The LungPoint Tools are endoscopic tools intended to be used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
Device Story
LungPoint Tools consist of a sheath and a dilation balloon used during bronchoscopy. The sheath provides a working channel for introducing endoscopic tools to targeted lung tissue; the balloon dilates bronchial tree tissue. The bronchoscope is navigated to a predefined target using LungPoint Software (K12051). The sheath includes a stylet to improve pushability and prevent airway mucosa entry. The device is used in a clinical setting by a physician. The output is physical access to the target site or mechanical dilation of tissue, which assists the physician in reaching targets or performing interventions. Benefits include improved access to lung tissue and ability to dilate bronchial structures.
Clinical Evidence
Bench testing only. Testing included dimensional analysis, joint/tensile strength, simulated use, balloon fatigue/burst pressure, deflation time, and radiopacity. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity). All results met pre-determined acceptance criteria.
Technological Characteristics
Materials include medical-grade plastics and platinum iridium markers. Sheath includes a stylet. Dilation balloon rated for 20 atm. Sterilization via e-beam. Connectivity: None (mechanical accessory).
Indications for Use
Indicated for use with LungPoint Software guided bronchoscopes to provide a working channel for endoscopic tools or to dilate bronchial tree tissue in adult patients. Not for pediatric use.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
K133385 — LUNGPOINT ATV PLANNING AND NAVIGATION SOFTWARE · Broncus Medical, Inc. · Feb 26, 2014
K090660 — AIRWAY BALLOON CATHETER INFLATION DEVICE · Acclarent, Inc. · Jun 12, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# 131234
## 510(k) Summary
1. Date of Summary
October 10, 2013
## 2. 510(k) Applicant
Broncus Medical, Inc. 1400 N. Shoreline Blvd, Suite A8 Mountain View, California 94043 (650) 428-1600 Phone: FAX: (650) 428-1542
OCT 15 2013
| Contact Person: | Gary Kaplan |
|-----------------|---------------------|
| Phone: | (650) 428-1600 |
| Fax: | (650) 428-1542 |
| e-mail: | gkaplan@broncus.com |
## 3. Device Overview
| Trade Name: | LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon) |
|----------------------|---------------------------------------------------------------------|
| Common Name: | Sheath and Dilation Balloon |
| Classification Name: | Bronchoscope and Accessories<br>21 CFR 874.4680<br>Product Code EOQ |
# 4. Predicate Device
The predicate devices identified are as follows:
| Trade Name | 510(k) Submitter | 510(k) Number |
|----------------------------------------------|----------------------------------------|----------------------------------------------------|
| CRE Pulmonary Balloon<br>Dilatation Catheter | Boston Scientific<br>Corporation | K023337, cleared to market on<br>November 18, 2002 |
| Olympus Guide Sheath | Olympus Medical<br>Systems Corporation | K060243, cleared to market on<br>June 23, 2006 |
## 5. Device Description
The LungPoint Tools are endoscopic tools used during bronchoscopy procedures. The LungPoint Sheath is designed to be used with a bronchoscope to provide a working
{1}------------------------------------------------
channel through which endoscopic tools, such as needles, dilation balloons, or other endoscopic devices may be introduced to the targeted lung tissue within the respiratory organs. The LungPoint Dilation Balloon is used to dilate tissue of the bronchial tree and may be inserted through the sheath or directly through the working channel of the bronchoscope. The bronchoscope is advanced to a predefined target following guidance of the LungPoint Software (previously cleared under 510(k) #K 12051 to guide endoscopic tools or catheters in the lungs or to enable marker placement in soft lung tissue).
The materials used in the LungPoint Tools are commonly used medical grade materials and include platinum iridium markers.
