible light. | | | | | | | | Appearance | Image: Synergetics needle in packaging | Image: Blue needle | Similar<br>For Reference #1, its needle tip resembles an<br>injection needle, and its inclusion of a light-<br>transmitting fiber is similar. | | Jacket Material | PTFE (Polytetrafluoroethylene) | PVC | Similar<br>The Reference #1 device has fiber optics inserted<br>in the needle, same as Lightin. The jacket<br>materials protecting the fiber optic of the two<br>devices are different, but both are commonly used. | | Jacket Color | Black/Green<br>Black/Blue | Blue<br>Orange | Similar<br>Jacket is color coded for needle identification.<br>Color coding was applied to Lightin according to<br>the needle gauge standard. | | Illumination Fiber -<br>OD | 0.015" | 0.125mm (0.004521") | Different<br>The difference of the illumination fiber outer<br>diameter (O.D.) is due to the functional | | | | | characteristics of the two devices. Reference #1 is<br>used for wide-angle illumination and Lightin is used<br>for marking the injection laryngoplasty site. Due to<br>this functional difference, the illumination fiber OD<br>of the reference #1 is bigger than that of Lightin. | | Illumination Fiber<br>Material | PMMA<br>(Polymethyl methacrylate with<br>Fluorinated Polymer Cladding) | PMMA<br>(Polymethyl methacrylate with<br>Fluorinated Polymer Cladding) | Same | | Needle size | 1. 23G<br>2. 25G | 1. 23G<br>2. 25G | Same | | Needle Material | 304 stainless Steel | 304 stainless Steel | Same | | Reusable/<br>Disposable | Disposable | Disposable | Same | | Sterilization | EO | EO | Same | Page 6 / 13 {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "solmedix" in gray font. There is a yellow bracket on the left side of the word and a yellow bracket on the right side of the word. The brackets are not complete, and they are only partially visible. Tel +82-2-717-7852 Page 7 / 13 {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "solmedix" in gray font. There is a yellow bracket on the left side of the word and a yellow "j" on the right side of the word. The font is sans-serif and the background is white. 03920 Rep. of Korea Tel +82-2-717-7852 Reference Device #2: K233037 | Attribute | Reference Device #2 | Proposed device | Remark | |---------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Product name | Sterile Hypodermic Needles for<br>Single Use | Lightin,<br>Lightin System | | | 510(k) no | K233037 | K232177 | | | Product Code | FMI | SBS | | | Manufacturer | Zhejiang Kindly Medical Device Co.,<br>Ltd. | Solmedix Co., Ltd | | | Indications for use | Sterile Hypodermic Needle for Single<br>Use are intended for use with<br>syringes and injection devices for | Lightin is a needle for injection<br>laryngoplasty with an optical fiber.<br>Lightin generator is an AC-powered light | Similar<br>Reference Device #2 and Lightin are<br>same as injection needles for a single | Page 8 / 13 {12}------------------------------------------------ Image /page/12/Picture/0 description: The image contains the word "solmedix" in gray font. There is a yellow bracket on the left side of the word and a yellow bracket on the right side of the word. The right bracket has a trademark symbol next to it. Tel +82-2-717-7852 | | general purpose fluid<br>injection/aspiration. | source including Light Guide which is fiber<br>optic cable that transmits light at the tip of<br>fiber optics embedded within Lightin.<br>Lightin System is used for injection<br>laryngoplasty to address glottic<br>incompetence to indicate the injection site<br>with visible light. | use.<br>Lightin is used for laryngeal injection and<br>RD #2 is used as an injection device for<br>general purpose.<br>Lightin has narrower indication of use<br>than RD#2. | |------------------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Needle size | 14G~34G | 1. 23G<br>2. 25G | Similar | | Needle Material | 304 Stainless Steel | 304 Stainless Steel | Same | | Reusable/<br>Disposable | Disposable | Disposable | Same | | Sterilization | EO | EO | Same | | Performance<br>Specification | ISO7864<br>ISO9626<br>ISO80369-7<br>ISO8039-20 | ISO7864<br>ISO9626<br>ISO80369-7<br>ISO8039-20 | Same | ### B. Lightin Generator | Attribute | Predicate device #2 | Proposed device | Remark | |--------------|----------------------------|--------------------------------------|--------| | Product name | AESCULAP® LED LIGHT SOURCE | Lightin Generator,<br>Lightin System | | | 510(k) no | 510K Exempt (CL 1) | K232177 | | | Product code | EQH | EQH | Same | | Manufacturer | AESCULAP, AG. | Solmedix co., Ltd | | {13}------------------------------------------------ Image /page/13/Picture/0 description: The image contains the word "solmedix" in gray font. The first and last letters of the word are enclosed in yellow brackets. The brackets are slightly raised above the word. 03920 Rep. of Korea | | | | Similar | |----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for use | The device is intended for use in<br>endoscopy systems for human<br>medicine. In combination with a light<br>lead, endoscope, camera and monitor,<br>its purpose is to illuminate the inside of<br>the human body. | Lightin generator is an AC-powered light<br>source including Light Guide which is fiber<br>optic cable that transmits light at the tip of<br>fiber optics embedded within Lightin. | Predicate Device #2 and Lightin<br>Generator are the same light sources that<br>generate LED light.<br><br>Lightin Generator, its intended use is<br>restricted solely to Lightin for injection<br>laryngoplasty.<br><br>Lightin Generator is used to transmits<br>light at the tip of fiber optics embedded<br>within Lightin for injection laryngoplasty.