← Product Code [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ) · K211947

# GlideScope BFlex 2.8 Single-Use Bronchoscope (K211947)

_Verathon Medical (Canada) Ulc · EOQ · Nov 3, 2021 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K211947

## Device Facts

- **Applicant:** Verathon Medical (Canada) Ulc
- **Product Code:** [EOQ](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ.md)
- **Decision Date:** Nov 3, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4680
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Pediatric

## Indications for Use

GlideScope® BFlex™ Single-Use Bronchoscopes are intended to work with a video monitor, in conjunction with nonpowered endoscopic accessories and other equipment for endoscopy within the airways and tracheobronchial tree.

## Device Story

Single-use flexible bronchoscope; provides real-time visualization of airways/tracheobronchial tree. Input: optical image data from distal tip camera; LED illumination. Transformation: transmits video signal via cable to reusable GlideScope video monitor (GVM or Core). Output: real-time video display for clinician. Used in hospital environments by clinicians. No working channel; lacks therapeutic/suction capabilities. Benefits: enables airway visualization in pediatric/adult patients without cross-contamination risk due to single-use design.

## Clinical Evidence

No clinical studies performed. Substantial equivalence supported by bench testing (electrical safety per AAMI/ANSI ES60601-1, EMC per IEC 60601-1-2, optical performance per ISO 8600 series, biocompatibility per ISO 10993-1) and a usability study involving simulated use scenarios with pediatric and adult populations.

## Technological Characteristics

Flexible endoscope; 2.8mm outer diameter; 0° direction of view; 120° diagonal field of view; 5-50mm depth of field; 640x480 resolution. Materials: medical-grade thermoplastic elastomers, cyanoacrylate adhesive, polyolefin primer. Power: rechargeable lithium-ion battery (via monitor). Connectivity: wired to reusable video monitor. Sterilization: Ethylene Oxide (EO).

## Regulatory Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- GlideScope® BFlex™ 3.8 Single-Use Bronchoscope System ([K193488](/device/K193488.md))
- GlideScope® BFlex™ 5.0 Single-Use Bronchoscope System ([K183256](/device/K183256.md))
- GlideScope® BFlex™ 5.8 Single-Use Bronchoscope System ([K191948](/device/K191948.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 3, 2021

Verathon Medical (Canada) ULC Teresa Davidson Director, Regulatory Affairs 2227 Douglas Road Burnaby, British Columbia V5C 5A9 Canada

#### Re: K211947

Trade/Device Name: GlideScope BFlex 2.8 Single-Use Bronchoscope Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: September 30, 2021 Received: October 4, 2021

#### Dear Teresa Davidson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

#### K211947

Device Name GlideScope® BFlex™ 2.8 Single-Use Bronchoscope

Indications for Use (Describe)

GlideScope® BFlex™ Single-Use Bronchoscopes are intended to work with a video monitor, in conjunction with nonpowered endoscopic accessories and other equipment for endoscopy within the airways and tracheobronchial tree.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Verathon logo. The logo consists of a stylized "V" shape in green and blue, followed by the word "VERATHON" in blue. The "V" shape is formed by two intersecting lines, with the left line in green and the right line in blue. The word "VERATHON" is in a sans-serif font, with the letters evenly spaced. There is a trademark symbol after the word "VERATHON".

# SECTION E: 510(K) SUMMARY

This summary of Safety and Effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR 807, Subpart E, section 807.92.

