← Product Code [EOB](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB) · K030857

# MAHE PEDIATRIC NASO PHARYNGOSCOPE ITS ASSOCIATED AND ANCILLARY EQUIPMENT AND ENDOSCOPIC ACCESSORIES (K030857)

_Mahe Intl., Inc. · EOB · Oct 16, 2003 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K030857

## Device Facts

- **Applicant:** Mahe Intl., Inc.
- **Product Code:** [EOB](/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB.md)
- **Decision Date:** Oct 16, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.4760
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MAHE Pediatric Naso-Pharyngoscope has been designed to be used with a light source, documentation equipment and video monitor for sinuscopy diagnosis and treatment within the nasal cavity and nasal pharynx.

## Device Story

Flexible nasopharyngoscope for pediatric sinuscopy; used with external light source, documentation equipment, and video monitor. Provides visual access to nasal cavity and nasopharynx for diagnostic and therapeutic procedures. Operated by qualified surgeons trained in flexible sinuscopy. Device features 2.7mm shaft, 2.8mm distal end, and 95-degree field of view. Enables real-time visualization for physician-guided interventions; facilitates diagnosis and treatment of nasal/pharyngeal conditions.

## Clinical Evidence

Bench testing only. Device designed and tested according to voluntary safety standards IEC 601-1 and IEC 601-2-18.

## Technological Characteristics

Flexible nasopharyngoscope; 2.7mm shaft diameter; 2.8mm distal end; 95-degree field of view; 3-50mm depth of view; 300-420mm working length. Complies with IEC 601-1 and IEC 601-2-18 safety standards. Manual operation; requires external light source and video monitor.

## Regulatory Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- KARL STORZ NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) [K945381](/device/K945381.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

030857

## OCT 1 6 2003

# 510 (k) Summary of Safety and Effectiveness

| DEVICE NAME:        | MAHE Pediatric Naso Pharyngoscope its associated and ancillary equipment and endoscopic accessories |
|---------------------|-----------------------------------------------------------------------------------------------------|
| COMMON/USUAL NAME   | Pediatric Naso-Pharyngoscope and Accessories                                                        |
| CLASSIFICATION NAME | Pediatric Naso-Pharyngoscope and Accessories                                                        |
| PREDICATE DEVICES   | KARL STORZ NASOPHARYNGOSCOPE (FLEXIBLE OR RIGID) K945381                                            |
| SUBMITTED BY:       | Paul Cart<br>MAHE International Inc<br>490 Craighead St<br>Nashville, TN 37204<br>615-269-7256      |
| ESTABLISHMENT #     | 9007726                                                                                             |

The MAHE Pediatric Naso-Pharyngoscope is intended for use only by qualified surgeons. Do not use the instrument for any purpose other than its intended application.

The MAHE Pediatric Naso-Pharyngoscope has been designed to be used with a light source, documentation equipment and video monitor for sinuscopy diagnosis and treatment within the nasal cavity and nasal pharynx.

The MAHE Pediatric Naso-Pharyngoscope is indicated for visual exam during diagnostic and therapeutic procedure within the nasal cavity and nasal pharynx.

The MAHE Pediatric Naso-Pharyngoscope are designed, manufactured, and tested according to Voluntary Safety Standards IEC 601-1 and IEC 601-2-18.

When compared with the predicate devices, the MAHE Pediatric Naso-Pharyngoscope does not incorporate any significant change in intended use, method of operation, material, or design that could affect the safety or effectiveness.

Only physicians who have received thorough previous training in the art of flexible sinuscopy should use this instrument.

{1}------------------------------------------------

# Predicate Device Comparison

| Feature                 | New Device           | Predicate     |
|-------------------------|----------------------|---------------|
| Optical System          |                      |               |
| Field of view           | 95 degree            | 80+/-4 degree |
| Depth of view           | 3-50mm               | 3-50mm        |
| Working length          | 300-420mm            | 320mm         |
| Sheath                  |                      |               |
| Shaft                   | 2.7mm                | 2.7 mm        |
| Distal end              | 2.8mm                | 2.8 mm        |
| Material                | Identical            | Identical     |
| Bending Section         |                      |               |
| Up                      | 120                  | 160+3,-10     |
| Down                    | 100                  | 130 +5,-10    |
| Sterilization Method(s) |                      |               |
| Weight                  | not more than 0.8 kg | .5-.7 kg      |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 6 2003

MAHE International, Inc. c/o Paul Cart 490 Craighead Nashville, TN 37204

Re: K030857

Trade/Device Name: MAHE Pediatric Naso Pharyngoscope its associated and Ancillary equipment and endoscopic accessories Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: September 24, 2003 Received: September 24, 2003

Dear Mr. Cart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Paul Cart

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

A. Károlyi Károly

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT FOR INDICATIONS FOR USE

510 (k) Number :

Device Name: MAHE International Pediatric NASOPHARYNGOSCOPE

Indications for Use:

The MAHE International Pediatric NASOPHARYNGOSCOPE is indicated for visual examination during a diagnostic and therapeutic procedure within the nasal cavity and nasal pharynx.

#### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRD, Office of Device Evaluation (ODE)

Karin Bohm

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, And Radiological Devices

510(k) Number: K03083

OR Over-the-Counter Use:_ Prescription Use: (Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K030857](https://fda.innolitics.com/submissions/EN/subpart-e%E2%80%94surgical-devices/EOB/K030857)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
