← Product Code [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ) · K223782

# Ambu aScope 5 Broncho HD 5.6/2.8 Sampler Set, Ambu aScope 5 Broncho HD 5.0/2.2 Sampler Set (K223782)

_Ambu A/S · EOQ · Jan 12, 2023 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K223782

## Device Facts

- **Applicant:** Ambu A/S
- **Product Code:** [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ.md)
- **Decision Date:** Jan 12, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

Ambu® aScope™ 5 Broncho HD Sampler Set is intended for endoscopic procedures and examination within the airways and tracheobronchial tree. Ambu® aScope™ 5 Broncho HD Sampler Set is intended to provide visualization via a compatible Ambu displaying unit. and to allow passing of endotherapy instruments via its working channel. Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration and collection of fluid samples.

## Device Story

Single-use, flexible bronchoscope system; includes aScope 5 Broncho HD endoscope, BronchoSampler 60 SC container, Bronchoscope Attachment Part (BAP), Suction Connection Tube (SCT), and luer lock adapters. Device inputs: visual signals from distal tip camera/LED; fluid samples via suction pathway. Operation: endoscope tip maneuvered by clinician; BronchoSampler inserted into suction pathway between endoscope and vacuum source; BAP maintains suction pathway connections. Output: real-time visualization on compatible Ambu display unit; collected fluid samples in container. Used in clinical settings for airway examination and sampling. Clinician uses visual output to guide endotherapy instruments and perform diagnostic aspiration. Benefits: sterile, single-use design eliminates reprocessing needs; integrated sampling capability allows multiple samples during one procedure.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification included transportation, packaging, suction, connectivity, functional testing of the sampler, biological evaluation, and sterilization validation.

## Technological Characteristics

Flexible, single-use endoscope with distal camera and LED light source. Materials sterilized via Ethylene Oxide. Includes suction pathway accessories (sampler container, BAP, SCT). Connectivity: wired to compatible Ambu display unit. Design verification performed per ISO 14971.

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- Ambu® aScope™ 5 Broncho HD ([K220606](/device/K220606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 12, 2023

Ambu A/S % Sanjay Parikh Director, QA/RA Ambu Inc. 6721 Columbia Gateway Drive, suite 200 Columbia, Maryland 21046

Re: K223782

Trade/Device Name: Ambu aScope 5 Broncho HD 5.6/2.8 Sampler Set, Ambu aScope 5 Broncho HD 5.0/2.2 Sampler Set Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: December 12, 2022 Received: December 16, 2022

Dear Sanjay Parikh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K223782

Device Name

Ambu® aScope™ 5 Broncho HD 5.6/2.8 Sampler Set Ambu® aScope™ 5 Broncho HD 5.0/2.2 Sampler Set

## Indications for Use (Describe)

Ambu® aScope™ 5 Broncho HD Sampler Set is intended for endoscopic procedures and examination within the airways and tracheobronchial tree.

Ambu® aScope™ 5 Broncho HD Sampler Set is intended to provide visualization via a compatible Ambu displaying unit. and to allow passing of endotherapy instruments via its working channel.

Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration and collection of fluid samples.

| Type of Use (Select one or both, as applicable)                                                                          |                                                          |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span> <span style="text-decoration: underline;"></span>Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This Special 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summary has been prepared in accordance with 21 CFR 807.92.

| Submitter                                                                            | Ambu A/S<br>Baltorpbakken 13<br>2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050                                                                                               |                          |                                             |                             |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|---------------------------------------------|-----------------------------|
| Contact Person                                                                       | Name: Kristian Moltved<br>Job Title: Associate Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, 2750 Ballerup<br>Telephone number: +45 7225 2116<br>Fax number: +45 7225 2050 |                          |                                             |                             |
| Date Summary<br>Prepared                                                             | December 12, 2022                                                                                                                                                                                    |                          |                                             |                             |
| Device Trade<br>Name                                                                 | Ambu® aScope™ 5 Broncho HD 5.6/2.8 Sampler Set<br>Ambu® aScope™ 5 Broncho HD 5.0/2.2 Sampler Set                                                                                                     |                          |                                             |                             |
| Device Common<br>Name                                                                | Flexible Endoscope and Specimen Sampling System - Single Use                                                                                                                                         |                          |                                             |                             |
| Device<br>Classification                                                             | Ambu® aScope™ 5 Broncho HD Sampler Set:<br>Bronchoscope (flexible or rigid) and accessories<br>Product Codes: EOQ<br>21 CFR 874.4680<br>Class II                                                     |                          |                                             |                             |
| Legally Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent | Predicate                                                                                                                                                                                            | Manufacturer<br>Ambu A/S | Trade Name<br>Ambu® aScope™<br>5 Broncho HD | 510(k)<br>number<br>K220606 |

