VISION-SCIENCES ENT-5000 VIDEO ENT SCOPE WITH ENDOSHEATH TECHNOLOGY, VISION-SCIENCES DPU-5000/DPU-5050 VIDEO PROCESSOR
K102733 · Vision-Sciences, Inc. · EOB · May 18, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K102733
Device Name
VISION-SCIENCES ENT-5000 VIDEO ENT SCOPE WITH ENDOSHEATH TECHNOLOGY, VISION-SCIENCES DPU-5000/DPU-5050 VIDEO PROCESSOR
Applicant
Vision-Sciences, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
May 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Intended Use
The flexible ENT-5000 Video ENT Scope with EndoSheath® Technology is intended for use in flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages; and for use in diagnostic arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening. The digital video processor is intended for use with the VSI flexible video scope.
Device Story
Flexible ENT-5000 video endoscope system; includes digital video processor, display monitor, and optional sterile, single-use EndoSheath® technology. EndoSheath covers insertion tube to provide barrier protection. System captures and displays real-time video images of internal body cavities. Used by clinicians in clinical settings for diagnostic endoscopic and arthroscopic procedures. Provides illumination and visualization of upper airway, vocal cords, nasal passages, and other body cavities. Output viewed on monitor to assist physician in diagnostic assessment. Benefits include enhanced visualization and reduced cross-contamination risk via disposable sheath.
Clinical Evidence
Bench testing only. No clinical data provided. Device passed electrical safety, thermal, and EMC testing requirements. Patient contact materials verified as identical to predicate device (K072073).
Technological Characteristics
Flexible video endoscope with digital video processor and disposable sheath. Materials identical to K072073. Illumination and visualization via fiber optics/digital sensor. Connectivity to external display monitor. System includes optional sterile, single-use protective sheath (with or without working channel).
Indications for Use
Indicated for flexible endoscopic examination of upper airway, vocal cords, and nasal passages; and diagnostic arthroscopic and endoscopic procedures requiring illumination and visualization of body cavities via natural or surgical openings. No specific age or gender restrictions stated.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Modified Flexible ENT Scopes with Digital Video Processor and Disposable EndoSheath® Technology (K072073)
Avis Surgical Technologies, Inc.: C-MOR Visualization System (K093717)
Biomet InnerVue™ Diagnostic Scope System (K072879)
SpineView SpineVu Endoscopic Spine System (SESS™) and SpineVu MiniScope (K081051)
Related Devices
K072073 — VISION-SCIENCES ENT-5000 AND ENT-5100 VIDEO ENT SCOPE WITH ENDOSHEATH SYSTEM, DPU-5000/DPU-5050 VIDEO PROCESSOR · Vision-Sciences, Inc. · Aug 29, 2007
K192090 — CMOS Video Rhino-Laryngoscope SSU · Karl Storz SE & CO. KG · Jan 27, 2020
K091768 — VIDEO SCIENCES BRS-5000 VIDEO BRONCHOSCOPE WITH ENDOSHEATH SYSTEM, DPU-5000/DPU-5050 VIDEO PROCESSOR & ACCESSORIES · Vision-Sciences, Inc. · Aug 3, 2009
K232435 — Rhinolaryngoscope system · Shenzhen HugeMed Medical Technical Development Co., Ltd. · Apr 26, 2024
K253184 — Video Rhinolaryngoscope System (Disposable Video Rhinolaryngoscope SP11A Endoscopic Video Processor M110B) · Zhuhai Wesee Meditech Co., Ltd. · Mar 25, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
Vision-Sciences, Inc. September 16, 2010
Kbiา
ditional Indication 510(k) Premarket Notification: Flexible ENT Scopes with Digital Video Processor and Disposable EndoSheath® Technology
# 510(k) Summary
# Date Prepared (21 CFR 807.92(a)(1): September 8, 2010
1 8 2011 MAY
Owner's Name (21 CFR 807.92(a)(1):
| | Vision-Sciences, Inc. |
|-------------------|----------------------------------|
| Address: | 40 Ramland Road South |
| | Orangeburg, NY 10962 |
| Telephone Number: | (845) 365-0600 |
| Fax Number: | (845) 365-0620 |
| Contact Person: | Lillian Quintero; Director QA/RA |
# Trade Name, Common Name, Classification (21 CFR 807.92(a)(2))
| Subject Device Name: | Flexible ENT-5000 Scopes with Digital Video Processor and<br>Disposable EndoSheath® Technology |
|------------------------|------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Flexible Video Endoscope with Sheath and Video Processor |