#### Submitter:

Verathon Medical (Canada) ULC 2227 Douglas Road Burnaby, BC V3W 1P2 Canada

#### Contact Person:

Teresa Davidson Director, Regulatory Affairs Phone: (425) 629-5516 Email: Teresa.davidson@verathon.com

### Date Summary Prepared:

October 21, 2021

### Establishment Registration Number:

Verathon Medical (Canada) ULC Registration Number: 9615393 Owner/Operator Number: 9095489

### Device Trade or Proprietary Name:

GlideScope® BFIex™ 2.8 Single-Use Bronchoscope

#### Device Common or Usual Name:

Flexible Bronchoscope and accessories

| Device Trade or Proprietary Name                  | Device Common or Usual Name           |
|---------------------------------------------------|---------------------------------------|
| GlideScope® BFlex™ 2.8 Single-Use<br>Bronchoscope | Flexible Bronchoscope and accessories |

### Device Classification:

| Classification Name                                 | Class | Product Code | Classification Regulation |
|-----------------------------------------------------|-------|--------------|---------------------------|
| Bronchoscope (Flexible or<br>Rigid) and Accessories | II    | EOQ          | 21 CFR 874.4680           |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the Verathon logo. The logo consists of a stylized symbol to the left of the word "VERATHON". The symbol is a green and blue abstract shape, and the word "VERATHON" is in blue. There is a registered trademark symbol to the right of the word "VERATHON".

### Review Panel:

Ear. Nose, and Throat

### Predicate Device:

The features and functions of the proposed GlideScope® BFlex™ 2.8 Single-Use Bronchoscope is substantially equivalent to the previously cleared GlideScope® BFlex™ 3.8. 5.0. and 5.8 Single-Use Bronchoscopes. The 510(k) clearance numbers and respective clearance dates for the predicate devices are included in the table below:

| Predicate Device                                         | 510(k) Number | Clearance Date   |
|----------------------------------------------------------|---------------|------------------|
| GlideScope® BFlex™ 3.8 Single-Use<br>Bronchoscope System | K193488       | January 16, 2020 |
| GlideScope® BFlex™ 5.0 Single-Use<br>Bronchoscope System | K183256       | January 04, 2019 |
| GlideScope® BFlex™ 5.8 Single-Use<br>Bronchoscope System | K191948       | August 21, 2019  |

## Device Description:

The GlideScope® BFlex™ 2.8 Single-Use Bronchoscope is one component of the GlideScope® BFlex™ Single-Use Bronchoscope System. The system consists of a single-use flexible bronchoscope, a reusable monitor, and a reusable cable. The GlideScope® BFlex™ Single-Use Bronchoscope System is intended to provide real time viewing and recording for a wide range of airway procedures.

Similar to the predicate GlideScope® BFlex™ 3.8, 5.0, and 5.8 Single-Use Bronchoscopes, the GlideScope® BFlex™ 2.8 Single-Use Bronchoscope is distributed sterile and is for single use only. The GlideScope® BFlex™ bronchoscopes operate with a portable reusable GlideScope video monitor (GVM or Core monitors) for purposes of image display.

### Indications for Use:

GlideScope® BFlex™ Single-Use bronchoscopes are intended to work with a video monitor, in coniunction with non-powered endoscopic accessories and other ancillary equipment, for endoscopy within the airways and tracheobronchial tree.

### Intended Patient Population:

The GlideScope® BFlex™ Single-Use system is for use in a hospital environment. The GlideScope BFlex bronchoscope is a single-use device designed for use in adults, with the BFlex 2.8 also designed for pediatric use (6 months to 6 years). It has been verified

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Verathon logo. The logo consists of a stylized green and blue symbol to the left of the word "VERATHON" in blue. The symbol appears to be a stylized letter "V" with a green portion and a blue portion. The word "VERATHON" is in a sans-serif font.

and validated for the following endotracheal tube (ETT) and endoscope accessory (EA) sizes:

| Model     | Minimum ETT Internal Diameter | EA Minimum Working Channel Width |
|-----------|-------------------------------|----------------------------------|
| BFlex 2.8 | 4.0 mm                        | Not applicable                   |
| BFlex 3.8 | 5.0 mm                        | 1.2 mm                           |
| BFlex 5.0 | 6.0 mm                        | 2.1 mm                           |
| BFlex 5.8 | 7.0 mm                        | 3.0 mm                           |

#### Contraindications:

The GlideScope® BFlex™ 2.8 Single-Use Bronchoscope does not have a working channel and therefore cannot be used for therapeutic purposes.

