K190525 · Acclarent, Inc. · LRC · May 3, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K190525
Device Name
RELIEVA ULTIRRA Sinus Balloon Catheter
Applicant
Acclarent, Inc.
Product Code
LRC · Ear, Nose, Throat
Decision Date
May 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic, Pediatric
Intended Use
The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures. For children aged 17 and under, the balloon catheter system is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures. The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT intended to irrigate from within a target sinus for therapeutic procedures nor to facilitate diagnostic procedures with TruDi™ NAV Wire.
Device Story
Flexible sinus balloon catheter; used to dilate sinus ostia and paranasal spaces; facilitates irrigation. Features hypotube on proximal end for rigidity during insertion. Manually operated by physician; balloon inflated with sterile saline or water. Compatible with TruDi™ NAV Wire and TruDi™ Navigation System for anatomical access and placement confirmation. When used with navigation wire, irrigation is contraindicated. Benefits patient by enabling minimally invasive sinus dilation and diagnostic access. Used in clinical settings.
Clinical Evidence
No clinical data was necessary. Substantial equivalence supported by bench testing, including simulated use, dimensional specifications, balloon burst pressure, joint separation force, deflation time, and cycle fatigue. Biocompatibility per ISO 10993-1 and sterilization validation per AAMI/ANSI/ISO 11135:2014.
Technological Characteristics
Nylon balloon; dual lumen shaft; 250mm length; 12 ATM max inflation pressure. Compatible with 0.035" guidewires. Materials identical to predicate. Sterilized via Ethylene Oxide (EtO). Packaging: HDPE backer card in Tyvek/Nylon pouch. Standards: ASTM F1980-07 (shelf life), ISO 10993-1 (biocompatibility), ISO 10993-7:2008 (EtO residuals).
Indications for Use
Indicated for dilation of sinus ostia and paranasal sinus spaces for diagnostic and therapeutic procedures in adults and children aged 17 and under (maxillary sinus only for children). Includes irrigation capability. Compatible with TruDi™ Navigation System for access and placement confirmation.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Predicate Devices
RELIEVA Solo Elite™ Sinus Balloon Catheter (K111254)
K111254 — RELIEVA SOLO ELITE SINUS BALLOON CATHETER · Acclarent, Inc. · Aug 22, 2011
K161698 — Relieva UltirraNav Sinus Balloon Catheter · Acclarent, Inc. · Oct 24, 2016
K143738 — DSS Sinusplasty Balloon Catheter · Intuit Medical Products, LLC · Aug 27, 2015
K201115 — Next Generation Balloon Dilation System · Acclarent, Inc. · Aug 27, 2020
K142252 — XprESS Multi-Sinus Dilation Tool · Entellus Medical, Inc. · Oct 17, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2019
Acclarent, Inc. Leena Sorathia Sr. Regulatory Affairs Specialist 33 Technology Drive Irvine, California 92618
Re: K190525
Trade/Device Name: RELIEVA ULTIRRA Sinus Balloon Catheter Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, nose, and throat manual surgical instrument Regulatory Class: Class I Product Code: LRC Dated: March 1, 2019 Received: March 4, 2019
Dear Leena Sorathia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### K190525
Device Name RELIEVA ULTIRRA® Sinus Balloon Catheter
#### Indications for Use (Describe)
The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.
For children aged 17 and under, the balloon catheter system is intended to dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures. It is also intended to irrigate from within a target sinus for therapeutic procedures and to facilitate diagnostic procedures.
The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT intended to irrigate from within a target sinus for therapeutic procedures nor to facilitate diagnostic procedures with TruDi™ NAV Wire.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple swoosh above the "ent" part of the word.
# 510(K) SUMMARY
### [807.92(a)(1)] Submitter Information
| Sponsor/Submitter: | Acclarent, Inc.<br>33 Technology Drive<br>Irvine, CA 92618 |
|--------------------|---------------------------------------------------------------------------------------------------------|
| Contact Person: | Leena Sorathia<br>Sr. Regulatory Affairs Specialist<br>Email: lsorathi@its.jnj.com<br>Tel: 949-923-4118 |
#### Date Summary Prepared: May 3, 2019
### [807.92(a)(2)] Name of Device
| Device Trade<br>Name: | RELIEVA ULTIRRA® Sinus Balloon Catheter |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Name: | Ear, Nose, and Throat Manual Surgical Instrument |
| Common Name: | Sinus Balloon Catheter |
| Device<br>Classification: | Class I |
| Regulation<br>Number: | 21 CFR 874.4420 |
| Review Panel: | Ear, Nose, and Throat |
| Product Code: | LRC |
| [807.92(a)(3)] Legally Marketed Devices | |
| Predicate Device: | RELIEVA Solo Elite™ Sinus Balloon Catheter (K111254) |
| Reference Device: | RELIEVA UltirraNav® Sinus Balloon Catheter, (K161698) |
| [807.92(a)(4)] Device Description | |
| Device Description: | The RELIEVA ULTIRRA® Sinus Balloon Catheter is a flexible catheter that is<br>intended to dilate sinus ostia. The shaft allows for inflation of the sinus balloon and<br>permits the passage of an electromagnetic navigable sinus guidewire or sinus<br>illumination system to facilitate access to the target sinus ostia. A hypotube is |
incorporated on the proximal end to provide rigidity during insertion through a sinus guide catheter.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Acclarent. The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "rent" portion in purple. Above the purple part of the word is a curved line, also in purple, that arches over the text.
