FUJIFILM Bronchoscope Model EB-530P

K171291 · Fujifilm Medical Systems U.S.A, Inc. · EOQ · Jan 19, 2018 · Ear, Nose, Throat

Device Facts

Record IDK171291
Device NameFUJIFILM Bronchoscope Model EB-530P
ApplicantFujifilm Medical Systems U.S.A, Inc.
Product CodeEOQ · Ear, Nose, Throat
Decision DateJan 19, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.

Device Story

Flexible video bronchoscope; used for visualization, diagnosis, and treatment of trachea and bronchial tree. Input: optical images captured via distal CCD sensor; illumination provided by internal glass fiber bundles. Device transforms optical input into electronic signals for display on external monitor via video processor. Operated by clinicians in clinical settings. Features control section for distal tip angulation; channels for air/suction and endoscopic accessories (e.g., forceps). Output: real-time endoscopic video feed. Assists clinicians in identifying pathology and performing interventions. Benefits: enables minimally invasive access to airway for diagnostic and therapeutic procedures.

Clinical Evidence

Bench testing only. Evaluated electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18), biocompatibility (ISO 10993-1, -5, -10), and endoscope-specific performance (ISO 8600-1, -3, -4). Validated reprocessing instructions (cleaning, high-level disinfection, EO sterilization). Performance metrics included field of view, bending capability, suction rate, working length, forceps channel diameter, viewing direction, and light guide output. All tests met specifications.

Technological Characteristics

Flexible video bronchoscope; CCD image sensor; glass fiber bundles for illumination. Distal end/flexible portion diameter 3.8mm; working length 600mm. Connectivity: compatible with FUJIFILM video processors (EPX-4440HD). Sterilization: EO sterilization. Biocompatibility per ISO 10993; electrical safety per IEC 60601 series. Mechanical performance per ISO 8600 series.