### Technological Characteristics:

The proposed subject GlideScope® BFlex™ 2.8 Single-Use Bronchoscope when compared to the predicate bronchoscopes has similar technological characteristics. See the comparison table below for similarities and differences between the proposed and predicate devices:

| Technological<br>Characteristic                                           | Predicate<br>GlideScope®<br>BFlex™ 5.0<br>Single-Use<br>Bronchoscope | Predicate<br>GlideScope®<br>BFlex™ 5.8<br>Single-Use<br>Bronchoscope | Predicate<br>GlideScope®<br>BFlex™ 3.8<br>Single-Use<br>Bronchoscope | Proposed<br>GlideScope®<br>BFlex™ 2.8<br>Single-Use<br>Bronchoscope |
|---------------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|---------------------------------------------------------------------|
| Flexible Endoscope                                                        | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |
| Size Distinguishing<br>Color<br>(Non-patient contacting)                  | Blue                                                                 | Yellow                                                               | Purple                                                               | Teal                                                                |
| Working Channel<br>Material<br>(Patient contacting)                       | Thermoplastic<br>Polyurethane<br>(Medical grade)                     | Thermoplastic<br>Polyurethane<br>(Medical grade)                     | Fluoropolymer<br>(Medical grade)                                     | Not applicable                                                      |
| Tip sheath adhesive<br>and primer material<br>(Patient contacting)        | Cyanoacrylate<br>adhesive and<br>polyolefin<br>primer                | Same                                                                 | Same                                                                 | Same                                                                |
| Tip Sheath Material<br>(Patient contacting)                               | Medical<br>thermoplastic<br>elastomers                               | Same                                                                 | Same                                                                 | Different Grade                                                     |
| Technological<br>Characteristic                                           | Predicate<br>GlideScope®<br>BFlex™ 5.0<br>Single-Use<br>Bronchoscope | Predicate<br>GlideScope®<br>BFlex™ 5.8<br>Single-Use<br>Bronchoscope | Predicate<br>GlideScope®<br>BFlex™ 3.8<br>Single-Use<br>Bronchoscope | Proposed<br>GlideScope®<br>BFlex™ 2.8<br>Single-Use<br>Bronchoscope |
| Braided Shaft Material<br>(Patient contacting)                            | Medical<br>thermoplastic<br>elastomers                               | Same                                                                 | Same                                                                 | Different Grade                                                     |
| Outside Diameter of<br>Flexible Insertion<br>Tube/Shaft and Distal<br>Tip | 5.0mm                                                                | 5.8mm                                                                | 3.8mm                                                                | 2.8mm                                                               |
| Minimum Internal<br>Diameter of Working<br>Channel                        | 2.1mm                                                                | 3.0mm                                                                | 1.2mm                                                                | Not Applicable                                                      |
| Suction Port                                                              | Yes                                                                  | Yes                                                                  | Yes                                                                  | Not Applicable                                                      |
| Accessory Port                                                            | Yes                                                                  | Yes                                                                  | Yes                                                                  | Not Applicable                                                      |
| Single Use<br>Bronchoscope                                                | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |
| Sterility                                                                 | Sterile by<br>Ethylene Oxide<br>(EO)                                 | Same                                                                 | Same                                                                 | Same                                                                |
| Control Button for Tip<br>maneuverability                                 | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |
| Power Source                                                              | Rechargeable<br>Lithium-ion<br>Battery                               | Same                                                                 | Same                                                                 | Same                                                                |
| Camera                                                                    | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |
| Direction of View,<br>Relative to Center Line<br>of Distal tip            | 0°                                                                   | Same                                                                 | Same                                                                 | Same                                                                |
| Field of view,<br>horizontal/vertical                                     | 85°                                                                  | Same                                                                 | Same                                                                 | Same                                                                |
| Field of View, diagonal                                                   | 120°                                                                 | Same                                                                 | Same                                                                 | Same                                                                |
| Depth of Field                                                            | 5-50mm                                                               | Same                                                                 | Same                                                                 | Same                                                                |
| Image Resolution                                                          | 640x480                                                              | Same                                                                 | Same                                                                 | Same                                                                |
| LED Light Source                                                          | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |
| Image Display                                                             | Displays image<br>on a Reusable<br>Video Monitor                     | Same                                                                 | Same                                                                 | Same                                                                |
| Extended Viewing                                                          | Yes                                                                  | Yes                                                                  | Yes                                                                  | Yes                                                                 |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Verathon. The logo consists of a stylized green "V" shape on the left, followed by the word "VERATHON" in blue, with a trademark symbol after the word. The logo is simple and modern, with a clean font and a bright color scheme.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Verathon logo. The logo consists of a stylized green and blue symbol to the left of the word "VERATHON" in blue, bold letters. The symbol appears to be a stylized letter "V" or a stylized plant.

