PeriView FLEX

K171232 · Olympus Surgical Technologies America · KTI · Sep 21, 2017 · Ear, Nose, Throat

Device Facts

Record IDK171232
Device NamePeriView FLEX
ApplicantOlympus Surgical Technologies America
Product CodeKTI · Ear, Nose, Throat
Decision DateSep 21, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2

Indications for Use

This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use.

Device Story

PeriView FLEX is a single-use aspiration needle for intrapulmonary tissue collection. Device comprises handle, sheath, needle, and removable stylet. Physician inserts distal end through bronchoscope working channel to target site; manually operates needle slider to extend/retract needle for tissue sampling. Sample expelled via air-filled syringe or stylet. Used in surgical, endoscopy, or bronchoscopy suites. Ergonomic handle design improves physician interaction; device reaches upper lung lobes. Benefits include minimally invasive tissue acquisition for diagnostic purposes.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), and performance verification (insertion/withdrawal forces, puncture force, handle durability, and vacuum decay).

Technological Characteristics

Materials: Stainless steel, PTFE, PEBAX, Nitinol. Manual actuation. 21G needle. Ethylene oxide sterilization per ISO 11135. Packaging per ISO 11607. Mechanical device; no software or electronic components.

Indications for Use

Indicated for tissue collection from intrapulmonary regions via compatible bronchoscope in patients requiring diagnostic sampling.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 21, 2017 Olympus Surgical Technologies America Ms. Mary Patella Senior Specialist. Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772-2104 Re: K171232 Trade/Device Name: PeriView FLEX Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: August 21, 2017 Received: August 22, 2017 Dear Ms. Patella: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Tara A. Ryan -S 2017.09.21 05:52:34 -04'00' for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name PeriView FLEX #### Indications for Use (Describe) This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Gyrus ACMI ViziShot 2 FLEX # General Information | Manufacturer: | Olympus Surgical Technologies<br>America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508-804-2600<br>Fax: 508-804-2624 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment Registration Number: | 3003790304 | | Contact Person: | Mary Anne Patella<br>Senior Specialist, Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com | | Date Prepared: | April 25, 2017 | | Classification Name: | Bronchoscope (flexible or rigid) and<br>accessories | | CFR Citation Number: | 21 CFR 874.4680 | | Product Code: | KTI | | Classification: | Class II | | Review Panel | Ear Nose & Throat | | Trade Name: | PeriView FLEX | | Generic/Common Name: | Aspiration Needle | Spiration, Inc. PeriFLEX K162611 {4}------------------------------------------------ # Product Description The PeriView FLEX Needle is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. The device is supplied sterile and is intended for single patient use. The device consists of a handle, sheath, needle, and stylet. The sheath and needle (together termed the insertion section) are attached to the handle and needle slider respectively. The removable stylet runs the full length of the device and is located within the lumen of the needle. Note that although, the device has a component called a needle, the device is often referred to as a needle as well. The distal end of the insertion portion (sheath and needle) of the PeriView FLEX needle is inserted into the working channel of the bronchoscope and advanced to the target site. The handle is connected to the insertion and has a needle slider component. The needle slider is controlled manually by the user to extend and retract the needle from the sheath at the target site. Once a sample is collected, an air filled syringe or the stylet can be used to expel the sample from the needle. The PeriView FLEX is available in one model only (NA-403D-2021), with a needle size of 21gauge (21G). The PeriView FLEX Needle can be used with the following optional accessories: vacuum syringe, biopsy valve, and guide sheath. These optional accessories are not included with the Peri View FLEX Needle's packaging and must be obtained separately by the user. # Intended Use This device is intended to be used through a compatible bronchoscope for the collection of tissue from the intrapulmonary regions. Do not use for any purpose other than its intended use. The Intended Use is identical to that of the predicate device. ### Comparison of Technological Characteristics The PeriView FLEX has the