# [807.92(a)(5)] Intended Use
| Indications for Use: | The RELIEVA ULTIRRA® Sinus Balloon Catheter is an instrument intended to<br>dilate sinus ostia and spaces within the paranasal sinus cavities for diagnostic and<br>therapeutic procedures. It is also intended to irrigate from within a target sinus for<br>therapeutic procedures and to facilitate diagnostic procedures. |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | For children aged 17 and under, the balloon catheter system is intended to dilate sinus<br>ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic<br>procedures. It is also intended to irrigate from within a target sinus for therapeutic<br>procedures and to facilitate diagnostic procedures. |
| | The RELIEVA ULTIRRA® Sinus Balloon Catheter may be utilized in conjunction<br>with TruDi™ NAV Wire and TruDi™ Navigation System to provide access to nasal<br>and sinus spaces, and to confirm placement in the accessed anatomy. It is NOT<br>intended to irrigate from within a target sinus for therapeutic procedures nor to<br>facilitate diagnostic procedures with TruDi™ NAV Wire. |
| Difference in<br>Indications from<br>Predicate Device | The indications for use statement of the subject device, RELIEVA ULTIRRA® Sinus<br>Balloon Catheter, is similar to predicate device, RELIEVA Solo Elite™ Sinus Balloon<br>Catheter (K111254). The subject and predicate devices are both intended to access and<br>dilate the sinus ostia and paranasal spaces using a balloon. |
| | The proposed revised indications for use for the subject device are largely unchanged<br>from the predicate device with the exception of the additional statement that the<br>device may be used in conjunction with TruDi™ NAV Wire and TruDi™ Navigation<br>System. In addition, the irrigation function of the subject device is not intended to be<br>used in conjunction with the TruDi™ NAV Wire. |
| | The changes in indications for use of the subject device do not raise any significant<br>issues of safety and effectivenessof the device when used as labeled, as demonstrated b<br>performance testing. |
| | The difference in indications for use between the subject device and the<br>predicate/reference devices is supported is presented in Table 1 of this summary. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Acclarent" in a stylized font. The first part of the word, "Acclar" is in black, while the "ent" is in purple. There is a purple curved line above the "ent" portion of the word.
# [807.92(a)(6)] Technical Characteristics
| Technological<br>Characteristics: | The design and materials of the subject device, RELIEVA ULTIRRA® Sinus<br>Balloon Catheter, are identical to the predicate device RELIEVA Solo Elite™ Sinus<br>Balloon Catheter (K111254). The added capability of the subject device to be used<br>with the TruDi™ NAV Wire and TruDi™ Navigation System did not require any<br>design changes from the predicate device. |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Like the predicate device, the subject device has the same ballon dilation pressure,<br>material composition, and size specifications. However, the subject device does not<br>allow for the irrigation feature. |
| | The subject device is substantially equivalent in technological characteristics, as<br>there are no significant differences in design, fundamental scientific technology, or<br>other features of the device as compared to the predicate device. |
| | See Table 1 for a comparison of the technological characteristics between the<br>subject device and the predicate/reference devices. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "Acclarent." The word "Acclarent" is written in a sans-serif font, with the first part of the word in black and the "ent" portion in purple. There is a purple curved line above the "ent" portion of the word, and a trademark symbol to the right of the "t".