# Performance Testing:

Performance testing has been completed to demonstrate that the proposed GlideScope® BFlex™ 2.8 Single-Use Bronchoscope meets the safety and performance requirements established in the design specifications. Comprehensive verification and validation testing included the following:

- Full System Requirements Testing ●
- Electrical Safety according to ●
	- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
	- IEC 60601-2-18: Edition 3.0 2009-08 o Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- Electromagnetic Compatibility according to ●
	- IEC 60601-1-2:2014 Ed.4.0/EN 60601-1-2:2015 о Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- Optical testing according to .
	- o ISO 8600-1:2015 (Ed. 4.0) Endoscopes - Medical endoscopes and endotherapy devices -- Part 1: General requirements
		- ISO 8600-3:2019 (Ed. 2.0) o Optics and Optical instruments - Medical endoscopes and endoscopic accessories - Part 3: Determination of field of view and direction of view of endoscopes with optics
		- ISO 8600-4:2014 (Ed. 2.0) o Endoscopes - Medical endoscopes and certain accessories - Part 4: Determination of maximum width of insertion portion
- Biocompatibility according to ●
	- o ANSI AAMI ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- Aging Performance Testing ●
- Sterile Packaging Integrity Testing .
- Cleaning Testing .
- Design Validation .
	- o Usability study
		- The usability study was conducted per:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the Verathon logo. The logo features a stylized "V" shape in blue and green, with the word "VERATHON" in blue, sans-serif font to the right of the "V" shape. The "V" shape is composed of two intersecting lines, with the left line in green and the right line in blue.

- IEC 60601-1-6. General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62366-1, 2015, Medical devices Application of usability engineering to medical devices, which calls for usability and validation testing during medical device development, and
- AAMI HE75: 2009, Human factors engineering Design of medical devices
- Guidance for Industry and Food and Drug Administration Staff, - Applying Human Factors and Usability Engineering to Medical Devices (February 03, 2016)

The study was focused on user requirements and risk control measures related to the introduction of the smaller size GlideScope® BFlex™ 2.8 Single-Use Bronchoscope. Both pediatric (ages 6 months to 6 years old) and adult patient populations were considered. A comprehension test in addition to the simulated use scenarios was completed. All participants, overall, were able to operate the GlideScope® BFlex™ 2.8 Single-Use Bronchoscope safely and effectively. The result of this usability testing demonstrated that the GlideScope® BFlex™ 2.8 Single-Use Bronchoscope and associated labeling are safe and effective for the intended use, its intended users, and use environments.

All testing resulted in acceptance criteria passed. Results:

### Summary of Clinical Tests:

The GlideScope® BFlex™ 2.8 Single-Use Bronchoscope, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.

#### Conclusion:

The information in this 510(k) Premarket Notification demonstrates that the proposed GlideScope® BFlex™ 2.8 Single-Use Bronchoscope is substantially equivalent to the previously cleared predicate GlideScope® BFIex™ 3.8, 5.0, and 5.8 Single-Use Bronchoscopes with respect to safety, effectiveness, and performance.

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K211947](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOQ/K211947)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