{4}------------------------------------------------

| Description of<br>the Device | The Ambu® aScope™ 5 Broncho HD Sampler Set consists of:<br>Ambu® aScope™ 5 Broncho HD Two Sample Containers (aScope BronchoSampler™ 60 SC) Suction Connection Tube (SCT) Bronchoscope Attachment Part (BAP) 2 Luer lock adapters (introducers)                                                                   |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              | Ambu® aScope™ 5 Broncho HD is currently being sold on its own as a sterile, single use, flexible endoscope intended for endoscopic procedures and examination within the airways and tracheobronchial tree.                                                                                                      |
|                              | The Ambu® BronchoSampler™ 60 SC is a sterile, single-use medical device designed as an accessory to aScope 5 Broncho HD single- use endoscopes. It enables aspiration and collection of fluid samples.                                                                                                           |
|                              | The Ambu® BronchoSampler™ 60 SC will be sold alone as an add-on to aScope 5 Broncho HD and as a part in the Ambu® aScope™ 5 Broncho HD Sampler Set.                                                                                                                                                              |
|                              | The Ambu® BronchoSampler™ 60 SC achieves its function by being inserted into the suction pathway between the endoscope and the vacuum source.                                                                                                                                                                    |
|                              | The Sample Container Ambu® BronchoSampler™ 60 SC can be removed and replaced by the user to support multiple samples being taken during the same procedure.                                                                                                                                                      |
|                              | The Bronchoscope Attachment Part (BAP) is inserted into the suction pathway between the Ambu® aScope™ 5 Broncho HD endoscope and the vacuum source. It functions to maintain the connections and the controls for the suction pathway and the Ambu® BronchoSampler™ 60 SC, where the aspirated sample is stored. |

{5}------------------------------------------------

The Suction Connection Tube (SCT) facilitates connection to suction tubing having end connectors of the male type, instead of the more typical female type.

The Ambu® BronchoSampler™ 60 SC plus an additional Sampler Container, BAP, SCT and two introducers will be packaged and sterilized together with a aScope 5 Broncho HD endoscope to form a self-contained set, the aScope 5 Broncho HD Sampler Set.

There will be two sets, corresponding to the two sizes of aScope 5 Broncho HD:

- a. Ambu® aScope 5 Broncho HD 5.6/2.8 Sampler Set
- b. Ambu® aScope 5 Broncho HD 5.0/2.2 Sampler Set

Ambu® aScope™ 5 Broncho HD Sampler Set has the following physical and performance characteristics:

- Maneuverable Ambu® aScope 5 Broncho HD endoscope tip . controlled by the user
- Flexible Ambu® aScope 5 Broncho HD endoscope insertion . cord
- . Camera and LED light source at the distal tip of the Ambu® aScope 5 Broncho HD endoscope
- . Sterilized by Ethylene Oxide
- For single use
- Enables aspiration and sample collection ●

The following characteristic of the endoscope varies between sizes:

- Distal end outer diameter
- Insertion tube outer diameter
- Working channel inner diameter ●
- . Angulation range

{6}------------------------------------------------

| Indications for<br>Use                                                                            | Ambu® aScope™ 5 Broncho HD Sampler Set is intended for<br>endoscopic procedures and examination within the airways and<br>tracheobronchial tree.<br><br>Ambu® aScope™ 5 Broncho HD Sampler Set is intended to<br>provide visualization via a compatible Ambu displaying unit and to<br>allow passing of endotherapy instruments via its working channel.<br><br>Ambu® aScope™ 5 Broncho HD Sampler Set enables aspiration<br>and collection of fluid samples.                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>device | The Ambu® aScope™ 5 Broncho HD endoscope in the Ambu®<br>aScope™ 5 HD Broncho Sampler Set is identical to the predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                   | Ambu® aScope™ 5 Broncho HD Sampler Set differs from the<br>predicate in the following areas:<br>Ambu® aScope™ BronchoSampler 60 SC, BAP, SCT and<br>two introducers are added to the Ambu® aScope™ 5<br>Broncho HD endoscope. Ambu® aScope™ BronchoSampler 60 SC, BAP, SCT and<br>two introducers are packaged and sterilized together with<br>aScope 5 Broncho HD endoscopes to form a self-contained<br>set, the Ambu® aScope™ 5 Broncho HD Sampler Set.                                                                                                                                                                                                                                                                                                                                                            |
| Performance<br>Data -Bench                                                                        | Performance tests related to the modifications was performed to<br>document the following properties of the modification to Ambu®<br>aScope™ 5 Broncho HD.<br><br>In addition to the non-clinical performance testing documented for<br>the Ambu® aScope™ 5 Broncho HD (K220606), additional bench<br>testing has been performed in accordance with ISO 14971 as<br>Design Verification of those Design Input requirements that are<br>specified as risk control measures for those risks arising from the<br>addition of sampling functionality to the endoscope. A table of this<br>special 510(k) cross references these additional risks, documented<br>in the Product Risk Evaluation for the Ambu® aScope™ 5 Broncho<br>Sampler Set, with the risk control measures and the subsequent<br>verification results. |
|                                                                                                   | A summary of the methods to verify the additional requirements<br>and identified risks arising from the risk control measures are as<br>follows:<br>• Transportation testing<br>• Packaging testing<br>• Suction Testing<br>• Connectivity testing<br>• Tests for basic function of BronchoSampler 60<br>• Biological evaluation<br>• Sterilization Validation<br>Result: All requirements are verified successfully.                                                                                                                                                                                                                                                                                                                                                                                                 |
| Performance<br>Data - Clinical                                                                    | Not applicable.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Conclusion                                                                                        | The Ambu® aScope™ 5 Broncho Sampler Set has the same<br>intended use and indications for use, and similar technological<br>characteristics and principles of operation as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                   | It is concluded that Ambu® aScope™ 5 Broncho Sampler Set is<br>substantial equivalent to its predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{7}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K223782](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K223782)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
