← Product Code [EOB](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB) · K103467

# CMOS VIDEO RHINO-LARYNGOSCOPE SYSTEM MODEL 11101CM (K103467)

_KARL STORZ Endoscopy-America, Inc. · EOB · Jun 28, 2012 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K103467

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [EOB](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB.md)
- **Decision Date:** Jun 28, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4760
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

## Device Story

System comprises flexible endoscope with integrated LED light source and distal CMOS imaging sensor; C-MAC monitor for display. Used for endoscopic diagnosis of nasal lumens and airway anatomy. Operated by qualified clinicians in clinical settings. CMOS sensor captures video signal; transmitted to monitor for real-time visualization. Enables direct inspection of airway/nasal structures to assist clinical diagnosis. Benefits include improved visualization compared to traditional fiber-optic systems.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993, and ISO 8600. Reprocessing validated per AAMI TIR 12:2010, AAMI TIR 30:2003, AAMI 7:1999, and AAMI E 1837.

## Technological Characteristics

Flexible endoscope with distal CMOS sensor; LED illumination; fiber optic light transmission. Compatible with C-MAC monitor. Reprocessing via manual cleaning and high-level disinfection (CIDEX OPA). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ISO 10993, ISO 8600.

## Regulatory Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- KARL STORZ VIDEO RHINO-LARYNGSCOPE SYSTEM. MODEL11101SERIES ([K072387](/device/K072387.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103467

JUN 2 8 2012

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Karl Storz Endoscopy-America, Inc. Applicant: 2151 E. Grand Avenue El Segundo. CA 90245

Contact: Shiven Gandhi Regulatory Affairs Associate Telephone: (424) 218-8322 FAX: (424) 218-8519 E-mail:shiven.gandhi@karlstorz.com

Date Prepared: 14th June 2012

## Device Identification:

Common Name: Video Rhino-Laryngoscope System

Trade Name: CMOS Video Rhino-Laryngoscope System

Predicate Device: KARL STORZ VIDEO RHINO-LARYNGSCOPE SYSTEM. MODEL11101SERIES (K072387)

Indication: The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

### Device Description:

The Karl Storz CMOS Video Rhino-Laryngoscope System includes a flexible endoscope and a monitor. The rhino-laryngoscope consists of an integrated light located in the handle. The light is transmitted thought the fiber optic bundles and illuminates the anatomy under visualization. A new Storz specific CMOS imaging sensor located at the distal tip of the endoscope shaft which is inserted into the body cavities. The captured video signal is then transferred to the C-MAC monitor for display purposes. The Karl Storz CMOS Video Rhino-laryngoscope is a Class II device under 21 CFR874.4760, Rhino-Laryngoscope and accessories.

### Substantial Equivalence:

The Karl Storz CMOS Video Rhino-Laryngoscope System is substantially equivalent to the predicate device (K072387) since the intended uses and the technological characteristics are the same. The minor differences between the Karl Storz CMOS Video Rhino-Laryngoscope System and predicate device raise no issue of safety and effectiveness, as these differences have no effect on the performance, function or general intended use of these devices.

{1}------------------------------------------------

|                     | Subject Device                                                             | Predicate Device (K072387)                                                 |
|---------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Applicant           | Karl Storz Endoscopy America,<br>Inc.                                      | Karl Storz Endoscopy America,<br>Inc.                                      |
| Device Name         | Karl Storz CMOS Video Rhino-<br>Laryngoscope                               | Karl Storz Video Rhino-<br>Laryngoscope                                    |
| Type of Device      | Videoscope System                                                          | Videoscope System                                                          |
| Indication for use  | For endoscopic diagnosis<br>within the nasal lumens and<br>airway anatomy. | For endoscopic diagnosis<br>within the nasal lumens and<br>airway anatomy. |
| Imaging Sensor      | CMOS                                                                       | CCD                                                                        |
| Illumination source | LED                                                                        | Xenon                                                                      |
| Working Length      | Same                                                                       | Same                                                                       |
| Shaft Diameter      | Same                                                                       | Same                                                                       |
| Data Storage        | Yes                                                                        | Yes                                                                        |

## Technological Characteristics Comparison Table

## Non-Clinical Performance Data:

The Karl Storz Flexible Video Rhino-Laryngoscope system has been successfully tested for its functions, performance and safety as per FDA recognized standards. Testing has been conducted as per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and ISO 10993 to address the safety aspect of the device. The performance of the device is tested by complying with ISO 8600.

## Reprocessing:

The CMOS Video Rhino-Laryngoscope is cleaned using manual cleaning with enzymatic detergent and is subjected to high level disinfection using CIDEX OPA. No methods of sterilization are validated for the CMOS Video Rhino-Laryngoscope. The cleaning and highlevel disinfection studies were perform in accordance with AAMI TIR 12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers), AAMI TIR 30:2003 (A compendium of processes, materials. test methods, and acceptance criteria for cleaning reusable medical devices), AAMI 7: 1999 (Chemical Sterilants and High-Level Sterilants: A Guide to Selection and Use), AAMI E 1837 (Standard Test Method to Determine Efficacy of Disinfection Processes for Reusable Medical Devices) and FDA draft guidance for Processing Medical devices in Healthcare settings (dated: May 2,2011)

## Conclusion :

Hence the CMOS Video Rhino-Laryngoscope system is as safe and as effective to the predicate device in terms of its intended use and technological characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a circular seal for the Department of Health & Human Services USA. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged around the emblem in a circular fashion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 2 8 2012

Karl Storz Endoscopy America Inc. c/o Mr. Shiven Gandhi, Regulatory Affairs Associate 2151 E. Grand Avenue El Segundo, CA 90245

Re: K103467

Trade/Device Name: Karl Storz CMOS Video Rhino Laryngoscope System Regulation Number: 21 CFR 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB Dated: June 14, 2012 Received: June 15, 2012

Dear Mr. Gandhi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, attresses, and include requirements for annual registration, listing of general volurely provincitice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adultieration. Teasoneter ODTE assess that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to additional controls. Existing major regulations affecting your device can be may be subject to additional vehicles, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease of advised that I Dr over and that your device complies with other requirements of the Act that I DA has made a associations administered by other Federal agencies. You must or any I carated and the regirements, including, but not limited to: registration and listing (21 comply with an the rece brequirements on 301); medical device reporting of medical CI'N I all 807), labeling (21 OFF 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Gandhi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

R. A. Khan, M.D., f

Malvina B. Eydelman, M Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indication for Use

# 510(k) Number (if known):

Device Name: Karl Storz CMOS Video Rhino-Laryngoscope System

Indication for use: The KARL STORZ CMOS Flexible Video Rhino-Laryngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR ·

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel C Cluff

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K103467

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K103467](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K103467)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
