ViziShot 2 FLEX

K163469 · Olympus Surgical Technologies America · KTI · Apr 6, 2017 · Ear, Nose, Throat

Device Facts

Record IDK163469
Device NameViziShot 2 FLEX
ApplicantOlympus Surgical Technologies America
Product CodeKTI · Ear, Nose, Throat
Decision DateApr 6, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2

Indications for Use

The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use.

Device Story

ViziShot 2 FLEX is a manual aspiration needle device used with ultrasound endoscopes for transbronchial needle aspiration (TBNA) of tracheobronchial lesions. Device consists of handle, sheath, needle, and removable stylet. Physician inserts flexible catheter into bronchoscope working channel; locks handle to adapter biopsy valve. Needle advanced to target site under real-time ultrasound visualization; echogenic tip facilitates guidance. Physician performs puncture and applies suction via syringe at proximal handle to obtain tissue sample. After sampling, vacuum released, handle unlocked, and device withdrawn. Used in endoscopy/bronchoscopy suites. Provides tissue samples for cytopathological or microbiological examination; aids in diagnosis of lesions. Benefits include precise, ultrasound-guided tissue acquisition.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and design verification. Performance tests included sheath/needle insertion/withdrawal force, bronchoscope angulation, activation force, plastic deformation, penetration force, transmission force, device/handle durability, and echogenicity. Biocompatibility testing performed per ISO 10993-4, -5, -10, and -11.

Technological Characteristics

Manual aspiration needle; 19G needle size. Materials: Stainless steel, PTFE, PEBAX, Nitinol. Features: Adjustable/lockable needle and sheath, echogenic needle tip, aspiration capability. Sterilization: Ethylene oxide (EO) per ISO 11135-1. Packaging per ISO 11607-1. No software or electronic components.