same basic technological characteristics as the predicate PeriFLEX cleared under K162611. The subject and predicate device operate in the same manner to obtain tissue using a bronchoscope. The indications for use of the PeriView FLEX are the same as that of the predicate. The PeriView FLEX handle profile design has been changed from the predicate device to improve ergonomics; as well as to utilize Design for Manufacturing principles to make assembly more efficient. The resulting device functionality and interaction with the physician are equivalent to those of the predicate device PeriFLEX. {5}------------------------------------------------ The bronchoscope angulation specification for the PeriView FLEX differs slightly from that of the predicate to accommodate bronchoscope to bronchoscope angulation variability. The proposed device has demonstrated the ability to reach the upper lobes of the lung; thereby affirming equivalence to the predicate device. A detailed comparison of the PeriView FLEX and the PeriFLEX is provided in the following table. | | Predicate Device (K162611) | Proposed Device | |---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Characteristic | PeriFLEX | PeriView FLEX | | Indications for Use | This device is intended to be used<br>through a compatible bronchoscope for<br>the collection of tissue from the<br>intrapulmonary regions. Do not use for<br>any purpose other than its intended use. | This device is intended to be used<br>through a compatible bronchoscope<br>for the collection of tissue from the<br>intrapulmonary regions. Do not use<br>for any purpose other than its<br>intended use. | | Anatomical Site | Lung | Identical | | Use Conditions | Surgical suite, endoscopy or<br>bronchoscopy suite, used with a<br>bronchoscope | Identical | | Intended User | By or under the supervision of a<br>physician | Identical | | Mechanics of Action | Manual | Identical | | Mode of Action | Single/multiple puncture and aspirate | Identical | | General design | Handle, Sheath, Needle, Stylet | Sheath, Needle & Stylet are identical.<br>The handle design has been changed to<br>improve ergonomics and to utilize<br>Design for Manufacturing principles. | | Patient Contacting<br>Materials | Stainless Steel, PTFE, PEBAX, Nitinol | Identical | | Biocompatible | Yes | Yes | | Sterilization | Ethylene Oxide | Ethylene Oxide | | Single Use Only | Yes | Identical | | Maximum Working<br>OD (mm) | 1.5 | Identical | | Catheter Length (cm) | 115 | Identical | | Needle Gauge | 21G | Identical | | Typical Needle Length<br>(mm) | 20 | Identical | | Stylet OD (in) | 0.0185 | Identical | Table 5.1 Comparison of Proposed and Predicate Device {6}------------------------------------------------ ## Summary of Non-Clinical Testing #### Biocompatibility: Biocompatibility testing on all patient contacting surfaces has been performed in compliance to relevant requirements of ISO-10993. Biocompatibility testing included the following tests: - . ISO 10993-4: 2002 Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood - . ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - ISO 10993-10: 2010 Biological evaluation of medical devices. Tests for . irritation and sensitization - ISO 10993-11:2006. Biological evaluation of medical devices. Tests for . systemic toxicity - United States Pharmacopeia 39, National Formulary 34, 2016. <151> Pyrogen Test - United States Pharmacopeia 39, National Formulary 34, 2016. <85> Bacterial ● Endotoxins Test ### Sterilization: The PeriView FLEX will be delivered in a sterile state and is intended for single patient use only. Sterilization (ethylene oxide) and packaging of the device was validated using the following standards: - ANSI/AAMI/ISO 11607-1:2006 Packaging for terminally sterilized medical ● devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems - ANSI/AAMI/ISO 11135-1:2014 Sterilization of health-care products -. ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices Packaging integrity and performance testing on devices that had undergone accelerated aging support a labeled one year shelf life. {7}------------------------------------------------ ## Bench testing: During design verification testing, the output of the design process was evaluated against the physical and performance specifications. The following performance tests were conducted: - Sheath and Needle Insertion and Withdrawal Force - Stylet Insertion and Withdrawal Force ● - Bronchoscope Angulation ● - Activation Force - Plastic Deformation Angle - Puncture Force - . Transmission Force - Handle Assembly Strength/Handle Durability - . Device Durability - Vacuum Decay Test ### Conclusion: In summary, the PeriView FLEX is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...