<br>Both PD #2 are technically identical with<br>Lightin Generator. | | Appearance | Image: Endoscopy system | Image: Lightin generator | Similar | | Light Source | LED | LED | Same | | Design | Dimension:<br>$330 * 146 * 367$ mm<br>Weight: 8.5kg | Dimension:<br>$300 * 200 * 100$ mm<br>Weight: 2.7kg(excluding adapter) | - | | Accessories | Image: Sterile adapter | Adapter(Power Cord) | Same<br>PD #2 has a cable to connect to the LED<br>light source, and Lightin Generator has a<br>Light Guide for the same purpose.<br><br>Different<br>The adapter of PD #2 is compatible to<br>connect various optical fiber cables to the | | | Optical Cable<br>Power cord | | LED light source, but the Lightin<br>Generator is exclusive to Lightin, so no<br>compatible parts are required. | | Technical<br>Specification | Output (at the end light guide) : 240W | Output (at the end light guide): $\le$ 8mW | Different<br>The output of PD #2 is higher than that<br>of Lightin Generator because PD #2<br>uses the LED light to illuminate the board<br>surgical site during surgical procedures.<br>However, Lightin Generator needs much<br>lower output than PD #2 because Lightin<br>Generator uses the LED light to indicate<br>the small injection laryngoplasty site and<br>the low output of Lightin Generator is<br>technically sufficient for use. | | Safety | IEC60601-1<br>IEC60601-1-2 | IEC60601-1<br>IEC60601-1-2 | Same | | Reusable/Disposable | Reusable | Reusable | Same | | Sterilization | N/A | N/A | Same | Page 10 / 13 {14}------------------------------------------------ Image /page/14/Picture/0 description: The image contains the word "solmedix" in gray font. The first and last letters of the word are enclosed in yellow brackets. The brackets are slightly raised above the word. 03920 Rep. of Korea Tel +82-2-717-7852 {15}------------------------------------------------ Image /page/15/Picture/0 description: The image shows the logo for Solmedix. The word "solmedix" is written in gray letters. There is a yellow bracket on the left side of the word and a yellow "J" on the right side of the word. #706, 330, Seongarm-ro, 03920 Rep. of Korea Tel +82-2-717-7852 ## 9. Conclusion on SE based on Technical Comparison and Performance Data The following performance testing has been performed for the proposed device; ### Biocompatibility Testing Lightin was accessed for conformity with the requirements of ISO 10993 series. - ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity (Recognition No: 2-245) - ISO 10993-10: 2021 Biological evaluation of medical devices Part 10: Tests for skin ● sensitization (Recognition No: 2-296) - ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation (Recognition No: 2-291) - ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity (Recognition No: 2-255) - . USP <151> Pyrogen Test (Recognition No:2-292) #### Electrical Safety & EMC Testing Lightin Generator was accessed for conformity with the requirements of IEC series. - IEC 60601-1:2005/AMD2:2012 Edition 3.0 2012 MEDICAL ELECTRICAL EQUIPMENT ● - Part 1: General requirements for safety and essential performance. (Recognition No: 19-4) - IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests (Recognition No: 19-8) - IEC TR 60601-4-2 Edition 1.0 2016-05 Medical electrical equipment Part 4-2: Guidance ● and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems (Recognition No: 19-19) #### Non-Clinical Performance Testing Lightin was accessed for conformity with the below requirements. - ISO 80369-7: 2021 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications (Recognition No: 5-133) - ISO 80369-20: 2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods (Recognition No: 5-97) {16}------------------------------------------------ Image /page/16/Picture/0 description: The image contains the logo for "solmedix". The logo is in gray, except for the left bracket and the trademark symbol, which are in yellow. The text is in a sans-serif font and is slightly blurred. #706, 330, Seongarm-ro, Mapo-qu, Seoul 03920 Rep. of Korea Tel +82-2-717-7852 - . ISO 7864: 2016 Sterile hypodermic needles for single use - Requirements and test methods (Recognition No : 6-379) - ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods(Recognition No: 6-380) - ISO 11135: 2014 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)] (Recognition No: 14-529) - ASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices (Recognition No: 14-497) - ASTM F1929-15: Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration (Recognition NO: 14-484) - ASTM F88 / F88M-15: Standard Test Method for Seal Strength of Flexible Barrier ● Materials (Recognition No: 14-482) - USP <71> Sterility Test (Recognition No: 14-569) - Internal test standard: light output test , Thermal measurement test - ASTM D4169-22: Standard Practice Performance Testing of Shipping Containers and ● Systems (Recognition No: 14-576) #### Performance Testing Lightin System was accessed for conformity with the below requirements. - . The Lightin System successfully directs the laryngoplastic injection material to the target area within the vocal fold. - . The tissue heating concerns by the tip illumination. Lightin and Lightin Generator was accessed for conformity with the below requirements. - IEC 62471: 2006 Photobiological safety of lamps and lamp systems (Recognition No; 12-● 249) - . Internal Test Standard: Light Output Test ### 10. Conclusions Based on a comparison of the various predicate devices and the reference device with the proposed Lightin System in terms of indication for use, technological characteristics, material and performance specifications, Lightin System raises no new issues of safety and effectiveness and is substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...