← Product Code [KTI](/submissions/EN/subpart-e%E2%80%94surgical-devices/KTI) · K170759

# CRE Pulmonary Balloon Dilatation Catheter (K170759)

_Boston Scientific Corporation · KTI · Jun 23, 2017 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/KTI/K170759

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [KTI](/submissions/EN/subpart-e%E2%80%94surgical-devices/KTI.md)
- **Decision Date:** Jun 23, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

The CRE Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree.

## Device Story

The CRE Pulmonary Balloon Dilatation Catheter is a medical device used to dilate strictures within the airway tree. It consists of an inflatable balloon mounted on a catheter shaft, which includes lumens for balloon inflation and guidewire passage. The device is inserted into the airway via the working channel of a flexible bronchoscope. Operated by a physician, the device is positioned at the site of the stricture; the balloon is then inflated to mechanically dilate the narrowed airway. This procedure aims to improve airway patency and patient breathing. The current device is a modification of a predicate, featuring an increased catheter shaft length to facilitate compatibility with flexible bronchoscope working channels.

## Clinical Evidence

Bench testing only. Performance data included in-vitro testing for kink resistance, scope insertion, scope removal, inflation time, and deflation time. All devices met required specifications.

## Technological Characteristics

Balloon dilatation catheter with inflation and guidewire lumens. Modification involves increased catheter shaft length for flexible bronchoscope working channel compatibility. Mechanical operation via manual inflation.

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- CRE™ Pulmonary Balloon Dilatation Catheter ([K023337](/device/K023337.md))

## Reference Devices

- ElationTM Pulmonary Balloon Dilatation Cathete ([K161392](/device/K161392.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2017

Boston Scientific Corporation Jia Huang Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752

Re: K170759

Trade/Device Name: CRE™ Pulmonary Balloon Dilatation Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: KTI Dated: May 23, 2017 Received: May 24, 2017

Dear Jia Huang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Denise L. Hampton -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K170759

Device Name

CRE Pulmonary Balloon Dilatation Catheter

Indications for Use (Describe)

The CRE Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree.

Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### 1. Submitter

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

| Contact:   | Jia Huang                         |
|------------|-----------------------------------|
|            | Sr. Regulatory Affairs Specialist |
| Email:     | Jia.Huang@bsci.com                |
| Telephone: | (508) 683-4360                    |
| Fax:       | (508) 683-5939                    |

Date Prepared: March 9, 2017

#### 2. Device

| Trade Name:          | CRE™ Pulmonary Balloon Dilatation Catheter        |
|----------------------|---------------------------------------------------|
| Common Name:         | Balloon Dilatation Catheter                       |
| Classification Name: | Bronchoscope (flexible and rigid) and accessories |
| Regulation Number:   | 21 CFR 874.4680                                   |
| Product Code:        | KTI                                               |
| Classification:      | Class II                                          |

#### 3. Predicate Device

| Trade Name:          | CRE™ Pulmonary Balloon Dilatation Catheter        |
|----------------------|---------------------------------------------------|
| Common Name:         | Balloon Dilatation Catheter                       |
| Manufacturer:        | Boston Scientific Corporation                     |
| Clearance Number:    | K023337                                           |
| Classification Name: | Bronchoscope (flexible and rigid) and accessories |
| Regulation Number:   | 21 CFR 874.4680                                   |
| Product Code:        | KTI                                               |
| Classification:      | Class II                                          |

#### 4. Reference Device

| Trade Name:       | ElationTM Pulmonary Balloon Dilatation Cathete |
|-------------------|------------------------------------------------|
| Common Name:      | Pulmonary Balloon Dilatation Catheter          |
| Manufacturer:     | Merit Medical Systems, Inc.                    |
| Clearance Number: | K161392                                        |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Boston Scientific. The logo is in a serif font and is a dark blue color. The words "Boston" and "Scientific" are stacked on top of each other. The logo is simple and professional.

Classification Name: Bronchoscope (flexible and rigid) and accessories Regulation Number: 21 CFR 874.4680 Product Code: KTI Classification: Class II

#### 5. Device Description

The CRE™ Pulmonary Balloon Dilatation Catheter is used to access the airway tree via a bronchoscope for the purpose of dilating strictures. It consists of an inflatable balloon on a catheter shaft with lumens for inflation and passage of a guidewire.

#### 6. Indications for Use

The CRE™ Pulmonary Balloon Dilatation Catheter is intended to be used endoscopically to dilate strictures of the airway tree.

#### 7. Technologies Characteristics

The changes are modifying the design of the CRE™ Pulmonary Balloon Dilatation Catheter to be able to use the device through a flexible bronchoscope working channel by increasing the length of the catheter shaft. The proposed CRE™ Pulmonary Balloon Dilatation Catheter is identical to the predicate CRE™ Pulmonary Balloon Dilatation Catheter (K023337), with the exception of catheter shaft working length.

#### 8. Performance Data

In-Vitro Testing has been performed and all full devices met the required specifications for the completed tests. A summary of the test results has been provided for Kink Resistance, Scope Insertion, Scope Removal, Inflation Time and Deflation Time.

#### 9. Conclusion

Boston Scientific Corporation has demonstrated that the proposed CRE™ Pulmonary Balloon Dilatation Catheter is substantially equivalent to Boston Scientific Corporation's currently marketed CRE™ Pulmonary Balloon Dilatation Catheter (K023337).

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/KTI/K170759](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/KTI/K170759)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