Indications for Use

Indicated for observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree in patients requiring bronchoscopic procedures.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 19, 2018 FUJIFILM Medical Systems U.S.A., Inc. Jeffrey Wan Specialist, Regulatory Affairs 10 High Point Drive Wayne, NJ 07470 Re: K171291 Trade/Device Name: FUJIFILM Bronchoscope Model EB-530P Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: December 18, 2017 Received: December 19, 2017 Dear Jeffrey Wan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {1}------------------------------------------------ (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Eric A. Mann -S for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171291 Device Name FUJIFILM Bronchoscope Model EB-530P Indications for Use (Describe) The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY # FUJIFILM Medical Systems U.S.A., Inc.'s FUJIFILM Bronchoscope Model EB-530P #### Date: January 15, 2018 #### Submitter's Information FUJIFILM Medical Systems U.S.A., Inc., Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293 #### Contact Person: Jeffrey Wan Specialist, Regulatory Affairs Telephone: (973) 709-2219 Facsimile: (201) 995-2452 E-Mail: jeffrey.wan@fujifilm.com ## Identification of the Proposed Device: | Proprietary/Trade Name: | FUJIFILM Bronchoscope Model EB-530F | |-------------------------|-------------------------------------| | Common Name: | Bronchoscope | | Device Class: | Class II | | Review Panel: | Ear, Nose, and Throat | Classification Information: | Classification Name | CFR Section | Product Codes | |--------------------------------------------------|-----------------|---------------| | Bronchoscope (flexible or rigid) and accessories | 21 CFR 874.4680 | EOQ | #### Predicate Devices - Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535) ● #### Intended Use / Indications for Use The FUJIFILM Bronchoscope Model EB-530P is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree. {4}------------------------------------------------ # Device Description The FUJIFILM Bronchoscope Model EB-530P is comprised of three general sections: a control/ operating section, an insertion and an umbilicus. The control/operating section controls the angulation (up/down) of the distal end of the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary charged coupled device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The subject device is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K122535. # Technological Characteristics A comparison of the technological characteristics between the subject and predicate devices is provided in the Table 7-1 below. ## Performance Data Electrical safety of the subject device was evaluated using following standards: ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009. Biocompatibility of the subject devices was evaluated using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010. Cleaning, high-level disinfection, and EO sterilization were performed to validate the reprocessing instructions recommended in the labeling. Endoscope specific testing was conducted using the following consensus standards: ISO 8600-1:2015, ISO 8600-3:1997, and ISO 8600-4:2014. Subject devices met performance specifications in the following additional testing: - . Field of view - . Bending capability - Rate of suction - Working length - Dianeter of forceps channel . - Viewing direction - . LG Output {5}------------------------------------------------ K171291 # Table 7.1: Comparison of FUJIFILM Bronchoscope Model EB-530P with its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscope Model EB-530S (K122535) | | | Predicate Device<br>Fujinon/Fujifilm EB-530 Series<br>Bronchoscopes Model EB-530S (K122535) | Proposed Device<br>FUJIFILM Bronchoscope Model EB-530P | |----------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------| | Device Name | | FUJIFILM Video Bronchoscope | FUJIFILM Video Bronchoscope | | | | EB-530S | EB-530P | | 510(k) Number | | K122535 | K171291 | | Indications for Use (IFU) | | The device is intended for the observation,<br>diagnosis, and endoscopic treatment of<br>trachea and bronchial tree. | Same as predicate device | | Viewing direction | | Forward/ 0 degree | Same as predicate device | | Observation range | | 3mm - 100mm | Same as predicate device | | Field of View | | 120 degrees | Same as predicate device | | Distortion characteristics | | Orthogonal Projection | Same as predicate device | | Image sensors | | CCD | Same as predicate device | | Distal end diameter | | 4.9mm | 3.8mm | | Flexible portion diameter | | 4.9mm | 3.8mm | | Maximum insertion diameter | | 5.9mm | 4.2mm | | Bending capability | Up | 180 degrees | Same as predicate device | | | Down | 130 degrees | Same as predicate device | | Forceps channel diameter | | 2.0mm | 1.2mm | | Working length | | 600mm | Same as predicate device | | Total length | | 870mm | 890mm | | Video Processor | | EPX-2500<br>EPX-4400<br>EPX-4400HD<br>EPX-4440HD | EPX-4440HD | | | Predicate Device<br>Fujinon/Fujifilm EB-530 Series<br>Bronchoscopes Model EB-530S (K122535) | Proposed Device<br>FUJIFILM Bronchoscope Model EB-530P | | | Peripherals | Endoscopic Accessory (i.e. Forceps) | Endoscopic accessory i.e. forceps caliber<br>under φ1.2mm | | | | Monitor | Same as predicate device | | | | Printer | Same as predicate device | | | | Electrosurgical Instruments | N/A | | | Peripherals | Foot Switch | Same as predicate device | | | | Cart | Same as predicate device | | | Standard Accessories | Channel Cleaning brush<br>WB3212FW2<br>WB3503FW | Channel Cleaning Brush<br>WB2212FW2<br>WB3503FW | | | | Valve Cleaning brush<br>(WB11002FW2) | Same as predicate device | | | | Cleaning Adapter Kit (CA-500C) | Same as predicate device | | | | Forceps Valve<br>(FOV-DV7) | Same as predicate device | | | | Suction button<br>(SB-500B/D) | Same as predicate device | | | | Ventilation Adapter<br>(AD-7) | Same as predicate device | | | Optional Accessories | Air leak tester LT-7F | Same as predicate device | | | | Sterile Suction button<br>(SB-602) | Same as predicate device | | {6}------------------------------------------------ {7}------------------------------------------------ # Substantial Equivalence FUJIFILM Bronchoscope Model EB-530P has the same intended use and indications for use, as well as similar technological characteristics and principles of operation as its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535). The modifications done to the subject device include changes in number of light quide fiber bundles, dimensions, accessories, transport and storage conditions, and materials. These minor dimensional and material differences between the FUJIFILM Bronchoscope Model EB-530P and its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S were made for the purpose of overall product enhancement and general technological advancement, and raise no new issues of safety or effectiveness. The material changes in the subject device do not raise new concerns regarding biocompatibility as discussed in section XVII. Performance data demonstrated that the FUJIFILM Bronchoscope Model EB-530P has substantially equivalent performance to the predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535). The subject device is successfully validated for the electrical safety and reprocessing instructions, supporting their safety equivalent to the predicate device. # Conclusions The subject device, Bronchoscope Model EB-530P is substantially equivalent to its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535), based on the same intended use/indications for use and similar technological characteristics. The differences in the dimensions and materials between the subject device and its predicate device raise no new issues of safety or effectiveness. Bench testing data demonstrated that the subject device has substantially equivalent performance to the predicate. Thus, the subject device FUJIFILM Bronchoscope Model EB-530P is substantially equivalent to its predicate device Fujinon/Fujifilm EB-530 Series Bronchoscopes Model EB-530S (K122535).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...