| Attribute | Primary Predicate Device: | Reference Device: | Subject Device: |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | RELIEVA Solo Elite™ Sinus Guide | RELIEVA ULTIRRA® Nav Sinus | RELIEVA ULTIRRA® Sinus Balloon |
| | Catheter (K111254) | Balloon Catheter (K161698) | Catheter (K190525) |
| Manufacturer | Acclarent | Acclarent | Acclarent |
| Common Name | Sinus Balloon Dilation System | Sinus Balloon Dilation System | Sinus Balloon Dilation System |
| Class | I | I | I |
| Product Code<br>Classification<br>Section | LRC<br>21 CFR 874.4420 | LRC<br>21 CFR 874.4420 | LRC<br>21 CFR 874.4420 |
| Indications for Use | The RELIEVA Ultirra® Sinus Balloon<br>Catheter is an instrument intended to dilate<br>sinus ostia and spaces within the paranasal<br>sinus cavities for diagnostic and<br>therapeutic procedures. It is also intended<br>to irrigate from within a target sinus for<br>therapeutic procedures and to facilitate<br>diagnostic procedures.<br>For children aged 17 and under, the<br>balloon catheter system is intended to<br>dilate sinus ostia and spaces associated<br>with the maxillary sinus for diagnostic and<br>therapeutic procedures. It is also intended<br>to irrigate from within a target sinus for<br>therapeutic procedures and to facilitate<br>diagnostic procedures. | The RELIEVA ULTIRRA® Nav Sinus<br>Balloon Catheter is an instrument intended<br>to dilate the sinus ostia and spaces within<br>the paranasal sinus cavities for diagnostic<br>and therapeutic procedures.<br>For children aged 17 and under, the balloon<br>catheter system is intended to dilate sinus<br>ostia and spaces associated with the<br>maxillary sinus for diagnostic and<br>therapeutic procedures.<br>The RELIEVA ULTIRRA® Sinus Balloon<br>Catheter may be utilized in conjunction<br>with the Acclarent NavWire™ Sinus<br>Navigation Guidewire and ACCLARENT®<br>ENT Navigation System to provide access to<br>nasal and sinus spaces, and to confirm<br>placement in the accessed anatomy. | The RELIEVA ULTIRRA® Sinus Balloon<br>Catheter is an instrument intended to dilate<br>sinus ostia and spaces within the paranasal<br>sinus cavities for diagnostic and therapeutic<br>procedures. It is also intended to irrigate from<br>within a target sinus for therapeutic procedures<br>and to facilitate diagnostic procedures.<br>For children aged 17 and under, the balloon<br>catheter system is intended to dilate sinus ostia<br>and spaces associated with the maxillary sinus<br>for diagnostic and therapeutic procedures. It is<br>also intended to irrigate from within a target<br>sinus for therapeutic procedures and to<br>facilitate diagnostic procedures.<br>The RELIEVA ULTIRRA® Sinus Balloon<br>Catheter may be utilized in conjunction with<br>TruDi™ NAV Wire and TruDi™ Navigation<br>System to provide access to nasal and sinus<br>spaces, and to confirm placement in the<br>accessed anatomy. It is NOT intended to<br>irrigate from within a target sinus for<br>therapeutic procedures nor to facilitate<br>diagnostic procedures with TruDi™ NAV Wire. |
| Indicated for<br>Children | Yes, for children 17 years and under in<br>maxillary sinus only | Yes, for children 17 years and under in<br>maxillary sinus only | Yes, for children 17 years and under in<br>maxillary sinus only |
| Attribute | Primary Predicate Device:<br>RELIEVA Solo Elite™ Sinus Guide | Reference Device:<br>RELIEVA ULTIRRA® Nav Sinus | Subject Device:<br>RELIEVA ULTIRRA® Sinus Balloon |
| | Catheter (K111254) | Balloon Catheter (K161698) | Catheter (K190525) |
| Single Patient Use | Yes | Yes | Yes |
| Direct Patient<br>Contact | Yes | Yes | Yes |
| Labeled as Non-<br>Pyrogenic? | No | No | No |
| Technological<br>Characteristics | Allows for dilation of sinus ostia with the<br>added capability to irrigate. | Allows for dilation of sinus ostia. | Allows for dilation of sinus ostia with the<br>added capability to irrigate when used with the<br>RELIEVA LUMA SENTRY®. It is NOT<br>intended to irrigate from within a target sinus<br>for therapeutic procedures nor to facilitate<br>diagnostic procedures with TruDi™ NAV Wire. |
| Constructed of<br>Materials<br>Commonly Used<br>in Patient<br>Contacting<br>Medical Devices | Yes | Yes | Yes |
| Balloon Material | Nylon | Nylon | Nylon |
| Balloon Diameter<br>and Length | 5x16mm, 6x16mm, 7x16mm and 7x24mm | 5x16mm | 5x16mm, 6x16mm, 7x16mm and 7x24mm |
| Maximum<br>Inflation Pressure | 12 ATM | 12 ATM | 12 ATM |
| Flexible Shaft | Yes | Yes | Yes |
| Deflation Time | < 5 seconds | < 3.6 seconds | < 5 seconds |
| Catheter Length | 250 ± 2 mm | 250 ± 2mm | 250 ± 2mm |
| Shaft Design | Dual Lumen | Dual Lumen | Dual Lumen |
| Used with<br>Guidewire | Yes…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.