← Product Code [ERL](/submissions/EN/subpart-e%E2%80%94surgical-devices/ERL) · K131053

# ANSPACH XMAX, EMAX2 AND EMAX2 PLUS SYSTEM WITH OTOLOGIC ATTACHMENT SYSTEM (K131053)

_The Anspach Effort, Inc. · ERL · Dec 31, 2013 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/ERL/K131053

## Device Facts

- **Applicant:** The Anspach Effort, Inc.
- **Product Code:** [ERL](/submissions/EN/subpart-e%E2%80%94surgical-devices/ERL.md)
- **Decision Date:** Dec 31, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4250
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are intended for cutting and shaping bone including spine and cranium. When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies.

## Device Story

System comprises pneumatic or electric surgical drill handpieces (XMax, eMax 2, eMax2 Plus) paired with Otologic Curved Micro (OCM) attachment and OCM Burr Support Sleeves. OCM attachment connects to handpieces via twist-lock; burr support sleeves snap into attachment. Sleeves feature 20° curve and small diameter for improved visibility; non-rotating sleeve protects soft tissue from rotating burr. Operated by surgeons in clinical/OR settings for bone dissection. System provides mechanical power to cutting burrs at speeds of 10,000–80,000 rpm. Benefits include precise bone shaping and enhanced access in delicate otologic procedures.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including biocompatibility testing per GLP (21 CFR Part 58), functional performance testing (repeated use, temperature safety, cleaning/sterilization validation), and a simulated use study confirming performance in bone cutting and shaping.

## Technological Characteristics

Pneumatic and electric surgical drill systems. OCM attachment uses standard twist-lock connection; OCM Burr Support Sleeves use snap-in connection. Materials are biocompatible, consistent with currently marketed attachments. Operating speed: 10,000–80,000 rpm. OCM sleeves available in 5.6 cm and 7.3 cm lengths. Non-rotating sleeve design for soft tissue protection. Reusable attachment; single-use sterile burr support sleeves.

## Regulatory Identification

An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.

## Predicate Devices

- ANSPACH® XMax Pneumatic System ([K965080](/device/K965080.md))
- ANSPACH® eMax 2 and eMax2 Plus Electric Systems ([K011444](/device/K011444.md))
- Medtronic, Inc., XPS® 3000 System ([K073255](/device/K073255.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEC 3 1 2013

510(k) PREMARKET NOTIFICATION ANSPACH® Systems with Otologic Curved Micro (OCM) Attachment and OCM Burr Support Sleeves

## 510(k) Summary

| Submitter<br>Information        | Company:               | The Anspach Effort, Inc.<br>4500 Riverside Drive<br>Palm Beach Gardens, FL 33410                                                                                  |
|---------------------------------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | Contact Person:        | Jeannette G. Dailey<br>Regulatory Affairs Manager                                                                                                                 |
|                                 | Email:                 | Dailey.Jeannette@synthes.com                                                                                                                                      |
|                                 | Telephone:             | 561-494-3710                                                                                                                                                      |
|                                 | FAX:                   | 561-625-9110                                                                                                                                                      |
|                                 | Date Prepared:         | November 8, 2013                                                                                                                                                  |
| Identification of<br>the Device | Common Names:          | Pneumatic and Electric Surgical Instruments<br>and Attachments                                                                                                    |
|                                 | Device Name:           | ANSPACH® XMax Pneumatic and eMax 2 and<br>eMax2 Plus Electric Systems with the<br>ANSPACH® Otologic Curved Micro (OCM)<br>Attachment and OCM Burr Support Sleeves |
|                                 | Classification Names:  | • Ear, nose and throat electric or pneumatic<br>surgical drill<br>• Ear, nose and throat bur                                                                      |
|                                 | Device Classification: | • Class II, 21 CFR 874.5250 (drill)<br>• Class 1, 510(k) exempt, 21 CFR 874.4140 (bur)                                                                            |

Predicate Devices(s) The subject devices are substantially equivalent to the following predicate devices:

The electric and pneumatic surgical drills are identical to the currently marketed electric and pneumatic surgical drills:

- ANSPACH® XMax Pneumatic System cleared via K965080 .
- ANSPACH® eMax 2 and eMax2 Plus Electric Systems cleared via . K011444

The Otologic Curved Micro (OCM) Attachment and OCM Burr Support Sleeves are substantially equivalent to the handpiece and burrs associated with the:

- . Medtronic. Inc., XPS® 3000 System cleared most recently via K073255 on March 24, 2008
Continued on next page