| Catalog<br>Number | Working Length | Maximum<br>Catheter Outer<br>Diameter (OD) | Catheter Internal<br>Diameter (ID) | Minimum accessory<br>length for use through<br>sheath |
|-------------------|----------------|--------------------------------------------|------------------------------------|-------------------------------------------------------|
| 10007-1 | 900mm | 2.65mm | 2.0mm | 965mm |
The LungPoint Sheath has the following specifications:
The LungPoint Dilation Balloon has the following specifications:
| Catalog<br>Number | Balloon Size<br>(OD x length) | Rated Balloon<br>Pressure | Maximum<br>Catheter Outer<br>Diameter (OD) | Catheter<br>Length/working length |
|-------------------|-------------------------------|---------------------------|--------------------------------------------|-----------------------------------|
| 10008-1 | 4mm x 6mm | 20atm | 1mm | 1430mm/975mm |
# 6. Intended Use
The LungPoint Tools are endoscopic tools intended to be used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
# 7. Comparison to Predicate Device
The LungPoint Tools are commonly used endoscopic tools with the same technological characteristics as the predicate devices. The indications for use of the LungPoint Tools for use with a bronchoscope guided by the LungPoint Software are all within the intended use of the predicate devices, which is to aid in reaching a target in the respiratory organ
{2}------------------------------------------------
either directly as is the case with the sheath or through dilating target tissue with the dilation balloon. The technological characteristics of the subject devices are the same as those of the predicate devices with the following exceptions
- LungPoint Balloon: balloon size and length, balloon burst pressure and balloon . catheter length as outlined.
| | CRE Pulmonary Balloon<br>Dilatation Catheter<br>(K023337) | LungPoint Dilation<br>Balloon |
|--------------------------------------------------|-----------------------------------------------------------|-------------------------------|
| Balloon Size and Inflation<br>(balloon pressure) | 8mm @ 3 ATM<br>9mm @ 5.5 ATM<br>10mm @ 9 ATM | 4mm @ 10 ATM |
| Catheter Length (cm) | 155 | 143 |
| Balloon Length (cm) | 3.0 | 0.6 |
| Rated Burst Pressure (atmospheres) | 9 | 20 |
- LungPoint Sheath: stylet, catheter ID/OD and catheter length as outlined. .
| | Olympus Guide Sheath<br>(K060243) | LungPoint Sheath |
|-----------------|-----------------------------------|-----------------------------------------------------------------------------------------|
| Stylet Provided | No | Yes - used to enhance<br>pushability and to prevent<br>airway mucosa entering<br>sheath |
| Catheter ID/OD | 2.1/2.7 mm | 2.0/2.6 mm |
| Catheter Length | 900 mm | 975 mm |
None of these differences raise new questions of safety and effectiveness. Performance of the subject devices has been verified by use of accepted methods.
### 8. Performance Data
The design and safety of the LungPoint Tools were verified by performing functional and performance testing. All tests were designed to subject the sheath and dilation balloon to stresses that exceed those which would be encountered during clinical use. Testing included the following:
- Dimensional testing -
- Joint/tensile test
- Simulated use -
- Balloon fatigue/burst pressure -
- Balloon deflation time -
- Radiopacity. -
All testing results met the pre-determined acceptance criteria that were established in the test protocols. Based on the testing the LungPoint Tools are as safe, as effective, and perform at least as safely and effectively as the predicate devices.
{3}------------------------------------------------
#### 9. Safety and Effectiveness
The LungPoint Tools labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the devices. The biocompatibility assessment of all patient contacting materials was performed in accordance with ISO 10993. Biological Evaluation of Medical Devices. Specifically, cytotoxicity, sensitization, intracutaneous reactivity and systemic toxicity (acute) were tested. In addition, the devices are sterilized using e-beam sterilization.
#### 10. Conclusion
Based on the testing the LungPoint Tools are as safe, as effective, and perform at least as safely and effectively as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS) in the United States. The logo consists of a stylized graphic element resembling an abstract human form or a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
October 15, 2013
Broncus Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K131234
Trade/Device Name: LungPoint™ Tools (LungPoint Sheath and LungPoint Dilation Balloon) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOQ Dated: August 26, 2013 Received: August 27, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
""Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A.Männ -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K131234
Device Name:
LungPoint Tools - LungPoint Dilation Balloon and LungPoint Sheath
The LungPoint Tools are endoscopic tools intended to be Indications for Use: used with LungPoint Software guided bronchoscopes. The LungPoint Sheath is intended to be used as a working channel through which endoscopic tools may be introduced to targeted tissue. The LungPoint Dilation Balloon is intended to dilate tissue of the bronchial tree and may be inserted through the LungPoint Sheath or directly through the working channel of the bronchoscope. Not for pediatric use.
× Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page __ of ___________________________________________________________________________________________________________________________________________________________________
Srinivaš
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.