Indications for Use

Indicated for ultrasound-guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree in patients requiring tissue sampling via ultrasound endoscope.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 6, 2017 Olympus Surgical Technologies America Mary Patella Senior Specialist, Regulatory Affairs 136 Turnpike Road Southborough, Massachusetts 01772-2104 Re: K163469 Trade/Device Name: Vizishot 2 Flex Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: March 9, 2017 Received: March 10, 2017 Dear Mary Patella: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Image /page/1/Picture/8 description: The image contains a signature and a name. The name reads "Tina Kiang-S". The signature is a series of looping lines to the left of the name. Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name ViziShot 2 FLEX #### Indications for Use (Describe) The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------------------------------------------| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary Gyrus ACMI ViziShot 2 FLEX # General Information | Manufacturer: | Olympus Surgical Technologies<br>America<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508-804-2600<br>Fax: 508-804-2624 | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment Registration Number: | 3003790304 | | Contact Person: | Mary Anne Patella<br>Senior Specialist, Regulatory Affairs<br>508-804-2771<br>Maryanne.patella@olympus-osta.com | | Date Prepared: | December 16, 2016 | Device Description | Classification Name: | Bronchoscope (flexible or rigid) and<br>accessories | |----------------------|-----------------------------------------------------| | CFR Citation Number: | 21 CFR 874.4680 | | Product Code: | KTI | | Classification: | Class II | | Review Panel | Ear Nose & Throat | | Trade Name: | ViziShot 2 FLEX | | Generic/Common Name: | Aspiration Needle | Predicate Devices | Spiration, Inc. ViziShot FLEX | K152922 | |-------------------------------|---------| |-------------------------------|---------| {4}------------------------------------------------ # Product Description The ViziShot 2 FLEX is intended for use with compatible ultrasound endoscopes for Transbronchial Needle Aspiration (TBNA) of submucosal and extramural lesions of the tracheobronchial tree. The device is supplied sterile and intended for single patient use. The device consists of a handle, sheath, needle, and stylet. The sheath and needle are attached to the handle, and the removable stylet is located within the needle. Note that although, the device has a component called a needle, the device is often referred to as a needle as well. Prior to a procedure, the flexible catheter portion is inserted into a bronchoscope's working channel (2.2mm) and advanced forward until fully inserted. The handle is then affixed to the channel port of the endoscope via a lever mechanism that locks onto the Adapter Biopsy Valve. The needle is advanced through the bronchoscope to the sampling site while visualizing both the target and the needle in real time with ultrasound. The handle facilitates advancement of the needle during puncture of the targeted biopsy site. The sample is obtained by penetrating the lesion with the needle while applying suction at the proximal end of the handle. After completing the sampling, the vacuum from the syringe is released to atmosphere, the handle unlocked from the bronchoscope, and the catheter and needle pulled out from the working channel. The removed tissue can then be prepared for cytopathological or microbiological examination and testing. The ViziShot 2 FLEX is available in one model only (NA-U403SX-4019), with a needle size of 19 gauge (19G). The two required accessories, the Adapter Biopsy Valve and the Merit Syringe with Stopcock, are packaged with the ViziShot 2 FLEX. ### Intended Use The ViziShot 2 FLEX has been designed to be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree. Do not use this device for any purpose other than its intended use. The Intended Use is identical to that of the predicate device. ### Technological Characteristics The technological characteristics of the ViziShot 2 FLEX are the same as those of the predicate ViziShot FLEX. That is they have substantially the same design, material, and energy source. They are both designed to obtain a sample from lung tissue or {5}------------------------------------------------ lymph node tissue under ultrasound visualization. Both devices have the following technological characteristics: - Needle length adjustable and lockable ● - Sheath length adjustable and lockable - Secure attachment to scope with controlled device orientation ● - Manual needle movement through bronchoscope to target site - Flexible needle - Stylet resistant to kinking - Echogenic needle tip visible with ultrasound ● - Sharp needle tip ● - Aspiration capability ● ### Comparison of Technological Characteristics The ViziShot 2 FLEX has the same basic technological characteristics as the predicate ViziShot FLEX cleared under K152922. The subject and predicate device operate in the same manner to obtain a tissue biopsy using an ultrasound endoscope. The indications for use of the ViziShot 2 FLEX are the same as that of the predicate. The ViziShot 2 FLEX handle profile design has been changed from the predicate device to improve ergonomics; as well as to utilize Design for Manufacturing principles to make the manufacturing assembly more efficient. The resulting device functionality, look, and interaction with the physician are equivalent to those of the predicate device ViziShot FLEX. A detailed comparison of the ViziShot 2 FLEX and the ViziShot FLEX is provided in the following table. | | Predicate Device (K152922) | Modified Device | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX | | Indications for Use | The ViziShot FLEX has been<br>designed to be used with<br>ultrasound endoscopes for<br>ultrasound guided fine needle<br>aspiration (FNA) of<br>submucosal and extramural<br>lesions of the<br>tracheobronchial tree. Do not<br>use this device for any<br>purpose other than its<br>intended use. | The ViziShot 2 FLEX has been<br>designed to be used with<br>ultrasound endoscopes for<br>ultrasound guided fine needle<br>aspiration (FNA) of<br>submucosal and extramural<br>lesions of the tracheobronchial<br>tree. Do not use this device for<br>any purpose other than its<br>intended use. | | Use Conditions | Surgical suite, endoscopy or<br>bronchoscopy suite, used with<br>a bronchoscope | Identical | | | Predicate Device (K152922) | Modified Device | | Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX | | Mechanics of Action | Manual | Identical | | Mode of Action | Single/multiple puncture and<br>aspirate | Identical | | General design | Handle, Sheath, Needle,<br>Stylet | Similar<br>The handle profile design has<br>been changed to improve<br>ergonomics and utilize Design<br>for Manufacturing principles.<br>Changes include:<br>• Handle shape and added<br>rubber finger grip<br>• Sheath adjuster rotates<br>• Snap bonds instead of<br>screw and glue | | Patient Contacting<br>Materials | Stainless Steel, PTFE,<br>PEBAX, Nitinol | Stainless Steel, PTFE, PEBAX,<br>Nitinol | | Biocompatible | Yes | Yes | | Product Specification<br>and Properties | Engineering bench testing confirms that the modified ViziShot 2<br>FLEX meets its critical product specifications, which are<br>substantially equivalent to those of the predicate ViziShot<br>FLEX. | | | Sterilization | EO | EO | | Single Use Only | Yes | Identical | | Working OD (mm) | 2.08 | Identical | | Catheter Length (cm) | 70 | Identical | | Needle Gauge | 19G | Identical | | Typical Needle Length<br>(mm) | 20 | Identical | | Max Needle Length<br>(mm) | 40 | Identical | | Stylet OD (in) | 0.0205 | Equivalent<br>0.0204 | | Stylet Surface Finish | Polished | Identical | | Accessories | Syringe with stopcock<br>Adapter biopsy valve<br>Devices provided with device | Identical | | Packaging | Needle assembly in tray with | Needle assembly, syringe, | | | Predicate Device (K152922) | Modified Device | | Device Name/<br>Characteristics | ViziShot FLEX | ViziShot 2 FLEX | | | snap downs. Tray placed in<br>pouch. Pouch placed in shelf<br>box (dust cover) for sterilization.<br>After sterilization, the sterile<br>pouched syringe and sterile<br>pouched adapter will be added<br>to the dust cover. | adapter placed in tray with<br>snap downs and tyvek lid.<br>Tray placed in shelf box<br>prior to sterilization. | {6}------------------------------------------------ . {7}------------------------------------------------ # Summary of Non-Clinical Testing # Biocompatibility: Biocompatibility testing on all patient contacting surfaces has been performed in compliance to relevant requirements of ISO-10993. Biocompatibility testing included the following tests: - ISO 10993-4: 2002 Biological evaluation of medical devices Part 4: ● Selection of tests for interactions with blood - ISO 10993-5: 2009 Biological evaluation of medical devices - Part5: Tests for in vitro cytotoxicity - ISO 10993-10: 2010 Biological evaluation of medical devices. Tests for ● irritation and sensitization - . ISO 10993-11:2006. Biological evaluation of medical devices. Tests for systemic toxicity ### Sterilization: The ViziShot 2 FLEX will be delivered in a sterile state and is intended for single patient use only. Sterilization (ethylene oxide) and packaging of the device was validated using the following standards: - . ANSI/AAMI/ISO 11607-1:2006 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems - . ANSI/AAMI/ISO 11135-1:2014 Sterilization of health-care products ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices Packaging integrity and performance testing on devices that had undergone accelerated aging support a labeled one year shelf life. {8}------------------------------------------------ # Bench testing: During design verification, the output of the design process was evaluated against the physical and performance specifications. The following performance tests were conducted: - Sheath and Needle Insertion and Withdrawal Force - Stylet Insertion and Withdrawal Force ● - Bronchoscope Angulation ● - Activation Force - Plastic Deformation Angle - Penetration Force - Transmission Force - Device Durability/Handle Durability - Bronchoscope Adapter Sliding Force - Handle Durability - Sheath to Handle Joint Strength ● - . Echogenicity ### Conclusion: In summary, the ViziShot 2 FLEX is substantially equivalent to the predicate device and presents no new questions of safety or efficacy.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...