{1}------------------------------------------------

## 510(k) Summary, Continued

| Description of Device | The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are surgical drills that handle a complete range of surgical procedures ranging from power demanding applications to the most delicate dissection.                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                       | The ANSPACH® Otologic Curved Micro (OCM) Attachment consists of the attachment which connects directly to the ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems handpieces with the standard twist lock mechanism, and the OCM Burr Support Sleeves. The OCM is provided non-sterile and is reusable. Cleaning and sterilization instructions are provided in the ANSPACH® OCM Attachment package insert.                                                                                                                                                                                                                                  |
|                       | The ANSPACH® OCM Burr Support Sleeves with integral cutting burrs mount to the OCM Attachment with a simple snap-in connection. The 20° curve and small diameter of the OCM Burr Support Sleeves provides excellent visibility. The non-rotating support sleeve protects soft tissue from the rotating cutting burr shaft. The OCM Burr Support Sleeves can be used at speeds from 10,000 to 80,000 rpm. They are available in two lengths (5.6 and 7.3 cm), each with a wide variety of different cutting burr styles and diameters to accommodate varying surgeon preferences. The OCM Burr Support Sleeves are provided sterile and for single use only. |
| Indications for Use   | The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electric Systems are intended for cutting and shaping bone including spine and cranium.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                       | When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

Continued on next page

.

.

{2}------------------------------------------------

## 510(k) Summary, Continued

| Technological<br>Characteristics | There are no new technological characteristics associated with the<br>ANSPACH® Pneumatic and Electric Systems, the OCM Attachment<br>and the OCM Burr Support Sleeves. The ANSPACH® XMax<br>Pneumatic and eMax 2 and eMax2 Plus Electric Surgical Drills are<br>identical to the currently marketed electric and pneumatic surgical drills.<br>The new ANSPACH® Otologic Curved Micro (OCM) Attachment and<br>OCM Burr Support Sleeves utilize the same biocompatible materials as<br>currently marketed attachments and burrs. The packaging,<br>manufacturing method, and cleaning and sterilization methods of the<br>OCM Attachment are the same as currently marketed ANSPACH®                                                                                                                                                                                            |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | attachments. The OCM Burr Support Sleeves packaging,<br>manufacturing method and sterilization method are also the same as<br>currently marketed ANSPACH® burrs.<br>The new indications for use are consistent with numerous other<br>currently marketed surgical drill systems and accessories.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Non-Clinical<br>Testing          | A series of biocompatibility tests demonstrated the safety of the OCM<br>Burr Support Sleeves patient contacting materials. All tests were<br>conducted in accordance with the GLP regulation (21 CFR Part 58).<br>Several performance tests demonstrated the safety and effectiveness of<br>the OCM Attachment and OCM Burr Support Sleeves. Specifically,<br>OCM Attachment testing demonstrated proper functionality after<br>repeated use, safe external temperatures, and the ability to clean and<br>sterilize effectively without adverse effects on functionality or<br>performance. OCM Burr Support Sleeves testing demonstrated proper<br>functionality and safe external temperatures. In addition, a simulated<br>use study demonstrated both the OCM Attachment and the OCM Burr<br>Support Sleeves perform as intended in cutting and shaping bone<br>properly. |
| Conclusions                      | Scientifically valid data demonstrates that the ANSPACH® XMax<br>Pneumatic and eMax 2 and eMax2 Plus Electric Systems with the<br>ANSPACH® Otologic Curved Micro (OCM) Attachment and OCM Burr<br>Support Sleeves is as safe and effective as the predicate devices when<br>used in accordance with the labeled directions for use and for the<br>requested indication.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

December 31, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

The ANSPACH Effort, Inc. % Ms. Annette M. Hillring 3012 St. Charles Drive Tampa, FL 33618

Re: K131053

Trade/Device Name: ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Electrical Systems with Otologic Attachment System (Otologic Attachment and Curved Burrs) Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose and Throat Electric or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL; EQJ Dated: November 26, 2013 Received: December 2, 2013

Dear Ms. Hillring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Ms. Annette M. Hillring

You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Bradley S. Cunningham -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear,

Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): _K131053_

Device Name: ANSPACH® XMax Pneumatic and eMax2 and eMax2 Plus Electrical Systems with Otologic Attachment System

The ANSPACH® XMax Pneumatic and eMax 2 and eMax2 Plus Indications for Use: Electric Systems are intended for cutting and shaping bone including spine and cranium.

When used with the ANSPACH® Systems, the OCM Attachment and OCM Burr Support Sleeves are intended for cutting and shaping bone primarily in otology procedures such as cochleostomies.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sageev George -S 2013.12.27 17:13:53 -05'00'

Page of of

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/ERL/K131053](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/ERL/K131053)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