| Product Codes: | EOB, HRX |
| FDA Regulations: | 21CFR874.4760, 21CFR888.1100 |
| Device Classification: | Class II |
# Predicate Device Names (21 CFR 807.92(a)(3))
Modified Flexible ENT Scopes with Digital Video Processor and Disposable EndoSheath® Technology (K072073)
Avis Surgical Technologies, Inc .: C-MOR Visualization System (K093717)
Biomet InnerVue™ Diagnostic Scope System (K072879)
Smith & Nephew Arthroscope (K043395)
SpineView SpineVu Endoscopic Spine System (SESS™) and SpineVu MiniScope (K081051)
| Common/Usual Name: | Flexible video endoscopes with sheaths and accessories |
|-------------------------|--------------------------------------------------------|
| Product Codes: | EOB, HRX |
| FDA Regulations: | 21CFR874.4760, 21CFR888.1100 |
| Device Classification: | Class II |
| Premarket Notification: | K072073/K093717/K072879/K043395/K081051 |
### Device Description
The VSI endoscopes are flexible endoscopes with connections to a video processor and display monitor. The EndoSheath® Techology for the ENT-5000 are optional, sterile, single-use protective sheath systems, with or without a working channel, that are intended to cover the entire insertion tube of the videoscope. The digital video processors are used with the flexible videoscopes for image visualization and capture.
#### Intended Use
The flexible ENT Videoscope with EndoSheath® Technology is intended for use in flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages; and for use in diagnostic arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.
26
{1}------------------------------------------------
Vision-Sciences, Inc. September 16, 2010
510(k) Premarket Notification: Flexible ENT Scopes with Digital Video Processor and Disposable EndoSheath® Technology
# Technological Characteristics
Vision Sciences believes that the subject device is substantially equivalent to both the Vision Sciences' predicate device, as well as those of other manufacturers, as outlined in this submission. The subject device has very similar functionality and indications as the predicate devices.
# Performance Testing
The subject device has been subjected to and passed electrical safety, thermal, and EMC testing requirements. The patient contact materials in the endoscope are identical to the materials used in predicate device Vision Sciences' ENT-5000 (K072073).
### Conclusion
Based on this request for additional indications for use, and the comparison to predicate devices, the VSI flexible video endoscopes with digital video processor and disposable EndoSheath® Technology have been shown to be safe and effective for their intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The eagle is positioned to the right of a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Vision-Sciences, Inc. % Ms. Lillian Quintero 40 Ramland Road, South Orangeburg, New York 10962
MAY 18 2011
Re: K102733
Trade/Device Name: Flexible ENT-5000 Video ENT Scope with Video Processor and EndoSheath® Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ, HRX, EOB Dated: April 27, 2011 Received: April 29, 2011
Dear Ms. Quintero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Lillian Quintero
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
A.g B. nk
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Vision-Sciences, Inc. September 16, 2010
Additional Indications - 510(k) Premarket Notification: Flexible ENT Scopes with Digital Video Processor and Disposable EndoSheath® Technology
510(k) Number (if known):
Flexible ENT-5000 Video ENT Scope with Video Processor and EndoSheath® Device Name: Technology
Indications for Use: The flexible ENT-5000 Video ENT Scope with EndoSheath® Technology is intended for use in flexible endoscopic examination of the upper airway, vocal cords and/or nasal passages; and for use in diagnostic arthroscopic and endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening.
The digital video processor is intended for use with the VSI flexible video scope.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over-the -Counter Use _ (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Net RP Ogden for mkm
(Division Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Page 1 of 1
21
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.